      Comparative analysis of hepatopulmonary shunt obtained from pretherapy 99mTc MAA scintigraphy and post-therapy 90Y Bremsstrahlung imaging in 90Y microsphere therapy.

      Nuclear Medicine Communications
      Adult, Aged, Embolization, Therapeutic, methods, Humans, Image Processing, Computer-Assisted, Liver Neoplasms, radionuclide imaging, radiotherapy, Lung, radiation effects, Microspheres, Middle Aged, Particle Size, Prognosis, Radiopharmaceuticals, diagnostic use, Reproducibility of Results, Technetium Tc 99m Aggregated Albumin, Yttrium Radioisotopes, therapeutic use

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          (99m)Tc macroaggregate albumin (MAA) scintigraphy is routinely used to estimate the hepatopulmonary shunt (HPS) of (90)Y microspheres because of their comparable average particle sizes (20-30 µm). However, the MAA particle size can vary from 10 to 90 µm. Therefore, HPS computed from (99m)Tc MAA scintigraphy may not accurately represent the HPS of (90)Y microspheres. In view of this, the present study was undertaken to investigate the accuracy of (99m)Tc MAA scintigraphy in estimating the HPS of (90)Y microspheres. Nineteen sessions of transarterial radioembolization using (90)Y therasphere were carried out in 17 patients for hepatic malignancies (both primary and secondary). For each session of therapy, a pretherapeutic (99m)Tc MAA scintigraphy and post-therapeutic (90)Y Bremsstrahlung scintigraphy were performed. The HPSs obtained from these images were compared. The mean HPS fractions calculated from the pretherapeutic (99m)Tc MAA study and the post-therapeutic (90)Y Bremsstrahlung images were 4.77 ± 2.81 and 4.52 ± 2.5%, respectively. The coefficient of correlation (r) was 0.96. (99m)Tc MAA scintigraphy accurately predicts the HPS of (90)Y microspheres.

          Related collections

          Author and article information

          Comments

          Comment on this article