135
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      First report of concomitant subcutaneous defibrillator and phrenic nerve stimulator implantation in a patient with severe central sleep apnea and left ventricular systolic dysfunction

      , , , , ,
      HeartRhythm Case Reports
      Elsevier BV

      Read this article at

          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Introduction There is a clinically significant association between central sleep apnea (CSA) and heart failure with reduced ejection fraction (HFrEF). 1 , 2 The presence of CSA has been associated with an increased risk of decompensated heart failure and death 3 and the prevalence of CSA increases in parallel with the severity of HFrEF. Phrenic nerve stimulation (PNS) using the remedē system (Respicardia, Minnetonka, MN) is a new therapeutic modality to treat CSA that involves placing transvenous leads to stimulate the phrenic nerve and activate the diaphragm. Since many patients with severe HFrEF have implanted defibrillators, a clinical scenario where concomitant CSA and defibrillator therapy are necessary is likely to occur. However, data are lacking concerning potential interaction between these devices. We present a case report of a patient with severe ischemic cardiomyopathy with a pre-existing subcutaneous implantable cardioverter-defibrillator (S-ICD) who was later diagnosed with severe CSA and underwent successful remedē system implantation. Case report A 54-year-old man with poorly controlled type 1 diabetes mellitus, end-stage kidney disease on hemodialysis, paroxysmal atrial fibrillation, and ischemic cardiomyopathy with HFrEF (left ventricular ejection fraction 30%) underwent a primary prevention Emblem S-ICD insertion (Boston Scientific, St. Paul, MN) in December 2015. The sensing vector was secondary. Conversion testing of ventricular fibrillation was successfully performed with 65 J shock and the shock impedance was measured as 46 ohms. In November 2018, an in-laboratory polysomnogram was obtained to evaluate for obstructive sleep apnea owing to complaints of daytime sleepiness, fatigue, and difficulty initiating and maintaining sleep. At that time his body mass index was 20.4 kg/m2. His most recent echocardiogram demonstrated further worsening of heart function with a left ventricle ejection fraction of 14%. The polysomnogram revealed severe sleep-disordered breathing with predominant CSA. The total apnea-hypopnea index was 61 events per hour (normal: less than 5 events per hour; severe: more than 30 events per hour). The central apnea index was 42 events per hour and the obstructive apnea-hypopnea index was 19 events per hour. Sleep-disordered breathing was also accompanied by significant sleep hypoxemia. Figure 1 illustrates a representative 2-minute window from the overnight polysomnogram demonstrating central apneas. After failing continuous positive airway pressure (CPAP) titration, the patient underwent bilevel positive airway pressure spontaneous-timed (PAP-ST) titration in the sleep laboratory and was prescribed home PAP-ST therapy with settings of 25/16 cm H2O with a backup respiratory rate of 16 breaths per minute with an oronasal mask. Adequate adherence was shown during the first 30 days of therapy, as obtained by downloading data from the patient’s device. The device-estimated residual apnea-hypopnea index was 22 events per hour (residual central apnea index of 14 and hypopnea index of 8). He did not feel any improvement in his symptoms despite adequate adherence to bilevel PAP-ST therapy. Given the persistence of symptoms and persistent CSA despite bilevel PAP-ST therapy, he was referred for phrenic nerve stimulator implantation. Figure 1 Polysomnogram showing central apneas. Two-minute window from overnight polysomnogram demonstrating frequent central apneas with Cheyne-Stokes respiration. There are 2 central apneas during the representative 2-minute window recorded during stage 2 non–rapid eye movement (N2) sleep based on the 6 electroencephalography leads (F3-M2, C3-M2, O1-M2, F4-M1, C4-M1, O2-M1) with the patient in the right lateral decubitus position. Each central apnea lasts approximately 30 seconds and is characterized by a lack of airflow on the nasal pressure transducer (NPT) and the oronasal thermistor (Therm), no effort on the chest and abdominal respiratory inductance plethysmography belts, followed by a 10% oxygen desaturation (in yellow). Each cycle, from the beginning of an apnea to the end of the next apnea, lasts approximately 60 seconds and the microarousals occur at the peak of ventilatory effort, consistent with central sleep apnea with Cheyne-Stokes respiration pattern. Implantation of the remedē system was performed in the electrophysiology laboratory under moderate sedation via the right subclavian vein access. The S-ICD was temporarily disabled to allow for electrocautery. The implantation technique of the remedē system has been described previously 4 but in short, we inserted a multipolar pacing lead Respistim LQS model 4065 (Respicardia, Minnetonka, MN,) into the left pericardiophrenic vein, which courses over the lateral border of the epicardium and runs parallel with the left phrenic nerve. A second lead, Medtronic Attain Ability model 4196 (Medtronic, Minneapolis, MN), was inserted into a low right intercostal vein via the azygos vein. This lead is used for monitoring transthoracic impedance, which is a surrogate for respiration. The generator was placed in a pectoral pocket (Figure 2). Figure 2 Chest radiograph showing implanted hardware. Anteroposterior (AP) chest radiograph of right pectoral position of phrenic nerve stimulator with multipolar pacing lead in left pericardiophrenic vein and sensing lead in low right intercostal vein via the azygos vein. Subcutaneous implantable cardioverter-defibrillator (S-ICD) is in its usual position on the left lateral thorax with corresponding suprasternal lead. A = distal electrode; AL = alternate S-ICD vector; B = proximal electrode; PR = primary S-ICD vector; SC = secondary S-ICD