      Long term complications in single and dual chamber pacing are influenced by surgical experience and patient morbidity.

      Heart
      Age Factors, Aged, Aged, 80 and over, Cardiac Pacing, Artificial, adverse effects, Clinical Competence, Equipment Failure, Female, Fluoroscopy, Humans, Intraoperative Period, Male, Middle Aged, Pacemaker, Artificial, Postoperative Complications, Reoperation, Retrospective Studies, Risk Factors, Ventricular Dysfunction, Left, complications

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          To determine how short and long term complication rates after pacemaker implantation are influenced by patient morbidity, operator experience, and choice of pacing system. Retrospective analysis of 1884 patients who received VVI (n = 610), VDD (n = 371), or DDD devices (n = 903) between 1990 and 2001. Follow up period was 64 (34) months. The influence of age, sex, coronary artery disease, myocardial infarction, reduced left ventricular (LV) function, right ventricular (RV) dilatation, atrial fibrillation, device type, and operator experience on operation time and complication rate were analysed. Operation time was prolonged in patients with coronary artery disease, inferior myocardial infarction, reduced LV function, and RV dilatation. Implantation of DDD pacemakers prolonged operation time, particularly among operators with a low or medium level of experience. The overall complication rate was 4.5%. Sixty seven per cent of these complications occurred within the first three months. Complication rate was increased by age, reduced LV function, and RV dilatation. Implantation of DDD systems led to a higher complication rate (6.3%) than implantation of VVI (2.6%) or VDD pacemakers (3.2%). These differences were present only among operators with a low or medium level of experience. Operation time and complication rate increased with age, impaired LV function, and RV dilatation. Complication rates were higher with DDD than with VVI or VDD implantation and were excessive among inexperienced but not experienced operators.

          Related collections

          Author and article information

          Comments

          Comment on this article

          scite_

          Similar content138

          Cited by28