15
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Free-breathing multi-slice myocardial T 2 mapping using slice-selective T 2 magnetization preparation

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background Quantitative myocardial T2 mapping allows non-invasive assessment of myocardial inflammation/edema [1]. Current implementations commonly use a T2-prepared (T2prep) SSFP sequence to acquire different T2 weighted images at different echo times to generate the T2 maps [2, 3]. However, all current techniques are designed for single slice acquisition with long rest cycles (3-6 sec) after each T2prep image acquisition to allow for full spin recovery. This markedly increases the overall scan time, especially if multiple slices are to be acquired in serial. In this study, we propose a novel multi-slice T2 mapping sequence, which uses slice-selective T2prep pulses combined with an interleaved slice acquisition scheme to provide a fast multi-slice T2 mapping. Methods Fig. 1 shows a schematic for the proposed sequence with the proposed slice selective T2prep pulses and the interleaved slice acquisition. Upon the acquisition of a specific slice, the other slices are selectively prepared, excited and acquired during the relaxation period of that slice. Thus, one T2prep image is acquired at every heartbeat. Prospective slice tracking and retrospective image registration were used to correct for respiratory motion. Phantom imaging was performed using NiCl2 doped agarose vials, whose T2/T1 values spanned the ranges of values found in the blood and myocardium. Ten healthy adults subjects (29±17 y, 4m) were imaged on a 1.5T Phillips scanner. A free-breathing single-shot ECG-triggered slice-selective T2prep bSSFP sequence with the following parameters was used for acquisition of five mid-ventricular short-axis slices, FOV=320×320 mm2, in-plane resolution=2.5×2.5mm2, slice thickness=8mm, slice gap=4mm, TR/TE=2.2/1.1ms, α=40°, SENSE rate=2, acquisition window=140 ms. For comparison, a conventional breath-hold single-slice T2prep bSSFP sequence was performed to image the middle of the 5-slices. All acquisitions were performed using the conventional 3-images with T2prep echo times = 0,25,50 ms (2), with a SAT image added to compensate for the T1 relaxation time during readout [4]. T2 maps were then generated using the 3-parameter fitting model [4]. Figure 1 a) Schematic diagram of the proposed multi-slice T2 mapping sequence. Multiple single-shot images are acquired using ECG-triggering, following T2prep of different echo lengths, TET2P. An image, ISAT, is acquired directly after a saturation pulse to simulate the effect of a very long T2prep echo time b) The slice selective T2prep pulse consists of a tip-down slice selective 90˚ pulse, followed by four non-selective 180˚ refocus pulses (only two are illustrated in the figure) and ends with a closing tip-up slice selective 90˚ pulse. c) Slices are acquired in an interleaved fashion to allow sufficient T1 recovery time for each slice. The slice selective 90˚ pulses are applied with a slice thickness twice as imaging slice to minimize the impact of slice imperfection. Results Fig. 2a shows the correlation between T2 measurements using the single and multi-slice sequences in phantom compared to spin echo. Figure 2b. shows an example T2 maps. Fig. 2c. shows a comparison between T2 maps generated using the single slice and multi slice sequences. The average scan time was 20 heartbeats for the 5 slices using the multi-slice and 13 heartbeats per slice using the single slice sequence. The average T2 across the myocardium and over all healthy subjects was 51ms and 48ms using the single and the multi-slice sequence respectively (p=0.1). Figure 2 a) Correlation between T2 measurements measured in different phantoms using the spin echo sequence and single and multi-slice T2 mapping b) A representative example for the multi-slice T2 maps in a healthy subject. c) Two example comparisons for T2 maps generated using the middle slice in the multi-slice acquisition and the corresponding single slice acquisition. The bullseye shows the mean ± standard deviation of T2 in each segment .The center values shows the mean T2 for all 5 slices. Conclusions The proposed multi-slice T2 mapping pulse sequence allows myocardial T2 measurements over the entire left ventricle by imaging of 5 interleaved slices in 20 heartbeats. Funding N/A.

          Related collections

          Most cited references4

          • Record: found
          • Abstract: not found
          • Article: not found

          MRM

            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            JACC

              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              MRM

                Bookmark

                Author and article information

                Journal
                J Cardiovasc Magn Reson
                J Cardiovasc Magn Reson
                Journal of Cardiovascular Magnetic Resonance
                BioMed Central (London )
                1097-6647
                1532-429X
                3 February 2015
                2015
                : 17
                : 1
                : W3
                Affiliations
                [ ]Cardiology, BIDMC, Boston, MA USA
                [ ]Harvard Medical School, Boston, MA USA
                Article
                4555
                10.1186/1532-429X-17-S1-W3
                4328396
                1162ac68-4a74-4cf5-938f-e60e7a685fc6
                © Basha et al; licensee BioMed Central Ltd. 2015

                This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

                History
                Categories
                Workshop Presentation
                Custom metadata
                © The Author(s) 2015

                Cardiovascular Medicine
                Cardiovascular Medicine

                Comments

                Comment on this article