      Development of real‐time motion verification system using in‐room optical images for respiratory‐gated radiotherapy

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Phase‐based respiratory‐gated radiotherapy relies on the reproducibility of patient breathing during the treatment. To monitor the positional reproducibility of patient breathing against a 4D CT simulation, we developed a real‐time motion verification system (RMVS) using an optical tracking technology. The system in the treatment room was integrated with a real‐time position management system. To test the system, an anthropomorphic phantom that was mounted on a motion platform moved on a programmed breathing pattern and then underwent a 4D CT simulation with RPM. The phase‐resolved anterior surface lines were extracted from the 4D CT data to constitute 4D reference lines. In the treatment room, three infrared reflective markers were attached on the superior, middle, and inferior parts of the phantom along with the body midline and then RMVS could track those markers using an optical camera system. The real‐time phase information extracted from RPM was delivered to RMVS via in‐house network software. Thus, the real‐time anterior‐posterior positions of the markers were simultaneously compared with the 4D reference lines. The technical feasibility of RMVS was evaluated by repeating the above procedure under several scenarios such as ideal case (with identical motion parameters between simulation and treatment), cycle change, baseline shift, displacement change, and breathing type changes (abdominal or chest breathing). The system capability for operating under irregular breathing was also investigated using real patient data. The evaluation results showed that RMVS has a competence to detect phase‐matching errors between patient's motion during the treatment and 4D CT simulation. Thus, we concluded that RMVS could be used as an online quality assurance tool for phase‐based gating treatments.

          PACS number: 87.55.Qr

          Related collections

          Most cited references27

          • Record: found
          • Abstract: found
          • Article: not found

          Determining parameters for respiration-gated radiotherapy.

          Respiration-gated radiotherapy for tumor sites affected by respiratory motion will potentially improve radiotherapy outcomes by allowing reduced treatment margins leading to decreased complication rates and/or increased tumor control. Furthermore, for intensity-modulated radiotherapy (IMRT), respiratory gating will minimize the hot and cold spot artifacts in dose distributions that may occur as a result of interplay between respiratory motion and leaf motion. Most implementations of respiration gating rely on the real time knowledge of the relative position of the internal anatomy being treated with respect to that of an external marker. A method to determine the amplitude of motion and account for any difference in phase between the internal tumor motion and external marker motion has been developed. Treating patients using gating requires several clinical decisions, such as whether to gate during inhale or exhale, whether to use phase or amplitude tracking of the respiratory signal, and by how much the intrafraction tumor motion can be decreased at the cost of increased delivery time. These parameters may change from patient to patient. A method has been developed to provide the data necessary to make decisions as to the CTV to PTV margins to apply to a gated treatment plan.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Audio-visual biofeedback for respiratory-gated radiotherapy: impact of audio instruction and audio-visual biofeedback on respiratory-gated radiotherapy.

            Respiratory gating is a commercially available technology for reducing the deleterious effects of motion during imaging and treatment. The efficacy of gating is dependent on the reproducibility within and between respiratory cycles during imaging and treatment. The aim of this study was to determine whether audio-visual biofeedback can improve respiratory reproducibility by decreasing residual motion and therefore increasing the accuracy of gated radiotherapy. A total of 331 respiratory traces were collected from 24 lung cancer patients. The protocol consisted of five breathing training sessions spaced about a week apart. Within each session the patients initially breathed without any instruction (free breathing), with audio instructions and with audio-visual biofeedback. Residual motion was quantified by the standard deviation of the respiratory signal within the gating window. Audio-visual biofeedback significantly reduced residual motion compared with free breathing and audio instruction. Displacement-based gating has lower residual motion than phase-based gating. Little reduction in residual motion was found for duty cycles less than 30%; for duty cycles above 50% there was a sharp increase in residual motion. The efficiency and reproducibility of gating can be improved by: incorporating audio-visual biofeedback, using a 30-50% duty cycle, gating during exhalation, and using displacement-based gating.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Patient training in respiratory-gated radiotherapy.

              Respiratory gating is used to counter the effects of organ motion during radiotherapy for chest tumors. The effects of variations in patient breathing patterns during a single treatment and from day to day are unknown. We evaluated the feasibility of using patient training tools and their effect on the breathing cycle regularity and reproducibility during respiratory-gated radiotherapy. To monitor respiratory patterns, we used a component of a commercially available respiratory-gated radiotherapy system (Real Time Position Management (RPM) System, Varian Oncology Systems, Palo Alto, CA 94304). This passive marker video tracking system consists of reflective markers placed on the patient's chest or abdomen, which are detected by a wall-mounted video camera. Software installed on a PC interfaced to this camera detects the marker motion digitally and records it. The marker position as a function of time serves as the motion signal that may be used to trigger imaging or treatment. The training tools used were audio prompting and visual feedback, with free breathing as a control. The audio prompting method used instructions to "breathe in" or "breathe out" at periodic intervals deduced from patients' own breathing patterns. In the visual feedback method, patients were shown a real-time trace of their abdominal wall motion due to breathing. Using this, they were asked to maintain a constant amplitude of motion. Motion traces of the abdominal wall were recorded for each patient for various maneuvers. Free breathing showed a variable amplitude and frequency. Audio prompting resulted in a reproducible frequency; however, the variability and the magnitude of amplitude increased. Visual feedback gave a better control over the amplitude but showed minor variations in frequency. We concluded that training improves the reproducibility of amplitude and frequency of patient breathing cycles. This may increase the accuracy of respiratory-gated radiation therapy.
                Bookmark

                Author and article information

                Contributors
                Koreasye@snu.ac.kr
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                06 September 2013
                September 2013
                : 14
                : 5 ( doiID: 10.1002/acm2.2013.14.issue-5 )
                : 25-42
                Affiliations
                [ 1 ] Department of Radiation Oncology Seoul National University Hospital Seoul
                [ 2 ] Interdisciplinary Program in Radiation Applied Life Science Seoul National University Seoul
                [ 3 ] Department of Mechanical and Aerospace Engineering Seoul National University Seoul
                [ 4 ] Program in Biomedical Radiation Sciences Department of Transdisciplinary Studies Graduate School of Convergence Science and Technology Seoul National University Seoul
                [ 5 ] Advanced Institutes of Convergence Technology Seoul National University Suwon
                Author notes
                [*] [* ]Corresponding author: Sung‐Joon Ye, 101 Daehak‐ro Jongno‐gu, Seoul, Korea, 110–744; phone: (82) (2) 2072 2819; fax: (82) (2) 741 2819; email: sye@ 123456snu.ac.kr
                Article
                ACM20025
                10.1120/jacmp.v14i5.4245
                5714558
                24036857
                02cd7499-c5c4-4276-bda2-234b3ee15eae
                © 2013 The Authors.

                This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 05 October 2012
                : 15 April 2013
                Page count
                Figures: 13, Tables: 6, References: 32, Pages: 18, Words: 6607
                Funding
                Funded by: SNU Brain Fusion Program Research
                Award ID: 400–20100049
                Funded by: National Research Foundation of Korea (NRF)
                Award ID: 800–20120109
                Award ID: 490–20120026
                Categories
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                acm20025
                September 2013
                Converter:WILEY_ML3GV2_TO_NLMPMC version:5.2.5 mode:remove_FC converted:16.11.2017

                gated radiotherapy,external marker tracking,quality assurance,4d ct

                Comments

                Comment on this article