      Impact of Energy and Access Methods on Extrahepatic Tumor Spreading and the Ablation Zone: An Ex vivo Experiment Using a Subcapsular Tumor Model

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Objective

          To evaluate the impact of energy and access methods on extrahepatic tumor spreading and the ablation zone in an ex vivo subcapsular tumor mimic model with a risk of extrahepatic tumor spreading.

          Materials and Methods

          Forty-two tumor-mimics were created in bovine liver blocks by injecting a mixture of iodine contrast material just below the liver capsule. Radiofrequency (RF) ablations were performed using an electrode placed parallel or perpendicular to hepatic surface through the tumor mimic with low- and high-power protocols (groups 1 and 2, respectively). Computed tomography (CT) scans were performed before and after ablation. The presence of contrast leak on the hepatic surface on CT, size of ablation zone, and timing of the first roll-off and popping sound were compared between the groups.

          Results

          With parallel access, one contrast leak in group 1 (1/10, 10%) and nine in group 2 (9/10, 90%) ( p < 0.001) were identified on post-ablation CT. With perpendicular access, six contrast leaks were identified in each group (6/11, 54.5%). The first roll-off and popping sound were significantly delayed in group 1 irrespective of the access method ( p = 0.002). No statistical difference in the size of the ablation zone of the liver specimen was observed between the two groups ( p = 0.247).

          Conclusion

          Low-power RF ablation with parallel access is proposed to be effective and safe from extrahepatic tumor spreading in RF ablation of a solid hepatic tumor in the subcapsular location. Perpendicular placement of an electrode to the capsule is associated with a risk of extrahepatic tumor spreading regardless of the power applied.

          Related collections

          Most cited references24

          • Record: found
          • Abstract: found
          • Article: not found

          Principles of and advances in percutaneous ablation.

          Image-guided tumor ablation with both thermal and nonthermal sources has received substantial attention for the treatment of many focal malignancies. Increasing interest has been accompanied by continual advances in energy delivery, application technique, and therapeutic combinations with the intent to improve the efficacy and/or specificity of ablative therapies. This review outlines clinical percutaneous tumor ablation technology, detailing the science, devices, techniques, technical obstacles, current trends, and future goals in percutaneous tumor ablation. Methods such as chemical ablation, cryoablation, high-temperature ablation (radiofrequency, microwave, laser, and ultrasound), and irreversible electroporation will be discussed. Advances in technique will also be covered, including combination therapies, tissue property modulation, and the role of computer modeling for treatment optimization. © RSNA, 2011.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Treatment of focal liver tumors with percutaneous radio-frequency ablation: complications encountered in a multicenter study.

            To report complications encountered by members of a collaborative group who performed radio-frequency (RF) ablation in patients with focal liver cancer. Members of 41 Italian centers that were part of a collaborative group used a percutaneous internally cooled RF ablation technique and a standardized protocol for follow-up. They completed a questionnaire regarding number of deaths, presumed cause of death, and likelihood of its relationship to the RF procedure; number and types of major complications; and types of minor complications and side effects. Enrollment included 2,320 patients with 3,554 lesions (size, 3.1 cm +/- 1.1 [SD] in diameter): 1,610 had hepatocellular carcinoma with chronic liver disease; 693 had metastases, predominantly from colorectal cancer (n = 501); and 17 had cholangiocellular carcinoma. Number and characteristics of complications (ie, deaths and major and minor complications) attributed to the procedure were reported. Data were subsequently analyzed with analysis of variance to determine whether the major complication rate was related to tumor size, number of ablation sessions, or electrode type (single or cluster). In total, 3,554 lesions were treated. Six deaths (0.3%) were noted, including two caused by multiorgan failure following intestinal perforation; one case each of septic shock following Staphylococcus aureus-caused peritonitis, massive hemorrhage following tumor rupture, liver failure following stenosis of right bile duct; and one case of sudden death of unknown cause 3 days after the procedure. Fifty (2.2%) patients had additional major complications. The most frequent of these were peritoneal hemorrhage, neoplastic seeding, intrahepatic abscesses, and intestinal perforation. An increased number of RF sessions were related to a higher rate of major complications (P <.01), whereas the number of complications was not significantly different when tumor size or electrode type were compared. Minor complications were observed in less than 5% of patients. Results of this study confirm that RF ablation is a relatively low-risk procedure for the treatment of focal liver tumors.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Radiofrequency ablation for hepatocellular carcinoma in so-called high-risk locations.

              We evaluated the efficacy and safety of radiofrequency (RF) ablation for hepatocellular carcinoma (HCC) in presumably high-risk locations. Between February 1999 and December 2001, we performed RF ablation on 1419 nodules in 636 consecutive HCC patients, of which 231 nodules in 207 patients were in high-risk locations, defined as less than 5 mm from a large vessel or an extrahepatic organ. Eighty-one patients had a nodule adjacent to a large vessel, 145 patients had a nodule adjacent to an extrahepatic organ, of whom 19 also had one adjacent to a large vessel. Early complications and local tumor progression were analyzed with regard to the location of each nodule. The mean nodule diameter and average number per patient were 27 mm and 2.3, respectively. Early complications, within 30 days after ablation, occurred in 12 of 207 patients (5.8%) with a nodule in a high-risk location and in 15 of 429 patients (3.5%) without (P = .1776). There was no significant difference in local tumor progression rate between nodules in high-risk locations (1 year: 2.1%, 2 years: 3.1%, 3 years: 3.1%) and those elsewhere (1 year: 0.6%, 2 years: 1.7%, 3 years: 2.5%) (P = .2745). In conclusion, HCC nodules adjacent to a large vessel or extrahepatic organ were treated with RF ablation without compromising the efficacy of the procedure. However, even though without significant difference, some complications occurred at risky locations and need to be carefully considered.
                Bookmark

                Author and article information

                Journal
                Korean J Radiol
                Korean J Radiol
                KJR
                Korean Journal of Radiology
                The Korean Society of Radiology
                1229-6929
                2005-8330
                April 2019
                25 February 2019
                : 20
                : 4
                : 580-588
                Affiliations
                Department of Radiology, College of Medicine, Ewha Womans University, Seoul, Korea.
                Author notes
                Corresponding author: Jeong Kyong Lee, MD, Department of Radiology, College of Medicine, Ewha Womans University, 1071 Anyangcheon-ro, Yangcheon-gu, Seoul 07985, Korea. Tel: (822) 2650-5174, Fax: (822) 2650-5302, kyongmd@ 123456ewha.ac.kr
                Author information
                https://orcid.org/0000-0002-3321-2507
                https://orcid.org/0000-0002-5507-0140
                Article
                10.3348/kjr.2018.0564
                6424823
                30887740
                32564c40-f0a8-4fff-81f9-1ed77639009c
                Copyright © 2019 The Korean Society of Radiology

                This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License ( http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 21 August 2018
                : 08 November 2018
                Categories
                Gastrointestinal Imaging
                Original Article

                Radiology & Imaging
                radiofrequency ablation,thermal ablation,experimental liver neoplasm,neoplasm seeding