18
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Subcutaneous implantable cardioverter-defibrillator implantation in a patient with bilateral pectoral deep brain stimulators

      Heartrhythm Case Reports
      Elsevier BV

      Read this article at

          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Related collections

          Most cited references3

          • Record: found
          • Abstract: found
          • Article: not found

          Simultaneous thalamic deep brain stimulation and implantable cardioverter-defibrillator.

          Thalamic deep brain stimulation is becoming increasingly popular for the control of drug-refractory tremor. Implantable cardiac pacemakers and defibrillators are commonly used therapeutic modalities. Concerns exist about the potential interactions between these 2 devices in the same patient, but no experience has been reported previously. We describe a patient with essential tremor who had a deep brain stimulator implanted into the left ventral intermediate nucleus of thalamus, who subsequently needed an implantable cardioverter-defibrillator. Despite concerns about possible interactions between the 2 types of implanted electrical devices (i.e., a situation similar to drug-drug interactions), the deep brain stimulator and the implanted pacemaker-defibrillator functioned appropriately, and no interaction occurred in our patient.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            Successful implantation of a subcutaneous cardiac defibrillator in a patient with a preexisting deep brain stimulator

            Introduction In September 2012, the United Stated Food and Drug Administration approved the use of a fully subcutaneous implantable cardiac defibrillator (S-ICD, Boston Scientific Inc). The device is implanted in the left midaxillary space and attached to a single lead that is tunneled subcutaneously from the xyphoid process in 2 directions, superiorly to the sternal manubrium joint to the left of the sternum and laterally to the pulse generator. The lead consists of a single coil in the portion of lead along the sternum and 2 sensing electrodes, 1 at the tip of the lead at the upper portion of the sternum and 1 at the xyphoid process. Sensing is achieved via 1 of 3 potential configurations: between the device and the lower electrode, between the device and the upper electrode, or between the 3 electrodes. The sensing vector is automatically chosen by the device to minimize the chance of T-wave oversensing, but it can be manually overriden. 1 A deep brain stimulator (DBS) is an electronic device consisting of a pulse generator and 1 or more electrodes implanted in the brain. It can be programmed to operate in a bipolar or unipolar stimulation mode. It is used for the treatment of Parkinson disease, among other neurologic conditions. 2 Manufacturer’s recommendations for concomitant use of a transvenous implantable cardiac defibrillator (ICD) and DBS include setting the ICD to bipolar sensing. Sensing in an S-ICD is achieved via much wider bipoles than in a transvenous ICD, raising the concern of adverse interaction between the 2 devices. To our knowledge, we present the first case report of successful implantation of an S-ICD in a patient with a previously implanted DBS. KEY TEACHING POINTS • The subcutaneous cardiac defibrillator (S-ICD) represents a major advance in ICD technology with the ability to provide sudden death prevention without transvenous leads. Because of its wide sensing bipole, interaction with other implanted electronic devices is a concern. This includes patients with a deep brain stimulator (DBS), which is used for treatment of neurologic disorders such as Parkinson disease. • Implantation of an S-ICD in patients with a preexisting DBS requires a multidisciplinary approach with the patient’s neurologist for programming the DBS to a bipolar mode if possible to limit the possibility of interaction with the S-ICD. In addition, technical support should be available during S-ICD implantation to test sensing with different DBS settings and for interrogation of the DBS after defibrillation threshold testing. • This case report outlines an approach that was successful when both devices coexisted in the same patient without any adverse effect on the S-ICD or the patient’s neurologic symptoms. Case report A 51-year-old man presented as an outpatient to our institution for consideration of an S-ICD implantation. His past medical history consisted of coronary artery disease, for which he had undergone placement of multiple coronary stents, and early-onset Parkinson disease, for which he had undergone implantation of a Medtronic Activa DBS in the right prepectoral area. In 1996, he had an episode of polymorphic ventricular tachycardia, which resulted in cardiac arrest. At that time, a single-chamber ICD was implanted in the left prepectoral area for secondary prevention of sudden cardiac death. His left ventricular ejection fraction was and remains normal. Between 1996 and 2005, he underwent 4 ICD generator replacements. His initial right ventricular lead was a Ventritex Cadence single-coil lead, which failed and was replaced by a dual-coil St. Jude Medical Riata lead in 2005. In view of the recent Food and Drug Administration recommendation, 3 the patient underwent routine surveillance imaging of the lead at another institution and was found to have externalization of a conductor on fluoroscopy. In addition, an acute rise in the right ventricular threshold from 1 to 3.5 V was noted. Because of a lack of confidence in the reliability of the Riata lead and the patient’s desire to continue to have protection from sudden cardiac death, the patient was given multiple options, including extraction of the transvenous lead and implantation of new transvenous lead, or abandonment of the leads and implantation of an S-ICD. The decision-making was complicated by the presence of the Medtronic Activa DBS, which had provided him with significant relief from parkinsonian symptoms. In patients with Parkinson disease, the DBS works by bilateral stimulation of the internal globus pallidus or the subthalamic nucleus. Our patient had a single unit with 2 leads, 1 to each cerebral hemisphere. Each lead has 4 electrodes, and the device can be programmed to stimulate in either a unipolar or bipolar fashion. The device can be programmed to a voltage mode or a current mode, and it can deliver 2 to 250 Hz at a pulse width of 60 to 450 μs and up 10.5 V (voltage mode) or 25.5 mA (current mode). The patient’s DBS had been chronically programmed to unipolar stimulation between the DBS pulse generator and the lead(s). As a first step to facilitate S-ICD implantation, we requested that the DBS be changed to a bipolar mode. Symptom relief from parkinsonian symptoms persisted in bipolar mode. The 2 DBS leads in the patient’s DBS can be programmed independently. The left hemisphere lead was programmed to an output of 3.V, and the right hemisphere lead was programmed to 2.1 V. The pulse width and frequency of both leads were the same at 90 μs and 180 Hz, respectively. During implantation of the S-ICD and defibrillation threshold (DFT) testing, these settings were not manipulated. Avoidance of T-wave oversensing by an S-ICD requires screening surface ECG recordings simulating the sensing vectors of the S-ICD. Application of a template provided by the manufacturer determines eligibility, which was adequate in this patient. The patient was taken to the electrophysiology laboratory for implantation of the S-ICD. A programmer and a technician were available to alter the programming of the DBS as needed. The procedure was performed with the patient under general anesthesia. The S-ICD implantation technique has been described elsewhere. 