      How Does Higher Ultrafiltration within the Conventional Clinical Range Impact the Volume Status of Hemodialysis Patients?

      research-article

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Aims: The higher ultrafiltration (UF) induces poor outcomes. The impact of higher UF on the volume status was investigated. Methods: 60 hemodialysis (HD) patients were divided into three groups according to the ratio of total UF to post-dialysis body weight (TUF/PDW) (<3%, 3–5%, ≥5%). ANP, the ratio of extracellular water to total body water and excess fluid mass (ExF/PDW) by bioimpedance spectroscopy, inferior vena cava diameter by ultrasound were measured at the end of HD. The ratio of post-HD blood volume to pre-HD (BVpost/BVpre) and standardized filtration coefficients (Lpst) of the microvasculature in the vicinity of PDW were calculated. Results: Only Lpst and BVpost/BVpre showed significant differences among the three groups. A stepwise multiple linear regression model revealed that BVpost/BVpre was correlated with TUF/PDW, ExF/PDW and Lpst (R = 0.778, p < 0.001), independently. Conclusion: Higher UF causes decreases in BVpost/BVpre and Lpst. BVpost/BVpre was determined by TUF/PDW, ExF/PDW and Lpst.

          Related collections

          Most cited references14

          • Record: found
          • Abstract: found
          • Article: not found

          Exploring the viability of using online social media advertising as a recruitment method for smoking cessation clinical trials.

          The aim of the present study was to explore the viability of using social media as a recruitment tool in a clinical research trial. Sociodemographic data and smoking characteristics were assessed in 266 participants recruited to investigate the effectiveness of a behavioral support program for smoking cessation. For analysis, participants were separated into 2 groups based on whether they were recruited either using traditional means (flyers, word of mouth, or newspaper advertisement; n = 125, 47.0%) or by advertisements in online social media (n = 138, 51.9%). Participants recruited via social media were significantly younger, but there were no differences in other socioeconomic variables or smoking characteristics compared with participants recruited via other traditional means. The findings of the present study suggest that using online social media is a viable recruitment method for smoking studies and compliments other more traditional recruitment methods.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Dialysis hypotension: a hemodynamic analysis.

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Relationship between volume status and blood pressure during chronic hemodialysis.

              The relationship between volume status and blood pressure (BP) in chronic hemodialysis (HD) patients remains incompletely understood. Specifically, the effect of interdialytic fluid accumulation (or intradialytic fluid removal) on BP is controversial. We determined the association of the intradialytic decrease in body weight (as an indicator of interdialytic fluid gain) and the intradialytic decrease in plasma volume (as an indicator of postdialysis volume status) with predialysis and postdialysis BP in a cross-sectional analysis of a subset of patients (N=468) from the Hemodialysis (HEMO) Study. Fifty-five percent of patients were female, 62% were black, 43% were diabetic and 72% were prescribed antihypertensive medications. Dry weight was defined as the postdialysis body weight below which the patient developed symptomatic hypotension or muscle cramps in the absence of edema. The intradialytic decrease in plasma volume was calculated from predialysis and postdialysis total plasma protein concentrations and was expressed as a percentage of the plasma volume at the beginning of HD. Predialysis systolic and diastolic BP values were 153.1 +/- 24.7 (mean +/- SD) and 81.7 +/- 14.8 mm Hg, respectively; postdialysis systolic and diastolic BP values were 136.6 +/- 22.7 and 73.9 +/- 13.6 mm Hg, respectively. As a result of HD, body weight was reduced by 3.1 +/- 1.3 kg and plasma volume was contracted by 10.1 +/- 9.5%. Multiple linear regression analyses showed that each kg reduction in body weight during HD was associated with a 2.95 mm Hg (P=0.004) and a 1.65 mm Hg (P=NS) higher predialysis and postdialysis systolic BP, respectively. In contrast, each 5% greater contraction of plasma volume during HD was associated with a 1.50 mm Hg (P=0.026) and a 2.56 mm Hg (P < 0.001) lower predialysis and postdialysis systolic BP, respectively. The effects of intradialytic decreases in body weight and plasma volume were greater on systolic BP than on diastolic BP. HD treatment generally reduces BP, and these reductions in BP are associated with intradialytic decreases in both body weight and plasma volume. The absolute predialysis and postdialysis BP levels are influenced differently by acute intradialytic decreases in body weight and acute intradialytic decreases in plasma volume; these parameters provide different information regarding volume status and may be dissociated from each other. Therefore, evaluation of volume status in chronic HD patients requires, at minimum, assessments of both interdialytic fluid accumulation (or the intradialytic decrease in body weight) and postdialysis volume overload.
                Bookmark

                Author and article information

                Journal
                BPU
                Blood Purif
                10.1159/issn.0253-5068
                Blood Purification
                S. Karger AG
                0253-5068
                1421-9735
                2009
                March 2009
                14 February 2009
                : 27
                : 3
                : 253-260
                Affiliations
                aDivision of Nephrology, Kyoto City Hospital, Kyoto, and bDivision of Nephrology, Kitano Hospital, Medical Institute of Tazuke Kofukai, Kyoto/Osaka, Japan
                Article
                202004 Blood Purif 2009;27:253–260
                10.1159/000202004
                19218793
                34bc0d94-c998-484a-94ce-e013f8897fef
                © 2009 S. Karger AG, Basel

                Copyright: All rights reserved. No part of this publication may be translated into other languages, reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying, recording, microcopying, or by any information storage and retrieval system, without permission in writing from the publisher. Drug Dosage: The authors and the publisher have exerted every effort to ensure that drug selection and dosage set forth in this text are in accord with current recommendations and practice at the time of publication. However, in view of ongoing research, changes in government regulations, and the constant flow of information relating to drug therapy and drug reactions, the reader is urged to check the package insert for each drug for any changes in indications and dosage and for added warnings and precautions. This is particularly important when the recommended agent is a new and/or infrequently employed drug. Disclaimer: The statements, opinions and data contained in this publication are solely those of the individual authors and contributors and not of the publishers and the editor(s). The appearance of advertisements or/and product references in the publication is not a warranty, endorsement, or approval of the products or services advertised or of their effectiveness, quality or safety. The publisher and the editor(s) disclaim responsibility for any injury to persons or property resulting from any ideas, methods, instructions or products referred to in the content or advertisements.

                History
                : 18 August 2008
                : 19 November 2008
                Page count
                Figures: 1, Tables: 6, References: 25, Pages: 8
                Categories
                Original Paper

                Cardiovascular Medicine,Nephrology
                Atrial natriuretic peptide,Bioimpedance spectroscopy,Hemodialysis,Vena cava diameter in quiet expiration,Plasma refilling,Ultrafiltration

                Comments

                Comment on this article