16
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Effects of Neurally Adjusted Ventilatory Assist (NAVA) levels in non-invasive ventilated patients: titrating NAVA levels with electric diaphragmatic activity and tidal volume matching

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Neurally adjusted ventilatory assist (NAVA) delivers pressure in proportion to diaphragm electrical activity ( Eadi). However, each patient responds differently to NAVA levels. This study aims to examine the matching between tidal volume ( Vt) and patients’ inspiratory demand ( Eadi), and to investigate patient-specific response to various NAVA levels in non-invasively ventilated patients.

          Methods

          12 patients were ventilated non-invasively with NAVA using three different NAVA levels. NAVA100 was set according to the manufacturer’s recommendation to have similar peak airway pressure as during pressure support. NAVA level was then adjusted ±50% (NAVA50, NAVA150). Airway pressure, flow and Eadi were recorded for 15 minutes at each NAVA level. The matching of Vt and integral of Eadi ( ʃEadi) were assessed at the different NAVA levels. A metric, Range90, was defined as the 5-95% range of Vt/ ʃEadi ratio to assess matching for each NAVA level. Smaller Range90 values indicated better matching of supply to demand.

          Results

          Patients ventilated at NAVA50 had the lowest Range90 with median 25.6 uVs/ml [Interquartile range (IQR): 15.4-70.4], suggesting that, globally, NAVA50 provided better matching between ʃEadi and Vt than NAVA100 and NAVA150. However, on a per-patient basis, 4 patients had the lowest Range90 values in NAVA100, 1 patient at NAVA150 and 7 patients at NAVA50. Robust coefficient of variation for ʃEadi and Vt were not different between NAVA levels.

          Conclusions

          The patient-specific matching between ʃEadi and Vt was variable, indicating that to obtain the best possible matching, NAVA level setting should be patient specific. The Range90 concept presented to evaluate Vt/ ʃEadi is a physiologic metric that could help in individual titration of NAVA level.

          Related collections

          Most cited references16

          • Record: found
          • Abstract: not found
          • Article: not found

          Neural control of mechanical ventilation in respiratory failure.

            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Early noninvasive ventilation averts extubation failure in patients at risk: a randomized trial.

            Respiratory failure after extubation and reintubation is associated with increased morbidity and mortality. To assess the efficacy of noninvasive ventilation in averting respiratory failure after extubation in patients at increased risk. A prospective randomized controlled trial was conducted in 162 mechanically ventilated patients who tolerated a spontaneous breathing trial after recovery from the acute episode but had increased risk for respiratory failure after extubation. Patients were randomly allocated after extubation to receive noninvasive ventilation for 24 h (n = 79), or conventional management with oxygen therapy (control group, n = 83). The primary end-point variable was the decrease in respiratory failure after extubation. In the noninvasive ventilation group, respiratory failure after extubation was less frequent (13, 16 vs. 27, 33%; p = 0.029) and the intensive care unit mortality was lower (2, 3 versus 12, 14%; p = 0.015). However, 90-d survival did not change significantly between groups. Separate analyses of patients without and with hypercapnia (arterial CO(2) tension greater than 45 mm Hg) during the spontaneous breathing trial showed that noninvasive ventilation improved intensive care unit mortality (0 vs. 4, 18%; p = 0.035) and 90-d survival (p = 0.006) in hypercapnic patients only; of them, 98% had chronic respiratory disorders. The early use of noninvasive ventilation averted respiratory failure after extubation and decreased intensive care unit mortality among patients at increased risk. The beneficial effect of noninvasive ventilation in improving survival of hypercapnic patients with chronic respiratory disorders warrants a new prospective clinical trial.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Patient-ventilator asynchrony during non-invasive ventilation for acute respiratory failure: a multicenter study.

              To determine the prevalence of patient-ventilator asynchrony in patients receiving non-invasive ventilation (NIV) for acute respiratory failure. Prospective multicenter observation study. Intensive care units in three university hospitals. Patients consecutively admitted to ICU were included. NIV, performed with an ICU ventilator, was set by the clinician. Airway pressure, flow, and surface diaphragmatic electromyography were recorded continuously for 30 min. Asynchrony events and the asynchrony index (AI) were determined from visual inspection of the recordings and clinical observation. A total of 60 patients were included, 55% of whom were hypercapnic. Auto-triggering was present in 8 (13%) patients, double triggering in 9 (15%), ineffective breaths in 8 (13%), premature cycling 7 (12%) and late cycling in 14 (23%). An AI > 10%, indicating severe asynchrony, was present in 26 patients (43%), whose median (25-75 IQR) AI was 26 (15-54%). A significant correlation was found between the magnitude of leaks and the number of ineffective breaths and severity of delayed cycling. Multivariate analysis indicated that the level of pressure support and the magnitude of leaks were weakly, albeit significantly, associated with an AI > 10%. Patient comfort scale was higher in pts with an AI < 10%. Patient-ventilator asynchrony is common in patients receiving NIV for acute respiratory failure. Our results suggest that leaks play a major role in generating patient-ventilator asynchrony and discomfort, and point the way to further research to determine if ventilator functions designed to cope with leaks can reduce asynchrony in the clinical setting.
                Bookmark

                Author and article information

                Contributors
                Journal
                Biomed Eng Online
                Biomed Eng Online
                BioMedical Engineering OnLine
                BioMed Central
                1475-925X
                2013
                2 July 2013
                : 12
                : 61
                Affiliations
                [1 ]Department of Mechanical Engineering, University of Canterbury, Christchurch, New Zealand
                [2 ]GIGA Cardiovascular Science, University of Liege, Liege, Belgium
                [3 ]Intensive Care Unit, Cliniques Universitaires St-Luc, Brussels, Belgium
                [4 ]La clinique Notre Dame de Grâce, Gosselies, Belgium
                Article
                1475-925X-12-61
                10.1186/1475-925X-12-61
                3707774
                23819441
                32ad641b-6425-41e2-a744-e7f85a496b5a
                Copyright © 2013 Chiew et al.; licensee BioMed Central Ltd.

                This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 25 April 2013
                : 18 June 2013
                Categories
                Research

                Biomedical engineering
                mechanical ventilation,nava,non-invasive ventilation,patient-ventilator interaction,matching

                Comments

                Comment on this article