      Application of the Exradin W1 scintillator to determine Ediode 60017 and microDiamond 60019 correction factors for relative dosimetry within small MV and FFF fields.

      1 , , ,
      Physics in medicine and biology
      IOP Publishing

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          In this work we use EBT3 film measurements at 10 MV to demonstrate the suitability of the Exradin W1 (plastic scintillator) for relative dosimetry within small photon fields. We then use the Exradin W1 to measure the small field correction factors required by two other detectors: the PTW unshielded Ediode 60017 and the PTW microDiamond 60019. We consider on-axis correction-factors for small fields collimated using MLCs for four different TrueBeam energies: 6 FFF, 6 MV, 10 FFF and 10 MV. We also investigate percentage depth dose and lateral profile perturbations. In addition to high-density effects from its silicon sensitive region, the Ediode exhibited a dose-rate dependence and its known over-response to low energy scatter was found to be greater for 6 FFF than 6 MV. For clinical centres without access to a W1 scintillator, we recommend the microDiamond over the Ediode and suggest that 'limits of usability', field sizes below which a detector introduces unacceptable errors, can form a practical alternative to small-field correction factors. For a dosimetric tolerance of 2% on-axis, the microDiamond might be utilised down to 10 mm and 15 mm field sizes for 6 MV and 10 MV, respectively.

          Related collections

          Author and article information

          Journal
          Phys Med Biol
          Physics in medicine and biology
          IOP Publishing
          1361-6560
          0031-9155
          Sep 07 2015
          : 60
          : 17
          Affiliations
          [1 ] Institut Universitaire du Cancer de Toulouse-Oncopole, 1 Avenue Irène Joliot-Curie, 31059 Toulouse, Cedex 9, France.
          Article
          10.1088/0031-9155/60/17/6669
          26271097
          15f80617-a4eb-4223-b044-caca6bffafe7
          History

          Comments

          Comment on this article