      Improved quantification in 123I cardiac SPECT imaging with deconvolution of septal penetration.

      Nuclear Medicine Communications

      Iodine Radioisotopes, Heart, radionuclide imaging, Humans, Image Enhancement, methods, Image Interpretation, Computer-Assisted, Imaging, Three-Dimensional, Algorithms, diagnostic use, Phantoms, Imaging, Quality Assurance, Health Care, Radiopharmaceuticals, Reproducibility of Results, Sensitivity and Specificity, Tomography, Emission-Computed, Single-Photon, instrumentation

      Read this article at

      ScienceOpenPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          (123)I is becoming an important radionuclide for cardiac imaging. Multiple, low-abundance, high-energy photons associated with (123)I imaging can cause septal penetration in the collimators and degrade quantification of the (123)I cardiac uptake. This study presents a method for the deconvolution of septal penetration (DSP) for improving quantification in (123)I cardiac single photon emission computed tomography (SPECT). Distance-dependent point spread functions were measured for low-energy high-resolution collimators on a dual-head SPECT system. The measured point spread functions were used in two-dimensional (2-D) and three-dimensional (3-D) models of the collimator response, respectively. 2-D DSP and 3-D DSP were then developed and implemented using iterative reconstruction. A cardiac torso phantom with an internal calibration source was designed with various heart-to-calibration ratios (HCRs) simulating different levels of a patient's uptake. SPECT acquisitions of the phantom were performed using optimized acquisition and processing parameters for (123)I cardiac SPECT. HCRs were calculated using planar projection and tomographic reconstructions. The paired t-test and regression analysis were used to compare the HCRs given by different calculation methods. SPECT produced more accurate HCRs than planar imaging. The slopes of the regression lines for SPECT using filtered back-projection were statistically significantly higher than those for planar imaging (0.2118 +/- 0.0297 vs. 0.0819 +/- 0.0070, P = 0.0001). 2-D DSP and 3-D DSP yielded similar HCRs that were close to the true HCR. The slopes of the regression lines for 2-D DSP and 3-D DSP were 0.9203 +/- 0.0523 and 0.9101 +/- 0.0304, respectively. The DSP HCRs were significantly more accurate than those calculated without DSP (P < 0.0001). DSP significantly improves quantification in (123)I cardiac SPECT imaging. 2-D DSP with its less computational burden shows promise for implementation in clinical practice so as to allow the use of the widely available low-energy, high-resolution collimators for quantitative I cardiac SPECT imaging.

          Related collections

          Author and article information

          Journal
          16794515

          Comments

          Comment on this article