      Accuracy and repeatability of anthropometric facial measurements using cone beam computed tomography.

      The Cleft palate-craniofacial journal : official publication of the American Cleft Palate-Craniofacial Association
      Anthropometry, instrumentation, Cadaver, Cone-Beam Computed Tomography, methods, Face, anatomy & histology, radiography, Humans, Imaging, Three-Dimensional, Radiographic Image Interpretation, Computer-Assisted, Reproducibility of Results, Software

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The purpose of this study was to determine the accuracy and repeatability of linear anthropometric measurements on the soft tissue surface model generated from cone beam computed tomography scans. The study sample consisted of seven cadaver heads. The accuracy and repeatability were assessed by means of a series of 21 standardized, linear facial measurements derived from 11 landmarks taken both directly on the face with a set of digital calipers and indirectly from a three-dimensional soft tissue surface model generated from a cone beam computed tomography scan of the heads using SimPlant® Ortho Pro software. The landmarks and measurements were chosen to cover various regions of the face with an emphasis on the oral-nasal region. The cone beam computed tomography measurements were compared with the physical measurements. Statistical analysis for the repeatability was done by means of the intraclass coefficient. Accuracy was determined by means of the absolute error and absolute percentage error. The cone beam computed tomography measurements were very accurate when compared with the physical measurements (0.962 to 0.999). Except for one measurement, between point tragion (t) and nasion (n) (mean, 1.52 mm), all the measurements had a mean absolute error of less than 1.5 mm. The three-dimensional surface models derived from cone beam computed tomography images are sufficiently precise and accurate for the anthropometric measurements.

          Related collections

          Author and article information

          Comments

          Comment on this article