      Umbral de impedancia transtorácica para la detección del paciente en desfibrilación Translated title: TRANSTHORACIC IMPEDANCE THRESHOLD FOR PATIENT DETECTION DURING DEFIBRILLATION

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          El valor de impedancia transtorácica para la desfibrilación se emplea en algunos algoritmos de compensación de impedancia para "personalizar" la descarga a cada paciente. El límite superior reportado para la impedancia puede ser sobrepasado si no se emplea gel conductor como interfaz entre las paletas y la piel. Los objetivos del presente trabajo son determinar un valor máximo medido de la impedancia que garantice la descarga aún cuando no se emplee gel conductor y establecer una comparación entre paletas para adultos y pediátricas, con y sin el empleo de gel conductor. Se midií la impedancia a un grupo de 27 voluntarios, encontrándose que el cociente del valor medido con las paletas pediátricas fue de 1,95 ± 0,61 veces el valor medido con paletas para adultos, cuando no se empleó gel como interfaz. La influencia del gel fue significativamente mayor para el caso de los hombres y en particular de los sujetos hirsutos. Similar resultado se obtuvo para la medición con gel al reducir el area de las paletas en 3,7 veces mediante el uso de un adaptador pediátrico. Se estimó un nuevo umbral de 400 Ω como limite superior para la detección de un paciente. Los resultados medidos sugieren que la ausencia de gel empleando las paletas pediátricas, equivale a la reducción, en casi cuatro veces, del área efectiva de las paletas para adultos.

          Translated abstract

          The value of transthoracic impedance for defibrillation is commonly employed by impedance compensation algorithms in order to �gtailor�h the discharge for each patient. The upper limit reported for the impedance may be exceeded if no conductive gel is used as an interface between paddles and skin. The present work is aimed to determining a maximum acceptable value when measured on a patient so that it can guarantee the discharge even if no gel is used and also to make a comparison between adult and pediatric paddles applying gel or not. The impedance of 27 volunteers was measured applying gel or not, finding that the value measured with pediatric paddles was 1,95 ± 0,61 times greater than the value measured with pediatric paddles, when no gel was applied. The influence of applying gel was significantly greater in men and particularly in hirsute individuals. A similar result was achieved when reducing the area of pediatric paddles 3,7 times by employing a pediatric adapter. Following these results, a top 400 Ω value was selected for patient detection. Results suggest that applying no gel to pediatric paddles is equivalent to the reduction in nearly four times of the adult paddle area.

          Related collections

          Most cited references10

          • Record: found
          • Abstract: not found
          • Article: not found

          "Factors affecting electrode-gelskin interface impedance in electrical impedance tomography"

            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            "Clinical Studies On Biphasic Defibrillation"

            M. Specht (2000)
              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              "Medición de la impedancia transtorácica para aplicaciones en desfibrilación"

                Bookmark

                Author and article information

                Journal
                uct
                Universidad, Ciencia y Tecnología
                uct
                Universidad Nacional Experimental Politécnica " Antonio José de Sucre", UNEXPO, Vicerrectorado Puerto Ordaz. (Puerto Ordaz, Bolívar, Venezuela )
                1316-4821
                September 2007
                : 11
                : 44
                : 129-134
                Affiliations
                [01] La Habana orgnameInstituto Central de Investigación Digital Cuba colorado@ 123456icid.cu
                [03] orgnameInstituto Central de Investigación Digital orgdiv11 jfolgueras@ 123456icid.cu
                [02] orgnameInstituto Central de Investigación Digital orgdiv1Departamento de Equipos Médicos 1 aportela@ 123456icid.cu
                Article
                S1316-48212007000300004 S1316-4821(07)01104404
                2456568a-01c1-4672-9a32-490e57db3c66

                http://creativecommons.org/licenses/by/4.0/

                History
                Page count
                Figures: 0, Tables: 0, Equations: 0, References: 12, Pages: 6
                Product

                SciELO Venezuela

                Categories
                Bioingeniería

                Desfibrilación,Compensación de impedancia,Impedancia transtorácica,Defibrillation,Impedance Compensation,Transthoracic Impedance

                Comments

                Comment on this article