      Usability of a new disposable autoinjector platform device: results of a formative study conducted with a broad user population

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          This article presents a late-stage formative usability study of an autoinjector platform device. Such devices are used for the subcutaneous delivery of biopharmaceuticals, primarily for self-administration by the patient. Previous usability work on autoinjectors reported in the literature has been specific to single indications. This study was instead conducted with a broad user population, defined to represent user characteristics across a range of indications. The goals of the study were to evaluate whether users could use the devices safely and effectively, and could understand the instructions for use (IFU) as well as the accompanying training. Further objectives were to capture any usability issues and to obtain participants’ subjective ratings on the IFU and training as well as the confidence and comfort in using the device. A total of 43 participants in 5 groups received training and performed simulated injections either into an injection pad or a mannequin. All participants were able to successfully use the device. The device was well appreciated by all users, with a reported degree of confidence in using the device of 98%, of user comfort of 93%, and of comprehensibility of IFU and training of 98%. These values are higher than other comparable results reported in the literature. The presence of both audible and visible feedback during injection was seen to be a significant factor contributing to injection success. The observation that the device can be safely and efficiently used by all tested user groups provides confidence that the device and IFU in their current form will pass future summative testing in specific applications.

          Related collections

          Most cited references34

          • Record: found
          • Abstract: found
          • Article: not found

          Patient-rated suitability of a novel electronic device for self-injection of subcutaneous interferon beta-1a in relapsing multiple sclerosis: an international, single-arm, multicentre, Phase IIIb study

          Background Multiple sclerosis (MS) currently requires long-term treatment with disease-modifying drugs, administered parenterally up to once daily. The need for regular self-injection can be a barrier to treatment for many patients. Autoinjectors can help patients overcome problems or concerns with self-injection and could, therefore, improve treatment adherence. This study was performed to assess the suitability of a new electronic device for the subcutaneous (sc) administration of interferon (IFN) beta-1a, 44 mcg three times weekly, for relapsing MS. Methods In this Phase IIIb, multicentre, single-arm study, patients with relapsing MS who had been consistently self-injecting sc IFN beta-1a using an autoinjector for at least 6 weeks were taught to use the new device and self-administered treatment for 12 weeks thereafter. Patient-rated suitability of the device was assessed at the end of Week 12 using the Patient User Trial Questionnaire. Patient satisfaction with, and evaluation of, the injection process was assessed using the MS Treatment Concern Questionnaire. Trainers evaluated the device using the Trainer User Trial Questionnaire. Results At Week 12, 71.6% (73/102) of patients considered the device 'very suitable' or 'suitable' for self-injection; 92.2% (94/102) reported some degree of suitability and only 7.8% (8/102) found the device 'not at all suitable'. At Weeks 4, 8 and 12, most patients reported that injection preparation and clean-up, performing injections and ease of device use in the previous 4 weeks compared favourably with, or was equivalent to, their previous experience of self-injection. Injection-related pain, injection reactions and 'flu-like' symptoms remained stable over the 12 weeks. Each device feature was rated 'very useful' or 'useful' by at least 80% of patients. All trainers and 95.2% (99/104) of patients found device functions 'very easy' or 'easy' to use. Overall convenience was considered the most important benefit of the device. Conclusions Most patients considered the new electronic injection device suitable for the sc injection of IFN beta-1a. They found the device easy to use with useful features, and reported benefits such as overall convenience. The device may, therefore, increase treatment adherence in patients with MS, particularly those with injection-related issues. Trial registration NCT00735007
            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers

            Objective: A disposable autoinjector was developed for subcutaneous (SC) self-injection by patients with chronic diseases. To verify its performance and evaluate its acceptance, a clinical study was conducted in healthy volunteers, comparing SC injections performed by subjects using the autoinjector with SC injections performed by nurses using a syringe. Methods: This was a randomized, single-center, crossover study comparing SC self-injection using an autoinjector with SC nurse-administered injection using a syringe. Two volumes (0.2 mL and 1 mL) were injected into healthy volunteers. Study objectives included assessment of the accuracy and consistency of the volume injected by the injection systems, and skin reaction and pain associated with the injection. The fluid depot in the SC tissue layer was evaluated by ultrasound. Subject acceptance was evaluated using questionnaires on attitudes and emotions towards the injection technique, and challenged by seeking the subjects’ preferred system for a final study injection or future treatment. Results: A total of 960 injections (480 with autoinjector, 480 with syringe) were performed in 40 subjects. There were no significant differences in mean fluid leakage and injected volumes between the systems. Pain associated with the injection was significantly lower with the auto-injector than with the syringe. Local skin reaction at the injection site was overall satisfactory. Injections were appropriately performed by all subjects. At study end, all 40 subjects preferred the autoinjector for a final study injection and for future treatment. Conclusion: This study indicated that the autoinjector used by the subject was similar to a syringe used by a nurse in terms of performance and safety in administering the injections, and better in terms of pain, overall acceptance, and preference.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              An open-label, multicenter study to evaluate the safe and effective use of the single-use autoinjector with an Avonex® prefilled syringe in multiple sclerosis subjects

