      Sizing of atrial septal defects in adults.

      Radiology
      Adult, Cardiac Catheterization, Echocardiography, Transesophageal, Female, Heart Septal Defects, Atrial, physiopathology, surgery, ultrasonography, Humans, Male, Middle Aged

      Read this article at

      ScienceOpenPublisherPubMed
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          In a retrospective study of 51 consecutive patients undergoing transcatheter closure of secundum type atrial septal defects (ASDs), we examined the reliability of transesophageal echocardiography (TEE) prior to catheterization and compared the diameter with that obtained by balloon measurement during catheterization. The TEE diameter was 16.3+/-4.6 mm compared with 22.5+/-6.0 mm for the stretched diameter obtained during catheterization (p<0.001). There was no gender difference. The degree of left-to-right shunting correlated poorly with the size of the defect. We conclude that although TEE is accurate for diagnosis of an ASD, the measurement of its size to determine the size of the closure device is at best inaccurate. Copyright (c) 2005 S. Karger AG, Basel.

          Related collections

          Most cited references9

          • Record: found
          • Abstract: found
          • Article: not found

          Echocardiographic estimation of balloon-stretched diameter of secundum atrial septal defect for transcatheter occlusion.

          Stretched diameter of the atrial septal defect (ASD), determined by balloon sizing at cardiac catheterization, is commonly used to select the sizes of the devices used for transcatheter closure of the secundum ASD. We have previously evaluated the utility of pulmonary/systemic flow ratio and angiographic and echocardiographic (echo) sizes of the ASD in estimating stretched ASD diameter in a group of 16 patients and determined that echo diameter had the best correlation with stretched diameter (r = 0.82; p less than 0.001). The stretched diameter can be estimated: 1.05 x echo diameter in millimeters + 5.49. In this study we have prospectively evaluated this formula in estimating the stretched ASD diameter by two-dimensional echo measurements obtained in two (long and short-axis) subcostal views in another group of 21 patients aged 2.5 to 29 years (median 4.5 years). The echo size of the ASD was 9.7 +/- 3.0 mm, whereas the measured stretched diameter was 15.3 +/- 4.0 mm. The predicted stretched ASD diameter was calculated according to the above formula and was 15.7 +/- 3.1 mm, not significantly different (p greater than 0.1) from the measured stretched diameter. The correlation between predicted and measured stretched ASD sizes was excellent (r = 0.9; p less than 0.001). The mean squared error was 2.4. The differences between measured and predicted values were within 2 mm in all but three patients. It is concluded that stretched ASD diameter can be estimated accurately by two-dimensional subcostal echo measurements, which in turn could be used for selection of device size for occlusion of the ASD.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Morphological study of defects of the atrial septum within the oval fossa: implications for transcatheter closure of left-to-right shunt.

            To determine the anatomical variability of the oval fossa in cases of atrial septal defect and to find out which factors might make such defects suitable or unsuitable for closure by umbrella or clamshell devices. 100 specimens with defects of the atrial septum within the oval fossa were studied, especially the position of the defects within the fossa; the area of the defect in relation to the total area of the oval fossa; the shape of the rims and flap valve of the oval fossa; and the anatomical variability in the eustachian and thebesian valves. The oval fossa was displaced to the mouth of the inferior caval vein in four cases; displaced to the mouth of superior caval vein in two cases; placed on the middle of the interatrial wall in 43; or placed slightly towards the inferior caval vein in 51. Because of their shape 29 of the hearts were considered to be unsuitable for transcatheter closure of the defect. This was because the defect was too large (16 hearts); the oval fossa was displaced to the mouth of the inferior caval vein (four hearts) or to the superior caval vein (two hearts); lacked its anterior rim (two hearts); lacked the posteroinferior rim (one heart); because a thick eustachian valve was displaced posteriorly forming a false posteroinferior rim (one heart); or because the strand of insertion of the floor of the oval fossa was too distant from the left atrial aspect (three hearts). 68 hearts appeared to be ideal candidates for transcatheter closure; 3 would probably have been suitable; but 29 were unsuitable. These morphological variations might explain why the procedure has been unsuccessful in so many cases. Patients should be screened before any attempt is made at transcatheter closure.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Relationship of echocardiographic, shunt flow, and angiographic size to the stretched diameter of the atrial septal defect.

              Stretched diameter of the atrial septal defect (ASD), measured by balloon sizing, is generally used as a guide to the selection of the size of the device utilized for transcatheter closure of the ASD. Balloon sizing is a cumbersome procedure and sometimes requires the use of very large size balloon catheters. Several methods of assessment of ASD size, namely, echographic, pulmonary-to-systemic flow ratio (Qp:Qs), and angiographic measures, were undertaken in a group of 16 patients, aged 7 months to 45 years (median, 4.5 years), who were being evaluated for transcatheter closure of ASD; the results were compared with the stretched diameter. Although the echographic size of the ASD (9.9 +/- 4.1 mm, mean +/- SD) is similar (p greater than 0.1) to the angiographic size (7.9 +/- 2.5 mm), it is much smaller (p less than 0.01) than the stretched diameter (16.1 +/- 5.3 mm). When the relationship between various measures of ASD was examined, although the Qp:Qs ratio and angiographic size have a significant (p less than 0.05) correlation with the stretched diameter (r = 0.55 and 0.54, respectively), the echo diameter has the best correlation coefficient, r = 0.82, p less than 0.001. The stretched diameter can be estimated by the equation: 1.05 x echo + 5.49 mm. It is concluded that the echographic diameter is a useful adjunct in the estimation of the stretch ASD diameter, which in turn can be used in the selection of the size of the device for transcatheter occlusion of the ASD.
                Bookmark

                Author and article information

                Comments

                Comment on this article