      Accelerometer and Gyroscope Based Gait Analysis Using Spectral Analysis of Patients with Osteoarthritis of the Knee

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          [Purpose] A wide variety of accelerometer tools are used to estimate human movement, but there are no adequate data relating to gait symmetry parameters in the context of knee osteoarthritis. This study’s purpose was to evaluate a 3D-kinematic system using body-mounted sensors (gyroscopes and accelerometers) on the trunk and limbs. This is the first study to use spectral analysis for data post processing. [Subjects] Twelve patients with unilateral knee osteoarthritis (OA) (10 male) and seven age-matched controls (6 male) were studied. [Methods] Measurements with 3-D accelerometers and gyroscopes were compared to video analysis with marker positions tracked by a six-camera optoelectronic system (VICON 460, Oxford Metrics). Data were recorded using the 3D-kinematic system. [Results] The results of both gait analysis systems were significantly correlated. Five parameters were significantly different between the knee OA and control groups. To overcome time spent in expensive post-processing routines, spectral analysis was performed for fast differentiation between normal gait and pathological gait signals using the 3D-kinematic system. [Conclusions] The 3D-kinematic system is objective, inexpensive, accurate and portable, and allows long-term recordings in clinical, sport as well as ergonomic or functional capacity evaluation (FCE) settings. For fast post-processing, spectral analysis of the recorded data is recommended.

          Related collections

          Most cited references25

          • Record: found
          • Abstract: found
          • Article: not found

          Symmetry and limb dominance in able-bodied gait: a review.

          As one of the most universal of all human activities, gait in the able-bodied has received considerable attention, but many aspects still need to be clarified. Symmetry or asymmetry in the actions of the lower extremities during walking and the possible effect of laterality on gait are two prevalent and controversial issues. The purpose of this study was to review the work done over the last few decades in demonstrating: (a) whether or not the lower limbs behave symmetrically during able-bodied gait; and (b) how limb dominance affects the symmetrical or asymmetrical behavior of the lower extremities. The literature reviewed shows that gait symmetry has often been assumed, to simplify data collection and analysis. In contrast, asymmetrical behavior of the lower limbs during able-bodied ambulation was addressed in numerous investigations and was found to reflect natural functional differences between the lower extremities. These functional differences were probably related to the contribution of each limb in carrying out the tasks of propulsion and control during able-bodied walking. In current debates on gait symmetry in able-bodied subjects, laterality has been cited as an explanation for the existence of functional differences between the lower extremities, although a number of studies do not support the hypothesis of a relationship between gait symmetry and laterality. Further investigation is needed to demonstrate functional gait asymmetry and its relationship to laterality, taking into consideration the biomechanical aspects of gait.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Human movement analysis using stereophotogrammetry. Part 3. Soft tissue artifact assessment and compensation.

            When using optoelectronic stereophotogrammetry, skin deformation and displacement causes marker movement with respect to the underlying bone. This movement represents an artifact, which affects the estimation of the skeletal system kinematics, and is regarded as the most critical source of error in human movement analysis. A comprehensive review of the state-of-the-art for assessment, minimization and compensation of the soft tissue artifact (STA) is provided. It has been shown that STA is greater than the instrumental error associated with stereophotogrammetry, has a frequency content similar to the actual bone movement, is task dependent and not reproducible among subjects and, of lower limb segments, is greatest at the thigh. It has been shown that in in vivo experiments only motion about the flexion/extension axis of the hip, knees and ankles can be determined reliably. Motion about other axes at those joints should be regarded with much more caution as this artifact produces spurious effects with magnitudes comparable to the amount of motion actually occurring in those joints. Techniques designed to minimize the contribution of and compensate for the effects of this artifact can be divided up into those which model the skin surface and those which include joint motion constraints. Despite the numerous solutions proposed, the objective of reliable estimation of 3D skeletal system kinematics using skin markers has not yet been satisfactorily achieved and greatly limits the contribution of human movement analysis to clinical practice and biomechanical research. For STA to be compensated for effectively, it is here suggested that either its subject-specific pattern is assessed by ad hoc exercises or it is characterized from a large series of measurements on different subject populations. Alternatively, inclusion of joint constraints into a more general STA minimization approach may provide an acceptable solution.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Accelerometry: providing an integrated, practical method for long-term, ambulatory monitoring of human movement.

              Accelerometry offers a practical and low cost method of objectively monitoring human movements, and has particular applicability to the monitoring of free-living subjects. Accelerometers have been used to monitor a range of different movements, including gait, sit-to-stand transfers, postural sway and falls. They have also been used to measure physical activity levels and to identify and classify movements performed by subjects. This paper reviews the use of accelerometer-based systems in each of these areas. The scope and applicability of such systems in unsupervised monitoring of human movement are considered. The different systems and monitoring techniques can be integrated to provide a more comprehensive system that is suitable for measuring a range of different parameters in an unsupervised monitoring context with free-living subjects. An integrated approach is described in which a single, waist-mounted accelerometry system is used to monitor a range of different parameters of human movement in an unsupervised setting.
                Bookmark

                Author and article information

                Journal
                J Phys Ther Sci
                J Phys Ther Sci
                JPTS
                Journal of Physical Therapy Science
                The Society of Physical Therapy Science
                0915-5287
                2187-5626
                30 July 2014
                July 2014
                : 26
                : 7
                : 997-1002
                Affiliations
                [1) ] Institute for Diagnostic and Interventional Radiology, Heart Center, University Medical Center Goettingen, Germany
                [2) ] DZHK, German Center for Heart Research, Germany
                [3) ] Department of Cardiology and Pneumology, Heart Center, University Medical Center Goettingen, Germany
                [4) ] Division of Imaging Sciences and Biomedical Engineering, King’s College London, United Kingdom
                [5) ] Department of Sports Medicine, University Medical Center Goettingen, Germany
                [6) ] Department of Orthopedics, University Medical Center Würzburg, Germany
                Author notes
                [* ]Corresponding author. Wieland Staab, Institute for Diagnostic and Interventional Radiology, Heart Center, University Medical Center Goettingen: Robert-Koch-Str. 40, 37099 Goettingen, Germany. (E-mail: wieland.staab@ 123456med.uni-goettingen.de )
                Article
                jpts-2013-562
                10.1589/jpts.26.997
                4135223
                25140082
                1bc5baca-960f-49a4-88cb-7ef63e695a75
                2014©by the Society of Physical Therapy Science

                This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (by-nc-nd) License.

                History
                : 29 November 2013
                : 08 January 2014
                Categories
                Original Article

                gait,accelerometer,gyroscope
                gait, accelerometer, gyroscope

                Comments

                Comment on this article