      Evaluation of Three Multiplex Real-time Reverse Transcription PCR Assays for Simultaneous Detection of SARS-CoV-2, Influenza A/B, and Respiratory Syncytial Virus in Nasopharyngeal Swabs

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          In the coronavirus disease 2019 (COVID-19) pandemic era, the simultaneous detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza virus (Flu), and respiratory syncytial virus (RSV) is important in the rapid differential diagnosis in patients with respiratory symptoms. Three multiplex real-time reverse transcription polymerase chain reaction (rRT-PCR) assays have been recently developed commercially in Korea: PowerChek™ SARS-CoV-2, Influenza A&B Multiplex Real-time PCR Kit (PowerChek; KogeneBiotech); STANDARD™ M Flu/SARS-CoV-2 Real-time Detection Kit (STANDARD M; SD BioSensor); and Allplex™ SARS-CoV-2/FluA/FluB/RSV Assay (Allplex; Seegene). We evaluated the analytical and clinical performances of these kits.

          Methods

          A limit of detection tests were performed and cross-reactivity analysis was executed using clinical respiratory samples. Ninety-seven SARS-CoV-2-positive, 201 SARS-CoV-2-negative, 71 influenza A-positive, 50 influenza B-positive, 78 RSV-positive, and 207 other respiratory virus-positive nasopharyngeal swabs were tested using the three assays. The AdvanSure™ respiratory viruses rRT-PCR assay (AdvanSure; LG Life Sciences) was used as a comparator assay for RSV.

          Results

          Except in influenza B, in SARS-CoV-2 and influenza A, there were no significant differences in detecting specific genes of the viruses among the three assays. All three kits did not cross-react with common respiratory viruses. All three kits had greater than 92% positive percent agreement and negative percent agreement and ≥ 0.95 kappa value in the detection of SARS-CoV-2 and flu A/B. Allplex detected RSV more sensitively than AdvanSure.

          Conclusion

          The overall performance of three multiplex rRT-PCR assays for the concurrent detection of SARS-CoV-2, influenza A/B, and RSV was comparable. These kits will promote prompt differential diagnosis of COVID-19, influenza, and RSV infection in the COVID-19 pandemic era.

          Graphical Abstract

          Related collections

          Most cited references20

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          A new coronavirus associated with human respiratory disease in China

          Emerging infectious diseases, such as severe acute respiratory syndrome (SARS) and Zika virus disease, present a major threat to public health 1–3 . Despite intense research efforts, how, when and where new diseases appear are still a source of considerable uncertainty. A severe respiratory disease was recently reported in Wuhan, Hubei province, China. As of 25 January 2020, at least 1,975 cases had been reported since the first patient was hospitalized on 12 December 2019. Epidemiological investigations have suggested that the outbreak was associated with a seafood market in Wuhan. Here we study a single patient who was a worker at the market and who was admitted to the Central Hospital of Wuhan on 26 December 2019 while experiencing a severe respiratory syndrome that included fever, dizziness and a cough. Metagenomic RNA sequencing 4 of a sample of bronchoalveolar lavage fluid from the patient identified a new RNA virus strain from the family Coronaviridae, which is designated here ‘WH-Human 1’ coronavirus (and has also been referred to as ‘2019-nCoV’). Phylogenetic analysis of the complete viral genome (29,903 nucleotides) revealed that the virus was most closely related (89.1% nucleotide similarity) to a group of SARS-like coronaviruses (genus Betacoronavirus, subgenus Sarbecovirus) that had previously been found in bats in China 5 . This outbreak highlights the ongoing ability of viral spill-over from animals to cause severe disease in humans.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Real-time RT-PCR in COVID-19 detection: issues affecting the results