4 We performed the standard technique with a modification: we used a sheath in conjunction with the tunneling tool to place the lead along the left side of the sternum, which avoids the superior third incision. After implantation, we tested for interaction of the S-ICD and the DBS. Changing between unipolar and bipolar stimulation on the DBS was immediately apparent on the surface ECG (Figure 1). The S-ICD sensing vectors were recorded with the DBS in both unipolar and bipolar configurations. There was no oversensing of DBS activity (in both bipolar and unipolar modes) by the S-ICD (Figure 2). DFT testing was performed with successful sensing and termination of induced ventricular fibrillation at 65, 50, and 35 J. The DBS was active in bipolar mode during DFT testing. We interrogated the DBS after DFT testing and found no interruption in normal function. On routine follow-up 12 months after implantation, the patient was doing well with no complications. He had not received any S-ICD shocks. Chest x-ray film showed a well-positioned S-ICD device and electrode (Figure 3). Discussion DBS is an increasingly common treatment for a variety of neurologic disorders, including Parkinson disease, so the possibility of a patient requiring both an ICD and a DBS is increasing. Three previous case reports have documented the safety and lack of interaction between transvenous ICDs and DBS. However, 1 case report did document resetting of a DBS to an off mode after DFT testing of a transvenous ICD. 5 To our knowledge, this is the first report of the safe implantation and follow-up of an S-ICD with a DBS. There was no issue acutely with interaction of the DBS and ICD with the DBS in either bipolar or unipolar mode, and in all 3 sensing vectors of the S-ICD. The lack of DBS artifact on the S-ICD (even with unipolar DBS stimulation) likely is due to signal filtering in the S-ICD. The DBS was programmed to a stimulation frequency of 180 Hz both chronically and during the implantation. The S-ICD allows frequencies between only 3 and 40 Hz to pass and thus eliminates the DBS signal. In contrast, the recording system in the electrophysiology laboratory where the case was performed allows frequencies of 30 to 250 Hz to pass, making DBS unipolar stimulation apparent on the surface ECG. Bipolar DBS stimulation was not seen on the ECG recording, probably because it was of much lower amplitude than unipolar stimulation. Sensing of ventricular fibrillation by the S-ICD was unaffected by active bipolar stimulation from the DBS. In addition, the 3 ICD shocks delivered for DFT testing did not adversely affect the DBS. It is important to note that, during follow-up, the DBS was left in bipolar mode and the S-ICD remained in its automatically selected ideal sensing vector. It is not clear from this report whether with the DBS in unipolar mode or a different S-ICD sensing vector and the DBS in either mode that oversensing and interaction may not have occurred during follow-up. DBS devices can also be programmed to frequencies that are well within the filter pass range of the S-ICD, which may result in a higher risk of interaction. In addition, it is not clear whether S-ICD sensing chronically or DBS function after an S-ICD shock would be unaffected if the DBS were on the left side of the patient. An additional issue that may arise with the combination of these 2 devices is the manner in which the DBS behaves after a power on reset event, which theoretically can occur after an ICD shock. The current generation of Medtronic devices, 1 of which was present in this patient, resets to the previously programmed parameters, even after multiple resets. However, older generations of DBS devices (which still are available but infrequently used) will revert to the default settings, which vary, but do include, in some instances, a stimulation frequency of 30 Hz, which is within the pass filters limits of the S-ICD and would increase the risk of oversensing of DBS stimulation by the ICD. For patients with an S-ICD, such older-generation DBS devices probably should be avoided. For patients with preexisting older-generation DBS, this issue should be taken into account when considering a new S-ICD implantation. Given the inherent limitations of a single case report, caution should be exercised in applying these findings to similar clinical situations. Ideally, a series of such cases would be useful to better understand the potential interactions and issues that may arise when these devices coexist in a single patient.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Implantation of a cardiac resynchronization therapy defibrillator in a patient with bilateral deep brain stimulator: feasibility and technique.

              Uncommonly, a patient with a NeuroStimulator Device (NSD) for a neurologic indication requires a cardiovascular implantable electronic device (CIED) for a cardiac indication. Typically in those with a unilateral pectoral NSD, the contralateral pectoral space is used for CIED implantation; however, in very rare occasions the patient has bilateral pectoral NSDs which makes subsequent implantation of a CIED challenging both because of placement and device interaction. Herein, we introduce the case of a 68-year-old gentleman with bilateral pectoral deep brain stimulators for Parkinsonism who received cardiac resynchronization therapy-defibrillator (CRT-D) for advanced heart failure. The CRT-D generator was implanted in the abdomen and the CRT-D leads were placed through a minimally invasive epicardial approach. Both devices were tested without any evidence of device interaction.
                Bookmark

                Author and article information

                Journal
                10.1016/j.hrcr.2017.12.005
                http://creativecommons.org/licenses/by-nc-nd/4.0/

                Comments

                Comment on this article