      Faster and improved 3-D head digitization in MEG using Kinect.

      Read this article at

          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Accuracy in localizing the brain areas that generate neuromagnetic activity in magnetoencephalography (MEG) is dependent on properly co-registering MEG data to the participant's structural magnetic resonance image (MRI). Effective MEG-MRI co-registration is, in turn, dependent on how accurately we can digitize anatomical landmarks on the surface of the head. In this study, we compared the performance of three devices-Polhemus electromagnetic system, NextEngine laser scanner and Microsoft Kinect for Windows-for source localization accuracy and MEG-MRI co-registration. A calibrated phantom was used for verifying the source localization accuracy. The Kinect improved source localization accuracy over the Polhemus and the laser scanner by 2.23 mm (137%) and 0.81 mm (50%), respectively. MEG-MRI co-registration accuracy was verified on data from five healthy human participants, who received the digitization process using all three devices. The Kinect device captured approximately 2000 times more surface points than the Polhemus in one third of the time (1 min compared to 3 min) and thrice as many points as the NextEngine laser scanner. Following automated surface matching, the calculated mean MEG-MRI co-registration error for the Kinect was improved by 2.85 mm with respect to the Polhemus device, and equivalent to the laser scanner. Importantly, the Kinect device automatically aligns 20-30 images per second in real-time, reducing the limitations on participant head movement during digitization that are implicit in the NextEngine laser scan (~1 min). We conclude that the Kinect scanner is an effective device for head digitization in MEG, providing the necessary accuracy in source localization and MEG-MRI co-registration, while reducing digitization time.

          Related collections

          Most cited references16

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          Accuracy and Resolution of Kinect Depth Data for Indoor Mapping Applications

          Consumer-grade range cameras such as the Kinect sensor have the potential to be used in mapping applications where accuracy requirements are less strict. To realize this potential insight into the geometric quality of the data acquired by the sensor is essential. In this paper we discuss the calibration of the Kinect sensor, and provide an analysis of the accuracy and resolution of its depth data. Based on a mathematical model of depth measurement from disparity a theoretical error analysis is presented, which provides an insight into the factors influencing the accuracy of the data. Experimental results show that the random error of depth measurement increases with increasing distance to the sensor, and ranges from a few millimeters up to about 4 cm at the maximum range of the sensor. The quality of the data is also found to be influenced by the low resolution of the depth measurements.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Magnetoencephalography: evidence of magnetic fields produced by alpha-rhythm currents.

            D. Cohen (1968)
            Weak alternating magnetic fields outside the human scalp, produced by alpha-rhythm currents, are demonstrated. Subject ard magnetic detector were housed in a multilayer magnetically shielded chamber. Background magnetic noise was reduced by signal-averaging. The fields near the scalp are about 1 x 10(-9) gauss (peak to peak). A course distribution shows left-right symmetry for the particular averaging technique used here.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Suppression of interference and artifacts by the Signal Space Separation Method.

              Multichannel measurement with hundreds of channels oversamples a curl-free vector field, like the magnetic field in a volume free of sources. This is based on the constraint caused by the Laplace's equation for the magnetic scalar potential; outside of the source volume the signals are spatially band limited. A functional solution of Laplace's equation enables one to separate the signals arising from the sphere enclosing the interesting sources, e.g. the currents in the brain, from the magnetic interference. Signal space separation (SSS) is accomplished by calculating individual basis vectors for each term of the functional expansion to create a signal basis covering all measurable signal vectors. Because the SSS basis is linearly independent for all practical sensor arrangements, any signal vector has a unique SSS decomposition with separate coefficients for the interesting signals and signals coming from outside the interesting volume. Thus, SSS basis provides an elegant method to remove external disturbances. The device-independent SSS coefficients can be used in transforming the interesting signals to virtual sensor configurations. This can also be used in compensating for distortions caused by movement of the object by modeling it as movement of the sensor array around a static object. The device-independence of the decomposition also enables physiological DC phenomena to be recorded using voluntary head movements. When used with properly designed sensor array, SSS does not affect the morphology or the signal-to-noise ratio of the interesting signals.
                Bookmark

                Author and article information

                Journal
                Front Neurosci
                Frontiers in neuroscience
                Frontiers Media SA
                1662-4548
                1662-453X
                2014
                : 8
                Affiliations
                [1 ] Biomedical Translational Imaging Centre (BIOTIC), IWK Health Centre Halifax, NS, Canada.
                [2 ] Biomedical Translational Imaging Centre (BIOTIC), IWK Health Centre Halifax, NS, Canada ; Department of Diagnostic Radiology, Dalhousie University Halifax, NS, Canada.
                [3 ] Biomedical Translational Imaging Centre (BIOTIC), IWK Health Centre Halifax, NS, Canada ; Faculty of Computer Science, Dalhousie University Halifax, NS, Canada.
                Article
                10.3389/fnins.2014.00326
                4211394
                25389382
                114377b0-9ba1-4cce-9b8b-66af12d848f5
                History

                Microsoft Kinect,magnetoencephalography (MEG),laser scanner,head position indicator (HPI),localization,alignment,color recognition

                Comments

                Comment on this article