      Assessment of left ventricular ejection fraction in patients eligible for ICD therapy: Discrepancy between cardiac magnetic resonance imaging and 2D echocardiography

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Objective

          Implantable cardioverter defibrillators (ICD) and cardiac resynchronisation therapy (CRT) have substantially improved the survival of patients with cardiomyopathy. Eligibility for this therapy requires a left ventricular ejection fraction (LVEF) <35 %. This is largely based on studies using echocardiography. Cardiac magnetic resonance imaging (CMR) is increasingly utilised for LVEF assessment, but several studies have shown differences between LVEF assessed by CMR and echocardiography. The present study compared LVEF assessment by CMR and echocardiography in a heart failure population and evaluated effects on eligibility for device therapy.

          Methods

          152 patients (106 male, mean age 65.5 ± 9.9 years) referred for device therapy were included. During evaluation of eligibility they underwent both CMR and echocardiographic LVEF assessment. CMR volumes were computed from a stack of short-axis images. Echocardiographic volumes were computed using Simpson’s biplane method.

          Results

          The study population demonstrated an underestimation of end-diastolic volume (EDV) and end-systolic volume (ESV) by echocardiography of 71 ± 53 ml (mean ± SD) and 70 ± 49 ml, respectively. This resulted in an overestimation of LVEF of 6.6 ± 8.3 % by echocardiography compared with CMR (echocardiographic LVEF 31.5 ± 8.7 % and CMR LVEF 24.9 ± 9.6 %). 28 % of patients had opposing outcomes of eligibility for cardiac device therapy depending on the imaging modality used.

          Conclusion

          We found EDV and ESV to be underestimated by echocardiography, and LVEF assessed by CMR to be significantly smaller than by echocardiography. Applying an LVEF cut-off value of 35 %, CMR would significantly increase the number of patients eligible for device implantation. Therefore, LVEF cut-off values might need reassessment when using CMR.

          Related collections

          Most cited references25

          • Record: found
          • Abstract: found
          • Article: not found

          Recommendations for quantitation of the left ventricle by two-dimensional echocardiography. American Society of Echocardiography Committee on Standards, Subcommittee on Quantitation of Two-Dimensional Echocardiograms.

          We have presented recommendations for the optimum acquisition of quantitative two-dimensional data in the current echocardiographic environment. It is likely that advances in imaging may enhance or supplement these approaches. For example, three-dimensional reconstruction methods may greatly augment the accuracy of volume determination if they become more efficient. The development of three-dimensional methods will depend in turn on vastly improved transthoracic resolution similar to that now obtainable by transesophageal echocardiography. Better resolution will also make the use of more direct methods of measuring myocardial mass practical. For example, if the epicardium were well resolved in the long-axis apical views, the myocardial shell volume could be measured directly by the biplane method of discs rather than extrapolating myocardial thickness from a single short-axis view. At present, it is our opinion that current technology justifies the clinical use of the quantitative two-dimensional methods described in this article. When technically feasible, and if resources permit, we recommend the routine reporting of left ventricular ejection fraction, diastolic volume, mass, and wall motion score.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Comparison of left ventricular ejection fraction and volumes in heart failure by echocardiography, radionuclide ventriculography and cardiovascular magnetic resonance; are they interchangeable?

            To prospectively compare the agreement of left ventricular volumes and ejection fraction by M-mode echocardiography (echo), 2D echo, radionuclide ventriculography and cardiovascular magnetic resonance performed in patients with chronic stable heart failure. It is important to know whether the results of each technique are interchangable, and thereby how the results of large studies in heart failure utilizing one technique can be applied using another. Some studies have compared cardiovascular magnetic resonance with echo or radionuclude ventriculography but few contain patients with heart failure and none have compared these techniques with the current fast breath-hold acquisition cardiovascular magnetic resonance. Fifty two patients with chronic stable heart failure taking part in the CHRISTMAS Study, underwent M-mode echo, 2D echo, radionuclude ventriculography and cardiovascular magnetic resonance within 4 weeks. The scans were analysed independently in blinded fashion by a single investigator at three core laboratories. Of the echocardiograms, 86% had sufficient image quality to obtain left ventricular ejection fraction by M-mode method, but only 69% by 2D Simpson's biplane analysis. All 52 patients tolerated the radionuclude ventriculography and cardiovascular magnetic resonance, and all these scans were analysable. The mean left ventricular ejection fraction by M-mode cube method was 39+/-16% and 29+/-15% by Teichholz M-mode method. The mean left ventricular ejection fraction by 2D echo Simpson's biplane was 31+/-10%, by radionuclude ventriculography was 24+/-9% and by cardiovascular magnetic resonance was 30+/-11. All the mean left ventricular ejection fractions by each technique were significantly different from all other techniques (P<0.001), except for cardiovascular magnetic resonance ejection fraction and 2D echo ejection fraction by Simpson's rule (P=0.23). The Bland-Altman limits of agreement encompassing four standard deviations was widest for both cardiovascular magnetic resonance vs cube M-mode echo and cardiovascular magnetic resonance vs Teichholz M-mode echo at 66% each, and was 58% for radionuclude ventriculography vs cube M-mode echo, 44% for cardiovascular magnetic resonance vs Simpson's 2D echo, 39% for radionuclide ventriculography vs Simpson's 2D echo, and smallest at 31% for cardiovascular magnetic resonance-radionuclide ventriculography. Similarly, the end-diastolic volume and end-systolic volume by 2D echo and cardiovascular magnetic resonance revealed wide limits of agreement (52 ml to 216 ml and 11 ml to 188 ml, respectively). These results suggest that ejection fraction measurements by various techniques are not interchangeable. The conclusions and recommendations of research studies in heart failure should therefore be interpreted in the context of locally available techniques. In addition, there are very wide variances in volumes and ejection fraction between techniques, which are most marked in comparisons using echocardiography. This suggests that cardiovascular magnetic resonance is the preferred technique for volume and ejection fraction estimation in heart failure patients, because of its 3D approach for non-symmetric ventricles and superior image quality. Copyright 2000 The European Society of Cardiology.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Combined cardiac resynchronization and implantable cardioversion defibrillation in advanced chronic heart failure: the MIRACLE ICD Trial.