vector. Because of this patient’s pre-existing S-ICD, the potential for device-device interaction was systematically assessed. The S-ICD was re-enabled. With maximum Respistim LQS bipolar pacing output (10 mA, 300 microseconds, 20 Hz) from all 4 pacing electrodes, subcutaneous signals were recorded from all 3 S-ICD vectors. Pacing artifact was evident in the primary and secondary vectors but absent in the alternate. While the pacing artifact was visible, it was not sensed as noise. During the remedē impedance check, 4 0.5-mA pulses were delivered and the resultant artifact was recorded on the subcutaneous signal. The sinus beat preceding the impedance check stimulus was labeled as noise and the stimulus was sensed (Figure 3). The S-ICD sensing vector was programmed as secondary. The leads were connected to the generator and the system was implanted in a right pectoral pocket with the aid of a TYRX antimicrobial envelope (Medtronic, St. Paul, MN). The remedē system was not activated, as is recommended, to allow for the leads to settle in and stabilize. The patient was discharged the next day and 6 weeks later was brought back for remedē system activation. With S-ICD sensing programmed to secondary vector, device-device interaction was assessed in the same fashion as at the time of the initial implant. There was no noise detected with up-titration of pacing output. With impedance testing, artifact similar to what was seen during implant was present, but this finding was not sensed reproducibly. The remedē system was activated. After 6 weeks of PNS, he is feeling better and S-ICD interrogation has not revealed any shocks. A repeat polysomnogram is scheduled in 2 months. Figure 3 Subcutaneous electrocardiography signals recorded from the subcutaneous implantable cardioverter-defibrillator during phrenic nerve stimulation impedance testing. During impedance testing by the remedē device (Respicardia, Minnetonka, MN), the stimulus (red arrow) was recorded and labeled as a sensed event (S). The preceding sinus beat (which is on time with the patient’s intrinsic rhythm) was labeled as noise (N). This finding was reproducible during implant but not evident at the follow-up visit. Discussion To our knowledge, this is the first report of concomitant S-ICD and phrenic nerve stimulator implantation. The prevalence of sleep apnea, both obstructive and central, is high among patients with congestive heart failure, approaching 50%–75% in some studies. 1 , 5 Untreated sleep apnea is associated with high morbidity and mortality. 6 , 7 Obstructive sleep apnea is more commonly recognized and is usually treated with positive pressure therapy such as CPAP. Prior to PNS, there were no effective treatments for CSA, as CPAP does not effectively treat central apneas 8 and adaptive servoventilation has been associated with a higher mortality in patients with HFrEF, 9 particularly in patients with an ejection fraction below 30%. 10 PNS with the remedē system has been shown to improve symptoms and reduce central apneas and is currently indicated for patients diagnosed with moderate-to-severe CSA. 11 , 12 Both sensing and pacing leads of the remedē system are programmable. The quadripolar left lead can be programmed in multipolar pacing configuration to deliver output between 0.1 and 10 mA at a pulse width of 60–300 microseconds. Most patients require up-titration of PNS pacing output during follow-up, which is why it is important to test for interaction at maximum output and with all available pacing options. There are 3 sensing vectors available on the S-ICD that are used for detecting malignant ventricular arrhythmias (primary, secondary, and alternate) (Figure 2). The S-ICD algorithm automatically chooses one based on subcutaneous signal amplitude, morphology, and quality. 13 A different sensing vector can be selected by the implanting physician if desired. Although an artifact was detected on primary and secondary vectors during pacing and impedance testing of the remedē system, the alternate vector was spared, as it senses signals from an area outside the PNS (ie, between distal and proximal lead electrodes; Figure 2). While it may seem desirable to manually select the alternate vector in these cases, it is often the one least chosen by the S-ICD algorithm owing to poor signal quality. 13 Regardless, even with maximum pacing output from the PNS, there was no noise or pacing artifact sensed. While artifact was recorded and intermittently sensed during PNS impedance testing, this finding was not reproducible. Because the S-ICD has limited storage capability, impedance testing artifact, even if it occurred during auto–impedance testing by the remedē device, would not likely be saved and available for review. Conclusion Our case is the first example of concomitant PNS and S-ICD implantation. Given the relatively high prevalence of CSA in the HFrEF population, situations where these 2 therapies could be utilized concurrently will become more commonplace. We did not see any significant interactions, but our follow-up is short. More data are needed before we can definitively say that concomitant PNS and S-ICD therapy is safe. Key Teaching Points • The prevalence of central sleep apnea (CSA) is relatively high in patients with heart failure with reduced ejection fraction and can lead to significant morbidity and mortality if left untreated. Phrenic nerve stimulation (PNS) is a desirable option in patients with systolic dysfunction, as continuous positive airway pressure is not effective in treating CSA. • The artifact generated during impedance testing by the remedē device (Respicardia, Minnetonka, MN) could theoretically interfere with subcutaneous implantable cardioverter-defibrillator function but is not demonstrated in this case despite rigorous testing. • Because up-titration of pacing output and/or pacing pole reprogramming is often needed after PNS implant, potential device-device interaction should be assessed at implant and during follow-up using maximum pacing output in all available PNS pacing configurations.