            Due to the rapid spread and increasing number of coronavirus disease 19 (COVID-19) cases caused by a new coronavirus, severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), rapid and accurate detection of virus and/or disease is increasingly vital to control the sources of infection and help patients to prevent the illness progression. Since December 2019, there has been considerable challenge regarding the use of nucleic acid test or clinical characteristics of infected patients as the reference standard to make a definitive diagnose of COVID-19 patients. As the early diagnosis of COVID-19 is critical for prevention and control of this pandemic, clinical characteristics cannot alone define the diagnosis of COVID-19, especially for patients presenting early-onset of symptoms. Along with the advancement in medical diagnosis, nucleic acid detection-based approaches have become a rapid and reliable technology for viral detection. Among nucleic acid tests, the polymerase chain reaction (PCR) method is considered as the ‘gold standard’ for the detection of some viruses and is characterized by rapid detection, high sensitivity, and specificity. As such, real-time reverse transcriptase-PCR (RT-PCR) is of great interest today for the detection of SARS-CoV-2 due to its benefits as a specific and simple qualitative assay [1–3]. Moreover, real-time RT-PCR has adequate sensitivity to help us much for diagnosing early infection. Therefore, the ‘criterion-referenced’ real-time RT-PCR assay can be considered as a main method to be applied to detect the causative agent of COVID-19, SARS-CoV-2. An important issue with the real-time RT-PCR test is the risk of eliciting false-negative and false-positive results. It is reported that many ‘suspected’ cases with typical clinical characteristics of COVID-19 and identical specific computed tomography (CT) images were not diagnosed [4]. Thus, a negative result does not exclude the possibility of COVID-19 infection and should not be used as the only criterion for treatment or patient management decisions. It seems that combination of real-time RT-PCR and clinical features facilitates management of SARS-CoV-2 outbreak. Several factors have been proposed to be associated with the inconsistency of real-time RT-PCR [5]. In the following, we attempt to discuss various challenges regarding the detection of SARS-CoV-2 by real-time RT-PCR. It is expected that this could provide beneficial information for the comprehension of the limitations of the obtained results and to improve diagnosis approaches and control of the disease. It is well known that results from real-time RT-PCR using primers in different genes can be affected by the variation of viral RNA sequences. Genetic diversity and rapid evolution of this novel coronavirus have been observed in different studies [6,7]. False-negative results may occur by mutations in the primer and probe target regions in the SARS-CoV-2 genome. Although it was attempted to design the real-time RT-PCR assay as precisely as possible based on the conserved regions of the viral genomes, variability causing mismatches between the primers and probes and the target sequences can lead to decrease in assay performance and potential false-negative results. In this regard, multiple target gene amplification could be used to avoid invalid results. Several types of SARS-CoV-2 real-time RT-PCR kit have been developed and approved rapidly, but with different quality. Importantly, the sensitivity and specificity of the real-time RT-PCR test is not 100%. All of them behind the laboratory practice standard and personnel skill in the relevant technical and safety procedures explain some of the false-negative results. According to the natural history of the COVID-19 and viral load kinetics in different anatomic sites of the patients, sampling procedures largely contribute to the false-negative results. Optimum sample types and timing for peak viral load during infections caused by SARS-CoV-2 remain to be fully determined. A study has reported sputum as the most accurate sample for laboratory diagnosis of COVID-19, followed by nasal swabs, while throat swabs were not recommended for the diagnosis [8]. They also suggested the detection of viral RNAs in bronchoalveolar lavage fluid (BALF) for the diagnosis and monitoring of viruses in severe cases. However, gathering of BALF needs both a suction tool and an expert operator, in addition to being painful to the patients. While BALF samples are not practical for the routine laboratory diagnosis and monitoring of the disease, collection of other samples such as sputum, nasal swab, and throat swab is rapid, simple, and safe. To avoid inconsistent results, it would be better to use different specimen types (stool and blood) besides respiratory specimen during different stages. It is worth noting that samples should be obtained by dacron or polyester flocked swabs and should reach the laboratory as soon as possible after collection. False-negative results may occur due to the presence of amplification inhibitors in the sample or insufficient organisms in the sample rising from inappropriate collection, transportation, or handling. Viral load kinetics of SARS-CoV-2 infection have been described in two patients in Korea, suggesting a different viral load kinetics from that of previously reported other coronavirus infections [9]. In the first patient, the virus was detected from upper respiratory tract (URT) and lower respiratory tract (LRT) specimens on days 2 and 3 of symptom onset, respectively. On day 5, the viral load was increased from day 3 in the LRT specimen. However, the viral loads decreased from around day 7 in both URT and LRT specimens. Real-time RT-PCR continued to be positive at a low level until day 13 (LRT specimens) and day 14 (URT specimens). Finally, the assay became undetectable for two consecutive days from day 14 (LRT specimen) and day 15 (URT specimen), respectively. In the second patient, SARS-CoV-2 was detected in both URT and LRT specimens on day 14 of symptom onset. However, the initial viral loads were relatively lower than those of patient 1 in whom the test was performed on day 2 of symptom onset. From day 18 (URT specimen) and day 20 (LRT specimen), real-time RT-PCR became undetectable for two consecutive days, respectively. URT sample of day 25 was again positive for RdRp and E genes. However, it was interpreted as negative due to high Ct value of the RdRp gene (Ct value of 36.69). These findings indicate the different viral load kinetics of SARS-coV-2 in different patients, suggesting that sampling timing and period of the disease development play an important role in real-time RT-PCR results. Finally, the Centers for Disease Control and Prevention (CDC) has designed a SARS-CoV-2 Real-Time RT-PCR Diagnostic Panel to minimize the chance of false-positive results [10]. In accordance, the negative template control (NTC) sample should be negative, showing no fluorescence growth curves that cross the threshold line. The occurrence of false positive with one or more of the primer and probe NTC reactions is indicative of sample contamination. Importantly, the internal control should be included to help identify the specimens containing substances that may interfere with the extraction of nucleic acid and PCR amplification. Because of the several risks to patients in the event of a false-positive result, all clinical laboratories using this test must follow the standard confirmatory testing and reporting guidelines based on their proper public health authorities. 1. Expert opinion In conclusion, according to the mentioned reasons, the results of real-time RT-PCR tests must be cautiously interpreted. In the case of real-time RT-PCR negative result with clinical features suspicion for COVID-19, especially when only upper respiratory tract samples were tested, multiple sample types in different time points, including from the lower respiratory tract if possible, should be tested. Importantly, combination of real-time RT-PCR and clinical features especially CT image could facilitate disease management. Proper sampling procedures, good laboratory practice standard, and using high-quality extraction and real-time RT-PCR kit could improve the approach and reduce inaccurate results.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: found