              Background The ability to self-inject in patients with multiple sclerosis (MS) has been associated with a reduced risk of missed injections and drug discontinuation, and a beneficial effect on patients' independence. However, injection anxiety, needle phobia and disease-related disability are major barriers to a patient's ability to self-administer treatment. Use of an autoinjector may improve patients' ability to self-inject. This study evaluated the safe and effective use of Avonex Pen™ (prefilled pen), a single use autoinjector, for intramuscular delivery of interferon beta-1a (IM IFNβ-1a, Avonex) in MS patients. Methods This was a Phase IIIb, open-label, single-country, multicenter trial in MS patients currently using IM IFNβ-1a prefilled syringes. Patients received weekly 30 mcg IM IFNβ-1a treatment over 4 weeks. On Day 1, patients self-administered IM IFNβ-1a using a prefilled syringe at the clinic. On Day 8, patients received training on the prefilled pen and self-administered IM IFNβ-1a using the device. On Day 15, patients self-administered IM IFNβ-1a at home using the prefilled pen. A final injection occurred at the clinic on Day 22 when patients self-administered IM IFNβ-1a using the prefilled pen while clinic staff observed and completed a detailed questionnaire documenting patients' ability to self-inject with the device. Serum neopterin levels were evaluated pre and post-injection on Days 1 and 8. Adverse events were monitored throughout. Results Seventy-one (96%) patients completed the study. The overall success rate in safely and effectively using the prefilled pen was 89%. No device malfunctions occurred. One unsuccessful administration occurred at Day 22 due to patient error; no patient injury resulted. Patients gave the prefilled pen high ratings (8.7-9.3) on a 10-point scale for ease of use (0 = extremely difficult, 10 = extremely easy). Ninety-four percent of patients preferred the prefilled pen over the prefilled syringe. Induction of serum neopterin levels, serving as a biomarker for type 1 interferon action, was similar to that of the prefilled syringe. The prefilled pen demonstrated a safety profile comparable to the prefilled syringe. Conclusions The prefilled pen is a safe and effective device for administration of IM IFNβ-1a and represents an alternative method for self-injection for MS patients using this therapy. Trial registration This study is registered at clinicaltrials.gov, identifier: NCT00828204
                Bookmark

                Author and article information

                Journal
                Med Devices (Auckl)
                Med Devices (Auckl)
                Medical Devices: Evidence and Research
                Medical Devices (Auckland, N.Z.)
                Dove Medical Press
                1179-1470
                2015
                03 June 2015
                : 8
                : 255-264
                Affiliations
                [1 ]Ypsomed AG, Burgdorf, Switzerland
                [2 ]BERGO UK, Bearsden, Glasgow, UK
                Author notes
                Correspondence: Jakob Lange, Ypsomed AG, Brunnmattstrasse 6, CH-3401 Burgdorf, Switzerland, Tel +41 34 424 3187, Fax +41 34 424 3151, Email jakob.lange@ 123456ypsomed.com
                Article
                mder-8-255
                10.2147/MDER.S85938
                4461121
                15cf1971-9b47-4997-9445-972e571c552e
                © 2015 Lange et al. This work is published by Dove Medical Press Limited, and licensed under Creative Commons Attribution – Non Commercial (unported, v3.0) License

                The full terms of the License are available at http://creativecommons.org/licenses/by-nc/3.0/. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed.

                History
                Categories
                Original Research

                Biotechnology
                human factors engineering,autoinjector,handling study,use error,instructions for use,ypsomate™

                Comments

                Comment on this article