      Closed loop anaesthesia at high altitude (3505 m above sea level): Performance characteristics of an indigenously developed closed loop anaesthesia delivery system

      other

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background:

          Closed loop anaesthesia delivery systems (CLADSs) are a recent advancement in accurate titration of anaesthetic drugs. They have been shown to be superior in maintaining adequate depth of anaesthesia with few fluctuations as compared with target-controlled infusion or manual titration of drug delivery.

          Methods:

          Twenty patients scheduled to undergo general abdominal or orthopaedic procedures under general anaesthesia at Leh (3505 m above sea level) were recruited as subjects. Anaesthesia was delivered by a patented closed loop system that uses the Bispectral Index (BIS™) as a feedback parameter to titrate the rate of propofol infusion. All vital parameters, drug infusion rate and the BIS™ values were continuously recorded and stored online by the system. The data generated was analysed for the adequacy of anaesthetic depth, haemodynamic stability and post-operative recovery parameters.

          Results:

          The CLADS was able to maintain a BIS™ within ±10 of the target of 50 for 85.0±7.8% of the time. Haemodynamics were appropriately maintained (heart rate and mean arterial blood pressure were within 25% of baseline values for 91.2±2.2% and 94.1±3% of the total anaesthesia time, respectively). Subjects were awake within a median of 3 min from cessation of drug infusion and achieved fitness to recovery room discharge within a median of 15 min. There were no adverse events or report of awareness under anaesthesia.

          Conclusions:

          The study demonstrates the safety of our CLADS at high altitude. It seeks to extend the use of our system in challenging anaesthesia environments. The system performance was also adequate and no adverse events were recorded.

          Related collections

          Most cited references12

          • Record: found
          • Abstract: found
          • Article: not found

          The heart and pulmonary circulation at high altitudes: healthy highlanders and chronic mountain sickness.

          More than 140 million people worldwide live >2500 m above sea level. Of them, 80 million live in Asia, and 35 million live in the Andean mountains. This latter region has its major population density living above 3500 m. The primary objective of the present study is to review the physiology, pathology, pathogenesis, and clinical features of the heart and pulmonary circulation in healthy highlanders and patients with chronic mountain sickness. A systematic review of worldwide literature was undertaken, beginning with the pioneering work done in the Andes several decades ago. Original articles were analyzed in most cases and English abstracts or translations of articles written in Chinese were reviewed. Pulmonary hypertension in healthy highlanders is related to a delayed postnatal remodeling of the distal pulmonary arterial branches. The magnitude of pulmonary hypertension increases with the altitude level and the degree of exercise. There is reversal of pulmonary hypertension after prolonged residence at sea level. Chronic mountain sickness develops when the capacity for altitude adaptation is lost. These patients have moderate to severe pulmonary hypertension with accentuated hypoxemia and exaggerated polycythemia. The clinical picture of chronic mountain sickness differs from subacute mountain sickness and resembles other chronic altitude diseases described in China and Kyrgyzstan. The heart and pulmonary circulation in healthy highlanders have distinct features in comparison with residents at sea level. Chronic mountain sickness is a public health problem in the Andean mountains and other mountainous regions around the world. Therefore, dissemination of preventive and therapeutic measures is essential.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            A simple study of awareness and dreaming during anaesthesia.

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Measuring the predictive performance of computer-controlled infusion pumps.

              Current measures of the performance of computer-controlled infusion pumps (CCIPs) are poorly defined, of little use to the clinician using the CCIP, and pharmacostatistically incorrect. We propose four measures be used to quantitate the performance of CCIPs: median absolute performance error (MDAPE), median performance error (MDPE), divergence, and wobble. These measures offer several significant advantages over previous measures. First, their definitions are based on the performance error as a fraction of the predicted (rather than measured) drug concentration, making the measures much more useful to the clinician. Second, the measures are defined in a way that addresses the pharmacostatistical issue of appropriate estimation of population parameters. Finally, the measure of inaccuracy, MDAPE, is defined in a way that is consistent with iteratively reweighted least squares nonlinear regression, a commonly used method of estimating pharmacokinetic parameters. These measures make it possible to quantitate the overall performance of a CCIP or to compare the predictive performance of CCIPs which differ in either general approach (e.g., compartmental model driven vs. plasma efflux approach), pump mechanics, software algorithms, or pharmacokinetic parameter sets.
                Bookmark

                Author and article information

                Journal
                Indian J Anaesth
                Indian J Anaesth
                IJA
                Indian Journal of Anaesthesia
                Medknow Publications & Media Pvt Ltd (India )
                0019-5049
                0976-2817
                May-Jun 2012
                : 56
                : 3
                : 238-242
                Affiliations
                [1]Department of Anaesthesia and Intensive Care, Post Graduate Institute of Medical Education and Research, Chandigarh, Punjab, India
                [1 ]Department of Anaesthesia and Intensive Care, SNM Hospital, Leh, Ladakh, India
                Author notes
                Address for correspondence: Prof. Goverdhan D Puri, Department of Anaesthesia and Intensive Care, PGIMER, Chandigarh - 160 012. Punjab, India. E-mail: gdpuri007@ 123456hotmail.com
                Article
                IJA-56-238
                10.4103/0019-5049.98765
                3425282
                22923821
                28611672-9318-4201-94f7-19f735cd0794
                Copyright: © Indian Journal of Anaesthesia

                This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                Categories
                Clinical Investigation

                Anesthesiology & Pain management
                closed loop anaesthesia,propofol,high altitude,bispectral index

                Comments

                Comment on this article