              Rates of Co-infection Between SARS-CoV-2 and Other Respiratory Pathogens

              This study describes the prevalence of SARS-CoV-2 co-infection with noncoronavirus respiratory pathogens in a sample of symptomatic patients undergoing PCR testing in March 2020.
                Bookmark

                Author and article information

                Journal
                J Korean Med Sci
                J Korean Med Sci
                JKMS
                Journal of Korean Medical Science
                The Korean Academy of Medical Sciences
                1011-8934
                1598-6357
                18 November 2021
                13 December 2021
                : 36
                : 48
                : e328
                Affiliations
                [1 ]Department of Laboratory Medicine, Seoul National University Hospital, Seoul, Korea.
                [2 ]Department of Laboratory Medicine, Seoul National University Boramae Medical Center, Seoul, Korea.
                [3 ]Department of Laboratory Medicine, Seoul National University College of Medicine, Seoul, Korea.
                Author notes
                Address for Correspondence: Hyunwoong Park, MD, PhD. Department of Laboratory Medicine, Seoul National University Boramae Medical Center, 20 Boramae-ro 5-gil, Dongjak-gu, Seoul 07061, Korea. bewithu2@ 123456snu.ac.kr
                Author information
                https://orcid.org/0000-0002-6306-3944
                https://orcid.org/0000-0003-0261-2185
                https://orcid.org/0000-0002-9851-1727
                https://orcid.org/0000-0002-1787-7509
                https://orcid.org/0000-0003-4791-8671
                https://orcid.org/0000-0002-1252-4739
                https://orcid.org/0000-0002-2093-1721
                https://orcid.org/0000-0001-7765-2259
                Article
                10.3346/jkms.2021.36.e328
                8668494
                34904407
                0bf5b470-7ecf-481f-90b9-9f026a8ff793
                © 2021 The Korean Academy of Medical Sciences.

                This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License ( https://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 16 July 2021
                : 10 November 2021
                Funding
                Funded by: Seegene
                Categories
                Original Article
                Infectious Diseases, Microbiology & Parasitology

                Medicine
                sars-cov-2,influenza a virus,influenza b virus,respiratory syncytial viruses,multiplex polymerase chain reaction

                Comments

                Comment on this article