Closed loop anaesthesia delivery systems (CLADSs) are a recent advancement in accurate titration of anaesthetic drugs. They have been shown to be superior in maintaining adequate depth of anaesthesia with few fluctuations as compared with target-controlled infusion or manual titration of drug delivery.
Twenty patients scheduled to undergo general abdominal or orthopaedic procedures under general anaesthesia at Leh (3505 m above sea level) were recruited as subjects. Anaesthesia was delivered by a patented closed loop system that uses the Bispectral Index (BIS™) as a feedback parameter to titrate the rate of propofol infusion. All vital parameters, drug infusion rate and the BIS™ values were continuously recorded and stored online by the system. The data generated was analysed for the adequacy of anaesthetic depth, haemodynamic stability and post-operative recovery parameters.
The CLADS was able to maintain a BIS™ within ±10 of the target of 50 for 85.0±7.8% of the time. Haemodynamics were appropriately maintained (heart rate and mean arterial blood pressure were within 25% of baseline values for 91.2±2.2% and 94.1±3% of the total anaesthesia time, respectively). Subjects were awake within a median of 3 min from cessation of drug infusion and achieved fitness to recovery room discharge within a median of 15 min. There were no adverse events or report of awareness under anaesthesia.