      Pseudomonas aeruginosa and Serratia marcescens contamination associated with a manufacturing defect in bronchoscopes.

      The New England journal of medicine
      Adult, Aged, Aged, 80 and over, Bronchoscopes, microbiology, Disinfection, Electrophoresis, Gel, Pulsed-Field, Equipment Contamination, Equipment Design, Equipment Failure, Female, Hospital Bed Capacity, 100 to 299, Hospitals, Community, Humans, Male, Middle Aged, Pneumonia, diagnosis, Pseudomonas Infections, etiology, Pseudomonas aeruginosa, classification, isolation & purification, Serratia Infections, Serratia marcescens, Tennessee

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Several outbreaks and pseudo-outbreaks of Pseudomonas aeruginosa and Serratia marcescens infections associated with bronchoscopy have been reported. We conducted an investigation of P. aeruginosa and S. marcescens isolates related to bronchoscopy at a community hospital. We reviewed the records of all bronchoscopic procedures at the community hospital from July to October 2001. Environmental samples were obtained. Pulsed-field gel electrophoresis (PFGE) was performed on isolates of P. aeruginosa. From July 1 to October 31, 2001, 66 bronchoscopic procedures were performed in 60 patients, and 43 specimens were obtained for bacterial culture; 20 of the specimens (47 percent) were positive for P. aeruginosa. Six (30 percent) of the specimens that were positive for P. aeruginosa also yielded S. marcescens. All 20 P. aeruginosa isolates were associated with procedures performed with three of four new bronchoscopes from the same manufacturer. Contrary to manufacturing specifications, the biopsy-port caps on all four bronchoscopes were easily removable, and P. aeruginosa was cultured from the biopsy ports of the three implicated bronchoscopes. The PFGE patterns of P. aeruginosa isolates from the bronchoscopes, patients, and two environmental samples were indistinguishable. One patient was hospitalized with P. aeruginosa pneumonia 11 days after bronchoscopy. The manufacturer reported a design change instituted in 1997, and production problems may have resulted in the distribution of bronchoscopes that did not meet specifications. We documented contamination of bronchoscopes with P. aeruginosa and S. marcescens and possible infection of patients at a community hospital as a result of the inadequate disinfection of bronchoscopes because of a manufacturing defect. Copyright 2003 Massachusetts Medical Society

          Related collections

          Author and article information

          Comments

          Comment on this article