      Troubleshooting electromagnetic interference in a patient with centrifugal flow left ventricular assist device and subcutaneous implantable cardioverter defibrillator.

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          A 25-year-old man with severe nonischemic dilated cardiomyopathy underwent subcutaneous implantable cardioverter defibrillator (S-ICD) implant and subsequently underwent HeartWare ventricular assist device (HVAD) placement. Postoperative interrogation revealed both primary and secondary S-ICD vectors inappropriately regarded sinus rhythm as "noise," and the alternate vector significantly undersensed sinus rhythm. The S-ICD was reinterrogated using high-resolution capture to visually confirm EMI with a dominant frequency in both the primary and secondary vectors of 46.67 Hz that fell within the S-ICD operational range of 9-60 Hz. The 46.67 Hz frequency correlated with the HVAD operational speed of 2,800 RPM. The HVAD pump speed was increased from 2,800 to 3,000 RPM, resulting in a dominant frequency of 50 Hz. The notch filter is nonprogrammable in S-ICDs. However, the built-in filter is 50 Hz for countries in European time zones as opposed to 60 Hz in US time zones due to differences in the anticipated noise from electrical sources within each continent. Thus, the S-ICD time zone was reprogrammed from EST to GMT, which reduced the notch filter from 60  to 50 Hz, resulting in S-ICD successfully eliminating EMI when the patient was in a supine position. The EMI interference was still intermittently present in the upright patient position. This case demonstrates the utility of high-resolution electrogram capture to identify the source and frequency of EMI in S-ICD and offers a potential avenue to troubleshoot dominant frequency oversensing by changing the device time zone.

          Related collections

          Author and article information

          Journal
          J. Cardiovasc. Electrophysiol.
          Journal of cardiovascular electrophysiology
          Wiley
          1540-8167
          1045-3873
          March 2018
          : 29
          : 3
          Affiliations
          [1 ] St. Vincent Medical Group, St. Vincent Hospital, Indianapolis, IN, USA.
          [2 ] Boston Scientific, St. Paul, MN, USA.
          Article
          10.1111/jce.13433
          29364552

          Comments

          Comment on this article