      Single Use or Disposable Flexible Bronchoscopes: Bench Top and Preclinical Comparison of Currently Available Devices

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Data regarding the risk of infection related to reusable bronchoscopes, the global drive toward disposable technology and the COVID-19 pandemic have led to an increase in the use and production of single use or disposable bronchoscopes. An in-depth comparison of all available devices has not been published.

          Methods

          A benchtop comparison of the Ambu ®aScope TM, Boston Scientific ® EXALT TM Model B, the Surgical Company Broncoflex © Vortex, Pentax ® Medical ONE Pulmo™, and Vathin ® H-Steriscope TM (all 2.8 mm inner dimension other than the Pentax single-use flexible bronchoscope (3 mm)) was undertaken including measurement of maximal flexion and extension angles, thumb force required and suction with and without biopsy forceps. Thereafter, preclinical assessment was performed with data collected including experience, gender, hand size, and scope preference.

          Results

          The Vathin single-use flexible bronchoscope had the biggest range of tip movement from flexion to extension with and without forceps. The Boston single-use flexible bronchoscope required the maximal thumb force but had the least reduction of tip movement with forceps. The Boston single-use flexible bronchoscope significantly outperformed all other scopes including the standard Pentax scope and was the only scope capable of suctioning pseudo-mucus around the forceps. Although there was no significant difference in preference in the overall group, females and those with smaller hand size preferred the Pentax and males the Broncoflex single-use flexible bronchoscope.

          Conclusions

          Currently available single-use flexible bronchoscopes differ in several factors other than scope sizes and monitor including suction, turning envelope, and handle size. Performance in the clinical setting will be key to their success.

          Related collections

          Most cited references28

          • Record: found
          • Abstract: found
          • Article: not found

          Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy.

          Flexible endoscopy is a widely used diagnostic and therapeutic procedure. Contaminated endoscopes are the medical devices frequently associated with outbreaks of health care-associated infections. Accurate reprocessing of flexible endoscopes involves cleaning and high-level disinfection followed by rinsing and drying before storage. Most contemporary flexible endoscopes cannot be heat sterilized and are designed with multiple channels, which are difficult to clean and disinfect. The ability of bacteria to form biofilms on the inner channel surfaces can contribute to failure of the decontamination process. Implementation of microbiological surveillance of endoscope reprocessing is appropriate to detect early colonization and biofilm formation in the endoscope and to prevent contamination and infection in patients after endoscopic procedures. This review presents an overview of the infections and cross-contaminations related to flexible gastrointestinal endoscopy and bronchoscopy and illustrates the impact of biofilm on endoscope reprocessing and postendoscopic infection.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            A systematic review and cost effectiveness analysis of reusable vs. single‐use flexible bronchoscopes

            Summary The cost effectiveness of reusable vs. single‐use flexible bronchoscopy in the peri‐operative setting has yet to be determined. We therefore aimed to determine this and hypothesised that single‐use flexible bronchoscopes are cost effective compared with reusable flexible bronchoscopes. We conducted a systematic review of the literature, seeking all reports of cross‐contamination or infection following reusable bronchoscope use in any clinical setting. We calculated the incidence of these outcomes and then determined the cost per patient of treating clinical consequences of bronchoscope‐induced infection. We also performed a micro‐costing analysis to quantify the economics of reusable flexible bronchoscopes in the peri‐operative setting from a high‐throughput tertiary centre. This produced an accurate estimate of the cost per use of reusable flexible bronchoscopes. We then performed a cost effectiveness analysis, combining the data obtained from the systematic review and micro‐costing analysis. We included 16 studies, with a reported incidence of cross‐contamination or infection of 2.8%. In the micro‐costing analysis, the total cost per use of a reusable flexible bronchoscope was calculated to be £249 sterling. The cost per use of a single‐use flexible bronchoscope was £220 sterling. The cost effectiveness analysis demonstrated that reusable flexible bronchoscopes have a cost per patient use of £511 sterling due to the costs of treatment of infection. The findings from this study suggest benefits from the use of single‐use flexible bronchoscopes in terms of cost effectiveness, cross‐contamination and resource utilisation.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              A prospective randomized equivalence trial of the GlideScope Cobalt® video laryngoscope to traditional direct laryngoscopy in neonates and infants.

              Intubation in children is increasingly performed using video laryngoscopes. Many pediatric studies examine novice laryngoscopists or describe single patient experiences. This prospective randomized nonblinded equivalence trial compares intubation time for the GlideScope Cobalt® video laryngoscope (GCV, Verathon Medical, Bothell, WA) with direct laryngoscopy with a Miller blade (DL, Heine, Dover, NH) in anatomically normal neonates and infants. The primary hypothesis was that intubation times with GCV would be noninferior to DL. Sixty subjects presenting for elective surgery were randomly assigned to intubation using GCV or DL. Intubation time, time to best view, percentage of glottic opening score, and intubation success were documented. We defined an intubation time difference of less than 10 s as clinically insignificant. There was no difference in intubation time between the groups (GCV median = 22.6 s; DL median = 21.4 s; P = 0.24). The 95% one-sided CI for mean difference between the groups was less than 8.3 s. GCV yielded faster time to best view (median = 8.1 s; DL 9.9 s; P = 0.03). Endotracheal tube passage time was longer for GCV (median = 14.3 s; DL 8.5 s; P = 0.007). The percentage of glottic opening score was improved with GCV (median 100; DL 80; P < 0.0001). Similar intubation times and success rates were achieved in anatomically normal neonates and infants with the GCV as with DL. The GCV yielded faster time to best view and better views but longer tube passage times than DL.
                Bookmark

                Author and article information

                Journal
                J Intensive Care Med
                J Intensive Care Med
                JIC
                spjic
                Journal of Intensive Care Medicine
                SAGE Publications (Sage CA: Los Angeles, CA )
                0885-0666
                1525-1489
                7 January 2023
                June 2023
                : 38
                : 6
                : 519-528
                Affiliations
                [1 ]Department of Respiratory Medicine, Ringgold 57983, universityCork University Hospital; , Cork, Ireland
                [2 ]College of Medicine and Health, University College Cork, Cork, Ireland
                Author notes
                [*]Marcus Kennedy, MD, FRCPI, FCCP, Department of Respiratory Medicine, Cork University Hospital, Wilton, Cork City, Ireland. Email: marcus.kennedy@ 123456hse.ie
                Author information
                https://orcid.org/0000-0002-8618-7499
                https://orcid.org/0000-0003-2119-7714
                Article
                10.1177_08850666221148645
                10.1177/08850666221148645
                10114257
                36609193
                3215841c-2577-48f9-84fb-e716cee58b65
                © The Author(s) 2023

                This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License ( https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access page ( https://us.sagepub.com/en-us/nam/open-access-at-sage).

                History
                Categories
                Original Research
                Custom metadata
                ts19

                single-use bronchoscopy,disposable bronchoscopy,intubation,bronchoalveolar lavage

                Comments

                Comment on this article