          Related collections

          Most cited references10

          • Record: found
          • Abstract: found
          • Article: not found

          Continuous positive airway pressure for central sleep apnea and heart failure.

          The Canadian Continuous Positive Airway Pressure for Patients with Central Sleep Apnea and Heart Failure trial tested the hypothesis that continuous positive airway pressure (CPAP) would improve the survival rate without heart transplantation of patients who have central sleep apnea and heart failure. After medical therapy was optimized, 258 patients who had heart failure (mean age [+/-SD], 63+/-10 years; ejection fraction, 24.5+/-7.7 percent) and central sleep apnea (number of episodes of apnea and hypopnea per hour of sleep, 40+/-16) were randomly assigned to receive CPAP (128 patients) or no CPAP (130 patients) and were followed for a mean of two years. During follow-up, sleep studies were conducted and measurements of the ejection fraction, exercise capacity, quality of life, and neurohormones were obtained. Three months after undergoing randomization, the CPAP group, as compared with the control group, had greater reductions in the frequency of episodes of apnea and hypopnea (-21+/-16 vs. -2+/-18 per hour, P<0.001) and in norepinephrine levels (-1.03+/-1.84 vs. 0.02+/-0.99 nmol per liter, P=0.009), and greater increases in the mean nocturnal oxygen saturation (1.6+/-2.8 percent vs. 0.4+/-2.5 percent, P<0.001), ejection fraction (2.2+/-5.4 percent vs. 0.4+/-5.3 percent, P=0.02), and the distance walked in six minutes (20.0+/-55 vs. -0.8+/-64.8 m, P=0.016). There were no differences between the control group and the CPAP group in the number of hospitalizations, quality of life, or atrial natriuretic peptide levels. An early divergence in survival rates without heart transplantation favored the control group, but after 18 months the divergence favored the CPAP group, yet the overall event rates (death and heart transplantation) did not differ (32 vs. 32 events, respectively; P=0.54). Although CPAP attenuated central sleep apnea, improved nocturnal oxygenation, increased the ejection fraction, lowered norepinephrine levels, and increased the distance walked in six minutes, it did not affect survival. Our data do not support the use of CPAP to extend life in patients who have central sleep apnea and heart failure. Copyright 2005 Massachusetts Medical Society.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Sleep-disordered breathing in patients with symptomatic heart failure: a contemporary study of prevalence in and characteristics of 700 patients.

