      In vitro non-rigid life-size model of aortic arch aneurysm for endovascular prosthesis assessment.

      European Journal of Cardio-Thoracic Surgery : Official Journal of the European Association for Cardio-thoracic Surgery
      Aorta, Thoracic, surgery, Aortic Aneurysm, Thoracic, Blood Vessel Prosthesis Implantation, education, Computer Simulation, Humans, Magnetic Resonance Angiography, methods, standards, Models, Anatomic, Prosthesis Design, Silicone Elastomers

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          It is essential to evaluate new stent designs before in vivo testing. The purpose of this study was to develop and validate a controlled and reproducible patient-derived process to produce a life-size in vitro model of aortic arch aneurysm for endovascular procedure simulation. A three-dimensional magnetic resonance angiography (3D MRA) image derived from a 60-year-old patient with aortic arch aneurysm was segmented using a home-made software package which allows one-click automatic segmentation of the aorta, meshing, and conversion to standard tessellation language (STL) format. A rapid prototyping technique established a stereolithographic model to produce a replica of the whole aorta, including the arch aneurysm and supra-aortic arteries. The final model was made by pouring silicone rubber to obtain a sturdy, life-size, soft, transparent, plastic cast, accurately reproducing both the internal and external anatomy of the aortic aneurysm. This model was used under perfusion by an extracorporeal circulation pump, to test ex vivo stent deployment. The combination of easy segmentation and conversion to the STL format with industrial stereolithography techniques enabled a realistic silicon vascular phantom to be created for endovascular procedure simulation, image modality calibration, and new stent design.

          Related collections

          Author and article information

          Comments

          Comment on this article