      Diagnostic performance of image navigated coronary CMR angiography in patients with coronary artery disease

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          The use of coronary MR angiography (CMRA) in patients with coronary artery disease (CAD) remains limited due to the long scan times, unpredictable and often non-diagnostic image quality secondary to respiratory motion artifacts. The purpose of this study was to evaluate CMRA with image-based respiratory navigation (iNAV CMRA) and compare it to gold standard invasive x-ray coronary angiography in patients with CAD.

          Methods

          Consecutive patients referred for CMR assessment were included to undergo iNAV CMRA on a 1.5 T scanner. Coronary vessel sharpness and a visual score were assigned to the coronary arteries. A diagnostic reading was performed on the iNAV CMRA data, where a lumen narrowing >50% was considered diseased. This was compared to invasive x-ray findings.

          Results

          Image-navigated CMRA was performed in 31 patients (77% male, 56 ± 14 years). The iNAV CMRA scan time was 7 min:21 s ± 0 min:28 s. Out of a possible 279 coronary segments, 26 segments were excluded from analysis due to stents or diameter less than 1.5 mm, resulting in a total of 253 coronary segments. Diagnostic image quality was obtained for 98% of proximal coronary segments, 94% of middle segments, and 91% of distal coronary segments. The sensitivity and specificity was 86% and 83% per patient, 80% and 92% per vessel and 73% and 95% per segment.

          Conclusion

          In this study, iNAV CMRA offered a very good diagnostic performance when compared against invasive x-ray angiography. Due to the short and predictable scan time it can add clinical value as a part of a comprehensive CAD assessment protocol.

          Related collections

          Most cited references23

          • Record: found
          • Abstract: found
          • Article: not found

          Whole-heart coronary MR angiography with 2D self-navigated image reconstruction.

          Several self-navigation techniques have been proposed to improve respiratory motion compensation in coronary MR angiography. In this work, we implemented a 2D self-navigation method by using the startup profiles of a whole-heart balanced Steady-state free precession sequence, which are primarily used to catalyze the magnetization towards the steady-state. To create 2D self-navigation images (2DSN), we added phase encoding gradients to the startup profiles. With this approach we calculated foot-head and left-right motion and performed retrospective translational motion correction. The 2DSN images were reconstructed from 10 startup profiles acquired at the beginning of each shot. Nine healthy subjects were scanned, and the proposed method was compared to a 1D self-navigation (1DSN) method with foot-head correction only. Foot-head correction was also performed with the diaphragmatic 1D pencil beam navigator (1Dnav) using a tracking factor of 0.6. 2DSN shows improved motion correction compared to 1DSN and 1Dnav for all coronary arteries and all subjects for the investigated diaphragmatic gating window of 10 mm. The visualized vessel length of the right coronary artery could be significantly improved with a multiple targeted 2D self-navigation approach, compared to 2DSN method. Copyright © 2011 Wiley Periodicals, Inc.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            "Soap-Bubble" visualization and quantitative analysis of 3D coronary magnetic resonance angiograms.

            In order to compare coronary magnetic resonance angiography (MRA) data obtained with different scanning methodologies, adequate visualization and presentation of the coronary MRA data need to be ensured. Furthermore, an objective quantitative comparison between images acquired with different scanning methods is desirable. To address this need, a software tool ("Soap-Bubble") that facilitates visualization and quantitative comparison of 3D volume targeted coronary MRA data was developed. In the present implementation, the user interactively specifies a curved subvolume (enclosed in the 3D coronary MRA data set) that closely encompasses the coronary arterial segments. With a 3D Delaunay triangulation and a parallel projection, this enables the simultaneous display of multiple coronary segments in one 2D representation. For objective quantitative analysis, frequently explored quantitative parameters such as signal-to-noise ratio (SNR); contrast-to-noise ratio (CNR); and vessel length, sharpness, and diameter can be assessed. The present tool supports visualization and objective, quantitative comparisons of coronary MRA data obtained with different scanning methods. The first results obtained in healthy adults and in patients with coronary artery disease are presented. Copyright 2002 Wiley-Liss, Inc.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Free-breathing whole-heart coronary MRA with 3D radial SSFP and self-navigated image reconstruction.

              Respiratory motion is a major source of artifacts in cardiac magnetic resonance imaging (MRI). Free-breathing techniques with pencil-beam navigators efficiently suppress respiratory motion and minimize the need for patient cooperation. However, the correlation between the measured navigator position and the actual position of the heart may be adversely affected by hysteretic effects, navigator position, and temporal delays between the navigators and the image acquisition. In addition, irregular breathing patterns during navigator-gated scanning may result in low scan efficiency and prolonged scan time. The purpose of this study was to develop and implement a self-navigated, free-breathing, whole-heart 3D coronary MRI technique that would overcome these shortcomings and improve the ease-of-use of coronary MRI. A signal synchronous with respiration was extracted directly from the echoes acquired for imaging, and the motion information was used for retrospective, rigid-body, through-plane motion correction. The images obtained from the self-navigated reconstruction were compared with the results from conventional, prospective, pencil-beam navigator tracking. Image quality was improved in phantom studies using self-navigation, while equivalent results were obtained with both techniques in preliminary in vivo studies. 2005 Wiley-Liss, Inc
                Bookmark

                Author and article information

                Contributors
                +44 (0)20 718 85441 , markus.henningsson@kcl.ac.uk
                joy.shome@kcl.ac.uk
                c_bratis@hotmail.com
                miguel.silvavieira@kcl.ac.uk
                eike.nagel@cardiac-imaging.org
                rene.botnar@kcl.ac.uk
                Journal
                J Cardiovasc Magn Reson
                J Cardiovasc Magn Reson
                Journal of Cardiovascular Magnetic Resonance
                BioMed Central (London )
                1097-6647
                1532-429X
                11 September 2017
                11 September 2017
                2017
                : 19
                : 68
                Affiliations
                [1 ]ISNI 0000 0001 2322 6764, GRID grid.13097.3c, Division of Biomedical Engineering and Imaging Sciences, , King’s College London, ; London, UK
                [2 ]ISNI 0000 0004 1936 9721, GRID grid.7839.5, Institute for Experimental and Translational Cardiovascular Imaging, , Goethe University, ; Frankfurt/Main, Germany
                [3 ]GRID grid.452396.f, DZHK (German Centre for Cardiovascular Research, Standort RheinMain), ; Berlin, Germany
                [4 ]ISNI 0000 0001 2157 0406, GRID grid.7870.8, Escuela de Ingeniería, , Pontificia Universidad Católica de Chile, ; Santiago, Chile
                Article
                381
                10.1186/s12968-017-0381-3
                5594598
                28893296
                0050e017-a0c5-49fc-a892-f0b6f5b32acb
                © The Author(s). 2017

                Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License ( http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

                History
                : 23 May 2017
                : 17 August 2017
                Funding
                Funded by: FundRef http://dx.doi.org/10.13039/501100000274, British Heart Foundation;
                Award ID: RE/08/03
                Award ID: RG/12/1/29262
                Funded by: FundRef http://dx.doi.org/10.13039/100004440, Wellcome Trust;
                Award ID: WT 088641/Z/09/Z
                Categories
                Research
                Custom metadata
                © The Author(s) 2017

                Cardiovascular Medicine
                coronary mr angiography,image navigators,respiratory motion correction,coronary artery disease

                Comments

                Comment on this article