      Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose:

          A known factor affecting the accuracy of radiochromic film dosimetry is the lateral response artifact (LRA) induced by nonuniform response of a flatbed scanner in the direction perpendicular to the scan direction. This work reports a practical solution to eliminate such artifacts for all forms of dose QA.

          Methods:

          EBT3 films from a single production lot (02181401) cut into rectangular 4 × 5 cm 2 pieces, with the long dimension parallel to the long dimension of the original 20.3 × 25.4 cm 2 sheets, were exposed at a depth of 5 cm on a Varian Trilogy at the center of a 20 × 20 cm 2 open field at seven doses between 50 and 1600 cGy using 6 MV photons. These films together with an unexposed film from the same production lot were lined one next to the other on an Epson 10000XL or 11000XL scanner in portrait orientation with their long dimension parallel to the scan direction. Scanned images were then obtained with the line of films positioned at seven discrete lateral locations perpendicular to the scan direction. The process was repeated in landscape orientation and on three other Epson scanners. Data were also collected for three additional production lots of EBT3 film (11051302, 03031401, and 03171403). From measurements at the various lateral positions, the scanner response was determined as a function of the lateral position of the scanned film. For a given color channel X, the response at any lateral position L is related to the response at the center, C, of the scanner by Response( C, D, X) = A L, X + B L, X ⋅Response( L, D, X), where D is dose and the coefficients A L, X and B L, X are determined from the film measurements at the center of the scanner and six other discrete lateral positions. The values at intermediate lateral positions were obtained by linear interpolation. The coefficients were determined for the red, green, and blue color channels, preserving the ability to apply triple-channel dosimetry once corrections were applied to compensate for the lateral position response artifact. To validate this method, corrections were applied to several films that were exposed to 15 × 15cm 2 open fields and large IMRT and VMAT fields and scanned at the extreme edges of the scan window in addition to the central location. Calibration and response data were used to generate dose maps and perform gamma analysis using single- or triple-channel dosimetry with FilmQAPro 2014 software.

          Results:

          The authors’ study found that calibration curves at the different lateral positions could be correlated by a simple two-point rescaling using the response for unexposed film as well as the response of film exposed at high doses between 800 and 1600 cGy. The coefficients A L, X and B L, X for each color channel X were found to be independent of dose at each lateral location L. This made it possible to apply the relationship Response( C, D, X) = A L, X + B L, X ⋅Response( L, D, X), to the raw film responses, permitting correction of the response values at any lateral position to an equivalent response, as if that part of the film was located at the center of the scanner. This correction method was validated for several films exposed to open as well as large IMRT and VMAT fields.

          Conclusions:

          The work reported elaborates on the process using the correction procedures to eliminate the lateral response artifact and demonstrates improvements in the accuracy of radiochromic film dosimetry for the radiation therapy quality assurance applications.

          Related collections

          Author and article information

          Journal
          Med Phys
          Med Phys
          MPHYA6
          Medical Physics
          American Association of Physicists in Medicine
          0094-2405
          January 2015
          30 December 2014
          30 December 2014
          : 42
          : 1
          : 416-429
          Affiliations
          RCF Consulting , LLC, 54 Benedict Road, Monroe, Connecticut 06468
          Memorial Sloan-Kettering Cancer Center , 136 Mountain View Boulevard, Basking Ridge, New Jersey 07920
          Author notes
          [a)]

          Author to whom correspondence should be addressed. Electronic mail: chanm@ 123456mskcc.org .; Telephone: (908) 542-3046.

          Article
          PMC5148133 PMC5148133 5148133 032501MPH 1.4903758 14-1007R1
          10.1118/1.4903758
          5148133
          25563282
          153545e4-f7ec-4a23-86cc-b4d11f4f6b04
          © 2015 Author(s).

          0094-2405/2015/42(1)/416/14/ $0.00

          All article content, except where otherwise noted, is licensed under a Creative Commons Attribution 3.0 Unported License.

          History
          : 05 June 2014
          : 04 November 2014
          : 23 November 2014
          Page count
          Pages: 14
          Funding
          Funded by: Ashland Specialty Ingredients
          Categories
          Radiation Therapy Physics
          Custom metadata

          lateral response,radiochromic film dosimetry,QA,IMRT,radiotherapy

          Comments

          Comment on this article