      Repeatability and reproducibility of anterior chamber angle measurement with anterior segment optical coherence tomography.

      The British Journal of Ophthalmology
      Adult, Anterior Chamber, anatomy & histology, Humans, Iris, Lighting, Photic Stimulation, methods, Reproducibility of Results, Tomography, Optical Coherence, Trabecular Meshwork

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          To evaluate the repeatability and reproducibility of anterior chamber angle measurement obtained by anterior segment optical coherence tomography. Twenty-five normal subjects were invited for anterior chamber angle imaging with an anterior segment optical coherence tomography (OCT) on one randomly selected eye in three separate visits within a week. Each eye was imaged three times under room light (light intensity = 368 lux) and three times in the dark during the first visit. In the subsequent visits, each eye was imaged once in the light and once in the dark. The angle opening distance (AOD 500) and the trabecular-iris angle (TIA 500) were measured by a single observer. Only the nasal angle was analysed. Intrasession and intersession within-subject standard deviation (Sw), precision (1.96xSw), coefficient of variation (CVw) (100xSw/overall mean), and intraclass correlation coefficient (ICC) were calculated to evaluate repeatability and reproducibility. For intrasession repeatability, the Sw, precision, CVw and ICC of AOD/TIA were 45 microm/2.4 degrees , 88 microm/4.7 degrees , 5.8%/4.8% and 0.97/0.95 in the light; and 45 microm/2.1 degrees , 88 microm/4.2 degrees , 7.0%/5.0% and 0.98/0.97 in the dark. For intersession reproducibility, the Sw, precision, CVw and ICC of AOD/TIA were 79 microm/3.5 degrees , 155 microm/6.8 degrees , 10.0%/7.0%, 0.91/0.89 in the light; and 64 microm/3.4 degrees , 124 microm/6.6 degrees , 9.9%/7.8% and 0.95/0.92 in the dark. The anterior segment OCT demonstrated reliable anterior chamber angle measurement in different lighting conditions with good repeatability and reproducibility.

          Related collections

          Author and article information

          Journal
          17475709
          2095411
          10.1136/bjo.2007.118901

          Chemistry
          Adult,Anterior Chamber,anatomy & histology,Humans,Iris,Lighting,Photic Stimulation,methods,Reproducibility of Results,Tomography, Optical Coherence,Trabecular Meshwork

          Comments

          Comment on this article