      Assessment of a wrist-worn device in the detection of obstructive sleep apnea.

      Sleep Medicine
      Adult, Aged, Algorithms, Case-Control Studies, Equipment Design, Equipment and Supplies, Female, Humans, Male, Middle Aged, Polysomnography, ROC Curve, Sensitivity and Specificity, Sleep Apnea, Obstructive, diagnosis, Wrist

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          To assess the accuracy of a wrist-worn device (Watch_PAT100) to diagnose obstructive sleep apnea (OSA). Thirty adult subjects with and without suspected OSA simultaneously had a standard in-laboratory polysomnogram (PSG) and wore the Watch_PAT100 during a full-night recording. PSG sleep and respiratory events were scored according to standard criteria. Watch_PAT data were analyzed with an automated computerized algorithm which calculated the frequency of respiratory events per hour of actigraphy measured sleep using a combination of peripheral arterial tonometry (PAT) signal attenuation, desaturation on pulse oximetry, and changes in heart rate. This yielded a PAT apnea hypopnea index (AHI). Mean age was 47.0+/-14.8 years, mean body mass index 31.0+/-7.6 kg/m(2), mean PSG AHI 23+/-23.9 events per hour, and mean PAT AHI 23+/-15.9 events per hour. There was a significant correlation between PAT AHI and AHI by PSG (r=0.87, P<0.001). To assess sensitivity and specificity of Watch_PAT, we constructed receiver operator characteristic curves using a variety of AHI threshold values (10, 15, 20, and 30 events per hour). Optimal combinations of sensitivity and specificity for the various thresholds were 82.6/71.4, 93.3/73.3, 90.9/84.2, and 83.3/91.7, respectively. The Watch_PAT is a device that can detect OSA with reasonable accuracy. Thus, the Watch_PAT may be a useful method to diagnose OSA.

          Related collections

          Author and article information

          Comments

          Comment on this article