      Effects of intraocular lenses with different diopters on chromatic aberrations in human eye models

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          In this study, the effects of intraocular lenses (IOLs) with different diopters (D) on chromatic aberration were investigated in human eye models, and the influences of the central thickness of IOLs on chromatic aberration were compared.

          Methods

          A Liou-Brennan-based IOL eye model was constructed using ZEMAX optical design software. Spherical IOLs with different diopters (AR40e, AMO Company, USA) were implanted; modulation transfer function (MTF) values at 3 mm of pupil diameter and from 0 to out-of-focus blur were collected and graphed.

          Results

          MTF values, measured at 555 nm of monochromatic light under each spatial frequency, were significantly higher than the values measured at 470 to 650 nm of polychromatic light. The influences of chromatic aberration on MTF values decreased with the increase in IOL diopter when the spatial frequency was ≤12 c/d, while increased effects were observed when the spatial frequency was ≥15 c/d. The MTF values of each IOL eye model were significantly lower than the MTF values of the Liou-Brennan eye models when measured at 555 nm of monochromatic light and at 470 to 650 nm of polychromatic light. The MTF values were also found to be increased with the increase in IOL diopter.

          Conclusion

          With higher diopters of IOLs, the central thickness increased accordingly, which could have created increased chromatic aberration and decreased the retinal image quality. To improve the postoperative visual quality, IOLs with lower chromatic aberration should be selected for patients with short axial lengths.

          Related collections

          Most cited references29

          • Record: found
          • Abstract: not found
          • Article: not found

          Aberrations and retinal image quality of the normal human eye

            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Aberrations and retinal image quality of the normal human eye.

            We have constructed a wave-front sensor to measure the irregular as well as the classical aberrations of the eye, providing a more complete description of the eye's aberrations than has previously been possible. We show that the wave-front sensor provides repeatable and accurate measurements of the eye's wave aberration. The modulation transfer function of the eye computed from the wave-front sensor is in fair, though not complete, agreement with that obtained under similar conditions on the same observers by use of the double-pass and the interferometric techniques. Irregular aberrations, i.e., those beyond defocus, astigmatism, coma, and spherical aberration, do not have a large effect on retinal image quality in normal eyes when the pupil is small (3 mm). However, they play a substantial role when the pupil is large (7.3-mm), reducing visual performance and the resolution of images of the living retina. Although the pattern of aberrations varies from subject to subject, aberrations, including irregular ones, are correlated in left and right eyes of the same subject, indicating that they are not random defects.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Anatomically accurate, finite model eye for optical modeling.

              There is a need for a schematic eye that models vision accurately under various conditions such as refractive surgical procedures, contact lens and spectacle wear, and near vision. Here we propose a new model eye close to anatomical, biometric, and optical realities. This is a finite model with four aspheric refracting surfaces and a gradient-index lens. It has an equivalent power of 60.35 D and an axial length of 23.95 mm. The new model eye provides spherical aberration values within the limits of empirical results and predicts chromatic aberration for wavelengths between 380 and 750 nm. It provides a model for calculating optical transfer functions and predicting optical performance of the eye.
                Bookmark

                Author and article information

                Contributors
                songh221@hotmail.com
                yuanxy_cn@hotmail.com
                86 22 2731 3336 , tangprofessor@aliyun.com
                Journal
                BMC Ophthalmol
                BMC Ophthalmol
                BMC Ophthalmology
                BioMed Central (London )
                1471-2415
                11 January 2016
                11 January 2016
                2016
                : 16
                : 9
                Affiliations
                Tianjin Eye Hospital, Tianjin Key Laboratory of Ophthalmology and Vision Science, Clinical College of Ophthalmology, Tianjin Medical University, No. 4 Gansu Rd, Heping District, Tianjin, 300020 China
                Article
                184
                10.1186/s12886-016-0184-6
                4707777
                26754111
                05d04f27-04b9-4ffe-860d-02c8acb38259
                © Song et al. 2016

                Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License ( http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

                History
                : 30 August 2014
                : 6 January 2016
                Categories
                Research Article
                Custom metadata
                © The Author(s) 2016

                Ophthalmology & Optometry
                intraocular lens,computer simulation,diopter,chromatic aberration,modulation transfer function

                Comments

                Comment on this article