      Quality of MR thermometry during palliative MR-guided high-intensity focused ultrasound (MR-HIFU) treatment of bone metastases

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Magnetic resonance (MR)-guided high-intensity focused ultrasound has emerged as a clinical option for palliative treatment of painful bone metastases, with MR thermometry (MRT) used for treatment monitoring. In this study, the general image quality of the MRT was assessed in terms of signal-to-noise ratio (SNR) and apparent temperature variation. Also, MRT artifacts were scored for their occurrence and hampering of the treatment monitoring.

          Methods

          Analyses were performed on 224 MRT datasets retrieved from 13 treatments. The SNR was measured per voxel over time in magnitude images, in the target lesion and surrounding muscle, and was averaged per treatment. The standard deviation over time of the measured temperature per voxel in MRT images, in the muscle outside the heated region, was defined as the apparent temperature variation and was averaged per treatment. The scored MRT artifacts originated from the following sources: respiratory and non-respiratory time-varying field inhomogeneities, arterial ghosting, and patient motion by muscle contraction and by gross body movement. Distinction was made between lesion type, location, and procedural sedation and analgesic (PSA).

          Results

          The average SNR was highest in and around osteolytic lesions (21 in lesions, 27 in surrounding muscle, n = 4) and lowest in the upper body (9 in lesions, 16 in surrounding muscle, n = 4). The average apparent temperature variation was lowest in osteolytic lesions (1.2°C, n = 4) and the highest in the upper body (1.7°C, n = 4). Respiratory time-varying field inhomogeneity MRT artifacts occurred in 85% of the datasets and hampered treatment monitoring in 81%. Non-respiratory time-varying field inhomogeneities and arterial ghosting MRT artifacts were most frequent (94% and 95%) but occurred only locally. Patient motion artifacts were highly variable and occurred less in treatments of osteolytic lesions and using propofol and esketamine as PSA.

          Conclusions

          In this study, the general image quality of MRT was observed to be higher in osteolytic lesions and lower in the upper body. Respiratory time-varying field inhomogeneity was the most prominent MRT artifact. Patient motion occurrence varied between treatments and seemed to be related to lesion type and type of PSA. Clinicians should be aware of these observed characteristics when interpreting MRT images.

          Electronic supplementary material

          The online version of this article (doi:10.1186/s40349-015-0026-7) contains supplementary material, which is available to authorized users.

          Related collections

          Most cited references46

          • Record: found
          • Abstract: found
          • Article: not found

          The safety and feasibility of extracorporeal high-intensity focused ultrasound (HIFU) for the treatment of liver and kidney tumours in a Western population

          High-intensity focused ultrasound (HIFU) provides a potential noninvasive alternative to conventional therapies. We report our preliminary experience from clinical trials designed to evaluate the safety and feasibility of a novel, extracorporeal HIFU device for the treatment of liver and kidney tumours in a Western population. The extracorporeal, ultrasound-guided Model-JC Tumor Therapy System (HAIFU™ Technology Company, China) has been used to treat 30 patients according to four trial protocols. Patients with hepatic or renal tumours underwent a single therapeutic HIFU session under general anaesthesia. Magnetic resonance imaging 12 days after treatment provided assessment of response. The patients were subdivided into those followed up with further imaging alone or those undergoing surgical resection of their tumours, which enabled both radiological and histological assessment. HIFU exposure resulted in discrete zones of ablation in 25 of 27 evaluable patients (93%). Ablation of liver tumours was achieved more consistently than for kidney tumours (100 vs 67%, assessed radiologically). The adverse event profile was favourable when compared to more invasive techniques. HIFU treatment of liver and kidney tumours in a Western population is both safe and feasible. These findings have significant implications for future noninvasive image-guided tumour ablation.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            MR imaging-guided focused ultrasound surgery of uterine leiomyomas: a feasibility study.

            The feasibility and safety of magnetic resonance (MR) imaging-guided focused ultrasound surgery for uterine leiomyomas is reported. Sequential sonications were delivered to nine targets. Temperature-sensitive phase-difference MR imaging monitored the location of the focus and measured tissue temperature elevations, ensuring therapeutic dose. MR images and hysterectomy specimens were evaluated. Six leiomyomas received full therapeutic doses, and 98.5% of the sonications were visualized. MR thermometry was successful in all sonications and cases. Focal necrotic lesions were seen in all cases at MR, and five were pathologically confirmed. MR imaging-guided focused ultrasound causes thermocoagulation and necrosis in uterine leiomyomas and is feasible and safe, without serious consequences.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Pain palliation in patients with bone metastases using MR-guided focused ultrasound surgery: a multicenter study.

              Noninvasive thermal ablation using magnetic resonance (MR)-guided focused ultrasound (MRgFUS) has been shown to be clinically effective in uterine fibroids, and is being evaluated for ablation of breast, liver, and brain lesions. Recently MRgFUS has been evaluated for palliation of pain caused by bone metastases. We present the clinical results of a multicenter study using MRgFUS for palliation of bone metastases pain. A multicenter study to evaluate the safety and efficacy of MRgFUS palliative treatment of bone metastases was conducted in patients suffering from painful metastatic bone lesions for which other treatments were either ineffective or not feasible. Thirty-one patients with painful bone metastases underwent the MRgFUS procedure in three medical centers. Treatment safety was evaluated by assessing the device-related complications. Effectiveness of pain palliation was evaluated using the visual analog pain score (VAS), and measurable changes in the intake of opioid analgesics. Thirty-six procedures were performed on 31 patients. Mean follow-up time was 4 months. 25 patients underwent the planned treatment and were available for 3 months post-treatment follow-up. 72% of the patients (18/25) reported significant pain improvement. Average VAS score was reduced from 5.9 prior to treatment to 1.8 at 3 months post treatment. 67% of patients with recorded medication data reported a reduction in their opioid usage. No device-related severe adverse events were recorded. The results suggest that MRgFUS has the ability to provide an accurate, effective, and safe noninvasive palliative treatment for patients with bone metastases.
                Bookmark

                Author and article information

                Contributors
                M.K.Lam-4@umcutrecht.nl
                M.Huisman-7@umcutrecht.nl
                Nijenhuis@maastrichtuniversity.nl
                MBosch@umcutrecht.nl
                M.Viergever@umcutrecht.nl
                C.Moonen@umcutrecht.nl
                W.Bartels@umcutrecht.nl
                Journal
                J Ther Ultrasound
                J Ther Ultrasound
                Journal of Therapeutic Ultrasound
                BioMed Central (London )
                2050-5736
                24 March 2015
                24 March 2015
                2015
                : 3
                : 5
                Affiliations
                [ ]Image Sciences Institute, University Medical Center Utrecht, Utrecht, The Netherlands
                [ ]Department of Radiology, University Medical Center Utrecht, Utrecht, The Netherlands
                Article
                26
                10.1186/s40349-015-0026-7
                4396149
                27e27a92-97bb-4964-bbac-1d6ce1ba4928
                © Lam et al.; licensee BioMed Central. 2015

                This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

                History
                : 21 November 2014
                : 7 March 2015
                Categories
                Research
                Custom metadata
                © The Author(s) 2015

                mr thermometry,bone metastases,mr-hifu,treatment monitoring,prfs

                Comments

                Comment on this article