            Evaluation of the prevalence and nature of sleep-disordered breathing (SDB) in patients with symptomatic chronic heart failure (CHF) receiving therapy according to current guidelines. We prospectively screened 700 patients with CHF (NYHA class> or =II, LV-EF< or =40%) for SDB using cardiorespiratory polygraphy (Embletta). Furthermore, echocardiography, cardiopulmonary exercise and 6-min walk testing were performed. Medication included ACE-inhibitors and/or AT1-receptor blockers in at least 94%, diuretics in 87%, beta-blockers in 85%, digitalis in 61% and spironolactone in 62% of patients. SDB was present in 76% of patients (40% central (CSA), 36% obstructive sleep apnoea (OSA)). CSA patients were more symptomatic (NYHA class 2.9+/-0.5 vs. no SDB 2.57+/-0.5 or OSA 2.57+/-0.5; p<0.05) and had a lower LV-EF (27.4+/-6.6% vs. 29.3+/-2.6%, p<0.05) than OSA patients. Oxygen uptake (VO(2)) was lowest in CSA patients: predicted peak VO(2) 57+/-16% vs. 64+/-18% in OSA and 63+/-17% in no SDB, p<0.05. 6-min walking distances were 331+/-111 m in CSA, 373+/-108 m in OSA and 377+/-118 m in no SDB (p<0.05). This study confirms the high prevalence of SDB, particularly CSA in CHF patients. CSA seems to be a marker of heart failure severity.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Central sleep apnea, right ventricular dysfunction, and low diastolic blood pressure are predictors of mortality in systolic heart failure.

              The purpose of this study was to determine whether central sleep apnea (CSA) contributes to mortality in patients with heart failure (HF). Cheyne-Stokes breathing with CSA commonly occurs in patients with systolic HF. Consequences of CSA, including altered blood gases and neurohormonal activation, could result in further left ventricular dysfunction. Therefore, we hypothesized that CSA might contribute to mortality of patients with HF. We followed 88 patients with systolic HF (left ventricular ejection fraction < or =45%) with (n = 56) or without (n = 32) CSA. The median follow-up was 51 months. The mean (+/-SD) of apnea-hypopnea index was significantly higher in patients with CSA (34 +/- 25/h) than those without CSA (2 +/- 1/h). Most of these events were central apneas. In Cox multiple regression analysis, 3 of 24 confounding variables independently correlated with survival. The median survival of patients with CSA was 45 months compared with 90 months of those without CSA (hazard ratio = 2.14, p = 0.02). The other 2 variables that correlated with poor survival were severity of right ventricular systolic dysfunction and low diastolic blood pressure. In patients with systolic HF, CSA, severe right ventricular systolic dysfunction, and low diastolic blood pressure might have an adverse effect on survival.
                Bookmark

                Author and article information

                Journal
                HeartRhythm Case Reports
                HeartRhythm Case Reports
                Elsevier BV
                22140271
                October 2019
                October 2019
                Article
                10.1016/j.hrcr.2019.10.004
                53ee9215-32cf-463b-b427-633cb1b798e7
                © 2019

                https://www.elsevier.com/tdm/userlicense/1.0/

                http://creativecommons.org/licenses/by-nc-nd/4.0/

                History

                Comments

                Comment on this article