      Full breast digital mammography with an amorphous silicon-based flat panel detector: physical characteristics of a clinical prototype.

      Medical physics
      Algorithms, Equipment Design, Mammography, instrumentation, methods, Radiographic Image Enhancement, Scattering, Radiation, Sensitivity and Specificity, Silicon, X-Rays

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The physical characteristics of a clinical prototype amorphous silicon-based flat panel imager for full-breast digital mammography have been investigated. The imager employs a thin thallium doped CsI scintillator on an amorphous silicon matrix of detector elements with a pixel pitch of 100 microm. Objective criteria such as modulation transfer function (MTF), noise power spectrum, detective quantum efficiency (DQE), and noise equivalent quanta were employed for this evaluation. The presampling MTF was found to be 0.73, 0.42, and 0.28 at 2, 4, and 5 cycles/mm, respectively. The measured DQE of the current prototype utilizing a 28 kVp, Mo-Mo spectrum beam hardened with 4.5 cm Lucite is approximately 55% at close to zero spatial frequency at an exposure of 32.8 mR, and decreases to approximately 40% at a low exposure of 1.3 mR. Detector element nonuniformity and electronic gain variations were not significant after appropriate calibration and software corrections. The response of the imager was linear and did not exhibit signal saturation under tested exposure conditions.

          Related collections

          Author and article information

          Journal
          10757607
          4280189
          10.1118/1.598895

          Chemistry
          Algorithms,Equipment Design,Mammography,instrumentation,methods,Radiographic Image Enhancement,Scattering, Radiation,Sensitivity and Specificity,Silicon,X-Rays

          Comments

          Comment on this article