3
views
0
recommends
+1 Recommend
1 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      The accuracy of a point-of-care test among different operators using the QBC Autoread Plus Analyzer for the measurement of a basic full blood count

      research-article
      , ,
      SAMJ: South African Medical Journal
      Health and Medical Publishing Group

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          BACKGROUND. Major issues around the implementation of point-of-care testing (POCT) include: user type, regulatory control, ongoing quality monitoring and limited guideline adherence.OBJECTIVES. To determine if there are significant differences in the results of a POC full blood count test between different levels of healthcare and non-medical 'lay users compared with laboratory users (technologists/pathologists).METHODS. This article retrospectively reviews the technical evaluation database of the intensive care unit (ICU), Chris Hani Baragwanath Academic Hospital, Johannesburg, South Africa. We searched for samples analysed by doctors, clerks and laboratory personnel. A minimum number of 60 comparisons were required. Bland-Altman plots, Spearman's correlation and Passing-Bablok fit were used to analyse the dataset.RESULTS. There were 72 comparisons for haematocrit (Hct), 98 for white cell count (WCC) and 137 for platelets (Pit) between the clerk and laboratory personnel. The correlations were 0.91, 0.96 and 0.92, respectively. All were statistically significant. Using the Bland-Altman method, there was good agreement between results of the clerk and those of the laboratory staff, with a mean bias of 0.5% (Hct), 0.1 χ 10(9)/L (WCC) and 10 χ 10(9)/L (Pit). An insufficient number of tests were performed by medical doctors for statistical comparison.CONCLUSIONS. With appropriate training, non-medical, non-healthcare lay users are able to perform a moderately complex POCT with similar accuracy as trained laboratory professionals. The focus should shift to equipment and quality management processes rather than the medical/technical qualification of the user.

          Related collections

          Most cited references10

          • Record: found
          • Abstract: not found
          • Article: not found

          Point of care testing.

            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients

            Introduction Rapid detection of abnormal biological values using point-of-care (POC) testing allows clinicians to promptly initiate therapy; however, there are concerns regarding the reliability of POC measurements. We investigated the agreement between the latest generation blood gas analyzer and central laboratory measurements of electrolytes, bicarbonate, hemoglobin, hematocrit, and glucose. Methods 314 paired samples were collected prospectively from 51 critically ill patients. All samples were drawn simultaneously in the morning from an arterial line. BD Vacutainer tubes were analyzed in the central laboratory using Beckman Coulter analyzers (AU 5800 and DxH 800). BD Preset 3 ml heparinized-syringes were analyzed immediately in the ICU using the POC Siemens RAPIDPoint 500 blood gas system. We used CLIA proficiency testing criteria to define acceptable analytical performance and interchangeability. Results Biases, limits of agreement (±1.96 SD) and coefficients of correlation were respectively: 1.3 (-2.2 to 4.8 mmol/L, r = 0.936) for sodium; 0.2 (-0.2 to 0.6 mmol/L, r = 0.944) for potassium; -0.9 (-3.7 to 2 mmol/L, r = 0.967) for chloride; 0.8 (-1.9 to 3.4 mmol/L, r = 0.968) for bicarbonate; -11 (-30 to 9 mg/dL, r = 0.972) for glucose; -0.8 (-1.4 to -0.2 g/dL, r = 0.985) for hemoglobin; and -1.1 (-2.9 to 0.7%, r = 0.981) for hematocrit. All differences were below CLIA cut-off values, except for hemoglobin. Conclusions Compared to central Laboratory analyzers, the POC Siemens RAPIDPoint 500 blood gas system satisfied the CLIA criteria of interchangeability for all tested parameters, except for hemoglobin. These results are warranted for our own procedures and devices. Bearing these restrictions, we recommend clinicians to initiate an appropriate therapy based on POC testing without awaiting a control measurement.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              Point-of-care method for total white cell count: an evaluation of the HemoCue WBC device

              Point-of-care testing (POCT) is becoming an important adjunct to haematology laboratory practice. An important component of the blood count is the total white cell count (WBC). Previously, this required laborious microscopic cell counting, but it can now be performed by means of automation; however, in many under-resourced countries, costly automated counters are only available in very few central hospitals. Moreover, neither method is practical in most POCT situations. The HemoCue WBC has been developed as a simplified alternative method, consisting of a reagent pre-loaded disposable cuvette together with basic image analysis technology. This report describes an assessment of its utility. The WBC of 500 routine blood samples from the hospital were tested in parallel by the HemoCue WBC and by a reference analyser to assess accuracy and utility of the former. The tests included precision, linearity, type of blood sample and anticoagulant and potential interfering substances in blood specimens. In the tests for accuracy, 192 of the 200 showed percentage difference from the NEQAS reference of 2% normoblasts or reticulocytosis showed significant differences from the reference measurements. The HemoCue WBC is reliable for WBC counts within the analytical range of 0.4–30.0 × 109/l, except in samples where there are significant numbers of normoblasts or reticulocytes. It is simple to use and provides a valuable advance in the facilities available for POCT in haematology.
                Bookmark

                Author and article information

                Contributors
                Role: ND
                Role: ND
                Role: ND
                Journal
                samj
                SAMJ: South African Medical Journal
                SAMJ, S. Afr. med. j.
                Health and Medical Publishing Group (Cape Town, Western Cape Province, South Africa )
                0256-9574
                2078-5135
                December 2019
                : 109
                : 12
                : 952-956
                Affiliations
                [02] Johannesburg orgnameIntensive Care Unit orgdiv1Chris Hani Baragwanath Academic Hospital South Africa
                [01] Johannesburg orgnameUniversity of the Witwatersrand orgdiv1Faculty of Health Sciences orgdiv2Academic Hospital and Intensive Care Unit South Africa
                [03] Johannesburg orgnameUniversity of the Witwatersrand orgdiv1Faculty of Health Sciences orgdiv2School of Clinical Medicine South Africa
                Article
                S0256-95742019001200013 S0256-9574(19)10901200013
                10.7196/samj.2019.v109i12.13981
                2c28c26e-4b8f-4280-9083-1b5cb10c7e89

                This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

                History
                Page count
                Figures: 0, Tables: 0, Equations: 0, References: 14, Pages: 5
                Product

                SciELO South Africa

                Categories
                Research

                Comments

                Comment on this article