      Comparative study of anterior segment measurements using 3 different instruments in myopic patients after ICL implantation

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          To compare and correlate anterior segment measurements of myopic eyes implanted with Implantable Collamer lens (ICL V4c) by using anterior segment optical coherence tomography (AS-OCT), Pentacam and ultrasound biomicroscopy (UBM).

          Methods

          Anterior chamber depth (ACD), distance between corneal endothelium and anterior surface of ICL(C-ICL) and central vault were measured in 82 phakic myopic eyes of 82 patients who underwent ICL surgery, by using AS-OCT, Pentacam and UBM consecutively at 3 months follow up. The correlation and agreement of instruments were accessed by using Intraclass correlation coefficient (ICC) and the Bland-Altman plot.

          Results

          AS-OCT showed higher ACD, C -ICL and central vault measurements than both of Pentacam and UBM ( P < 0.001), while Pentacam showed lower measurements than UBM ( P < 0.05). The Pearson correlation coefficient (r) was 0.91 to 0.96, and ICC was 0.95 to 0.98 for all measurements between difference devices (all P < 0.001). The 95% limits of agreement of ACD, C-ICL, vault measurements were 0.13 to 0.38 mm, − 0.07 to 0.27 mm, 0.08 to 0.34 mm between AS-OCT and Pentacam, − 0.03 to 0.33 mm, − 0.16 to 0.31 mm, − 0.10 to 0.26 mm between AS-OCT and UBM, and − 0.29 to 0.07 mm, − 0.25 to 0.20 mm, − 0.31 to 0.05 mm between Pentacam and UBM, respectively.

          Conclusions

          AS-OCT demonstrated significantly higher value, while Pentacam demonstrated significantly lower value than UBM for ACD, C-ICL and central vault measurements in myopic eyes after ICL surgery. Measurements with these instruments were highly correlated, but could not replace each other especially for vault. This study provided valuable information about how to judge the results of anterior segment parameters of eyes implanted with ICL V4c from different devices.

          Trial registration

          Registration number: ChiCTR-OOC-16008987. Retrospectively registered: 08 August 2016.

          Related collections

          Most cited references23

          • Record: found
          • Abstract: found
          • Article: not found

          Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment.

          To compare anterior segment parameters using quantitative imaging by anterior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscopy (UBM). Tertiary-care glaucoma research center. Sixty-three eyes of 63 subjects had anterior segment evaluation by AS-OCT (Visante-Zeiss) and UBM (Paradigm). Central corneal thickness (CCT), anterior chamber depth (ACD) (measured from the central corneal endothelium to the anterior lens capsule), and the peripheral iridocorneal angles (temporal and nasal) were assessed and compared. There was an excellent correlation between AS-OCT and UBM measurements for the nasal angle (r = 0.84; P<.0001), temporal angle (r = 0.86; P<.0001), ACD (r = 0.97; P<.0001), and CCT (r = 0.91; P<.0001). There was no significant difference (paired t test) between the mean ACD, CCT, and angle parameters measured by AS-OCT or UBM. The mean values of the parameters measured by AS-OCT and UBM were, respectively, as follows: nasal angle, 26.25 degrees +/- 11.0 (SD) and 28.27 +/- 11.3 degrees (P = .3); temporal angle, 25.1 +/- 11.4 degrees and 28.3 +/- 13.5 degrees (P = .15); ACD, 2.85 +/- 0.5 mm and 2.78 +/- 0.5 mm (P = .2); and CCT, 512 +/- 46 microm and 502 +/- 46 microm (P = .25). The AS-OCT images showed sharper definition of the scleral spur than the UBM images. Anterior segment optical coherence tomography and UBM can both be used for anterior segment measurements and yielded comparable results.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Four-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia.

            To assess the long-term clinical outcomes of implantation of a lens consisting of a biocompatible collagen copolymer (Visian implantable Collamer lens [ICL]; STAAR Surgical, Nidau, Switzerland) for moderate to high myopia. We evaluated 56 eyes of 34 patients with myopic refractive errors of -4.00 to -15.25 diopters (D) who underwent ICL implantation and routine postoperative examinations. Before and 1, 3, and 6 months and 1, 2, and 4 years after surgery, we assessed the safety, efficacy, predictability, stability, and adverse events of the surgery. Mean (SD) logMAR uncorrected and best spectacle-corrected visual acuities were -0.03 (0.23) and -0.21 (0.09), respectively, at 4 years after surgery. The mean (SD) safety and efficacy indexes were 1.19 (0.25) and 0.83 (0.29), respectively. At 4 years, 44 (79%) and 52 (93%) of the eyes were within +/-0.5 and +/-1.0 D, respectively, of the targeted correction. Mean (SD) manifest refraction changes of -0.24 (0.57) D occurred from 1 month to 4 years after surgery. No vision-threatening complications occurred during the observation period. Implantation of ICLs is safe and effective and provides predictable and stable refractive results in the treatment of moderate to high myopia during a 4-year observation period, suggesting its viability as a surgical option for the treatment of such eyes.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Prevalence of cataract after collagen copolymer phakic intraocular lens implantation for myopia, hyperopia, and astigmatism.

              To evaluate the prevalence of cataract after Implantable Collamer Lens phakic intraocular lens (pIOL) implantation with different models at different ages and for different refractive powers.
                Bookmark

                Author and article information

                Contributors
                wanting1982@zju.edu.cn
                yinhoufa@zju.edu.cn
                164877814@qq.com
                fangwuu@126.com
                361090557@qq.com
                +86-571-87783897 , yangyabo@zju.edu.cn
                Journal
                BMC Ophthalmol
                BMC Ophthalmol
                BMC Ophthalmology
                BioMed Central (London )
                1471-2415
                13 August 2019
                13 August 2019
                2019
                : 19
                : 182
                Affiliations
                ISNI 0000 0004 1759 700X, GRID grid.13402.34, Eye Center, Second Affiliated Hospital, School of Medicine, , Zhejiang University, ; 88 Jiefang Road, Hangzhou, 310009 China
                Author information
                http://orcid.org/0000-0003-2084-4422
                Article
                1194
                10.1186/s12886-019-1194-y
                6693247
                31409385
                251308c6-34f7-4796-a336-0181dc166de6
                © The Author(s). 2019

                Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License ( http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

                History
                : 4 December 2018
                : 2 August 2019
                Funding
                Funded by: FundRef http://dx.doi.org/10.13039/501100001809, National Natural Science Foundation of China;
                Award ID: 81500694
                Award ID: 81870558
                Award Recipient :
                Categories
                Research Article
                Custom metadata
                © The Author(s) 2019

                Ophthalmology & Optometry
                implantable collamer lens,anterior chamber depth,vault,optical coherence tomography,ultrasound biomicroscopy,pentacam

                Comments

                Comment on this article