      Two-dimensional scanning probe driven by a solenoid-based single actuator for optical coherence tomography.

      Optics letters
      Computer-Aided Design, Equipment Design, Equipment Failure Analysis, Lenses, Micro-Electrical-Mechanical Systems, instrumentation, Tomography, Optical Coherence, Transducers

      Read this article at

      ScienceOpenPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          We propose the two-dimensional scanning probe operating with a single actuator, which is thought useful as a sample probe for optical coherence tomography (OCT). The probe was designed to use a single-body lensed fiber cantilever loaded with an iron-bead and driven by a single-solenoid actuator. Elliptic spiral trace patterns were achieved using off-axis magnetic fields of the solenoid. A three-dimensional OCT image was obtained for a scanning area of 3.8 mm × 3.4 mm at an acquisition speed of 16.7 s/V. Up to 27 Hz B-scan rate, the proposed probe worked well, and 1000 A-scans were made per each B-scan. © 2011 Optical Society of America