      Systematic review and meta-analysis of method comparison studies of Masimo pulse co-oximeters (Radical-7™ or Pronto-7™) and HemoCue® absorption spectrometers (B-Hemoglobin or 201+) with laboratory haemoglobin estimation.

      1 , 1 , 2
      Anaesthesia and intensive care
      haemoglobin, oximetry

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          We assessed agreement in haemoglobin measurement between Masimo pulse co-oximeters (Rad-7™ and Pronto-7™) and HemoCue® photometers (201+ or B-Hemoglobin) with laboratory-based determination and identified 39 relevant studies (2915 patients in Masimo group and 3084 patients in HemoCue group). In the Masimo group, the overall mean difference was -0.03 g/dl (95% prediction interval -0.30 to 0.23) and 95% limits of agreement -3.0 to 2.9 g/dl compared to 0.08 g/dl (95% prediction interval -0.04 to 0.20) and 95% limits of agreement -1.3 to 1.4 g/dl in the HemoCue group. Only B-Hemoglobin exhibited bias (0.53, 95% prediction interval 0.27 to 0.78). The overall standard deviation of difference was larger (1.42 g/dl versus 0.64 g/dl) for Masimo pulse co-oximeters compared to HemoCue photometers. Masimo devices and HemoCue 201+ both provide an unbiased, pooled estimate of laboratory haemoglobin. However, Masimo devices have lower precision and wider 95% limits of agreement than HemoCue devices. Clinicians should carefully consider these limits of agreement before basing transfusion or other clinical decisions on these point-of-care measurements alone.

          Related collections

          Author and article information

          Journal
          Anaesth Intensive Care
          Anaesthesia and intensive care
          0310-057X
          0310-057X
          May 2015
          : 43
          : 3
          Affiliations
          [1 ] Department of Anaesthesia, Mercy Hospital for Women, Heidelberg, Victoria.
          [2 ] Department of Anaesthesia, Mercy Hospital for Women, Heidelberg and Department of Pharmacology, University of Melbourne, Victoria.
          Article
          20140520
          10.1177/0310057X1504300310
          25943608
          35e2ce18-f596-468d-887d-aa586a1c6994
          History

          haemoglobin,oximetry
          haemoglobin, oximetry

          Comments

          Comment on this article