              Cardiac resynchronization therapy (CRT) through biventricular pacing is an effective treatment for heart failure (HF) with a wide QRS; however, the outcomes of patients requiring CRT and implantable cardioverter defibrillator (ICD) therapy are unknown. To examine the efficacy and safety of combined CRT and ICD therapy in patients with New York Heart Association (NYHA) class III or IV congestive HF despite appropriate medical management. Randomized, double-blind, parallel-controlled trial conducted from October 1, 1999, to August 31, 2001, of 369 patients with left ventricular ejection fraction of 35% or less, QRS duration of 130 ms, at high risk of life-threatening ventricular arrhythmias, and in NYHA class III (n = 328) or IV (n = 41) despite optimized medical treatment. Of 369 randomized patients who received devices with combined CRT and ICD capabilities, 182 were controls (ICD activated, CRT off) and 187 were in the CRT group (ICD activated, CRT on). The primary double-blind study end points were changes between baseline and 6 months in quality of life, functional class, and distance covered during a 6-minute walk. Additional outcome measures included changes in exercise capacity, plasma neurohormones, left ventricular function, and overall HF status. Survival, incidence of ventricular arrhythmias, and rates of hospitalization were also compared. At 6 months, patients assigned to CRT had a greater improvement in median (95% confidence interval) quality of life score (-17.5 [-21 to -14] vs -11.0 [-16 to -7], P =.02) and functional class (-1 [-1 to -1] vs 0 [-1 to 0], P =.007) than controls but were no different in the change in distance walked in 6 minutes (55 m [44-79] vs 53 m [43-75], P =.36). Peak oxygen consumption increased by 1.1 mL/kg per minute (0.7-1.6) in the CRT group vs 0.1 mL/kg per minute (-0.1 to 0.8) in controls (P =.04), although treadmill exercise duration increased by 56 seconds (30-79) in the CRT group and decreased by 11 seconds (-55 to 12) in controls (P<.001). No significant differences were observed in changes in left ventricular size or function, overall HF status, survival, and rates of hospitalization. No proarrhythmia was observed and arrhythmia termination capabilities were not impaired. Cardiac resynchronization improved quality of life, functional status, and exercise capacity in patients with moderate to severe HF, a wide QRS interval, and life-threatening arrhythmias. These improvements occurred in the context of underlying appropriate medical management without proarrhythmia or compromised ICD function.
                Bookmark

                Author and article information

                Contributors
                +31 20 444 3847 , s.dehaan@vumc.nl
                Journal
                Neth Heart J
                Neth Heart J
                Netherlands Heart Journal
                Bohn Stafleu van Loghum (Houten )
                1568-5888
                1876-6250
                4 September 2014
                4 September 2014
                October 2014
                : 22
                : 10
                : 449-455
                Affiliations
                Department of Cardiology and Institute for Cardiovascular Research (ICaR-VU), VU University Medical Center, De Boelelaan 1118, 1081 HV Amsterdam, the Netherlands
                Article
                594
                10.1007/s12471-014-0594-0
                4188847
                25187012
                05bb348b-5ac7-4730-add3-a2599b53d883
                © The Author(s) 2014

                Open Access This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.

                History
                Categories
                Original Article-E-Learning
                Custom metadata
                © The Author(s) 2014

                Cardiovascular Medicine
                cardiac magnetic resonance imaging,echocardiography,left ventricular ejection fraction

                Comments

                Comment on this article