Introduction
KEY TEACHING POINTS
Key Teaching Points
•
Dissimilar ventricular rhythms have been observed in cardiac resynchronization implantable
cardioverter-defibrillators (ICDs) in the presence of a left ventricular lead. The
same arrhythmic phenomenon may also occur in patients with a single- or dual-chamber
ICD in the absence of a left ventricular lead.
•
Electrograms (EGMs) recorded from pace/sense electrodes (or “near-field” EGMs) show
bipolar signals from the local area in the near vicinity of the electrode position.
In contrast, the far-field EGM recording from the shocking electrodes (right ventricular
coil vs can), integrating a much larger area of myocardium, provide a more global
visualization of electrical activity, which includes both right and left ventricular
deflections. The far-field EGM becomes of special importance and is of clinical relevance
for the diagnosis of dissimilar ventricular rhythms even if the far-field channel
is not used for automatic detection.
•
Right ventricular rates slower than the left ventricular rate may lead to nondetection
and lack of ICD therapy during a life-threatening ventricular tachycardia. Repositioning
of the ICD lead and a renewed ventricular fibrillation induction may be helpful measures.
Implantable cardioverter-defibrillators (ICDs) represent the treatment of choice for
the prevention of sudden cardiac death. Low signal amplitudes are occasionally encountered
during placement of ICDs. ICD implant guidelines state that the amplitude of the sinus
rhythm R-wave recorded from the ventricular electrogram should be higher than 5 mV.
If no sufficient R signal amplitude can be found despite several attempts of repositioning
the ICD lead, an induction of ventricular fibrillation (VF) is useful to ensure that
the ICD will sense, detect, and defibrillate VF.
Case report
A 51-year-old woman with prior anterior myocardial infarction, apical and septal dyskinesis,
a left ventricular ejection fraction of 26%, and NYHA class III symptoms received
an ICD device (Itrevia 7 VR-T DX; Biotronik SE & Co KG, Berlin, Germany).
1
The implanted ICD lead (Linox Smart ProMRI DX S 65/15; Biotronik SE & Co KG) has a
pentapolar design, which also enables atrial sensing via the 2 floating atrial sensing
rings.
2
Lead positioning was difficult, as no sufficient R-wave amplitude could be measured.
Several repositionings of the ICD lead were tried in the right ventricular (RV) apex
and on the septal wall. Maximum signal amplitudes between 2.9 and 3.6 mV were found
in a low septal position, as shown in Figure 1. Induction of VF was performed by shock-on-T
to check the VF sensitivity in the presence of very low signal amplitudes. VF was
clearly displayed on the electrocardiogram monitor, but no automatic VF detection
could be fulfilled. Since no automatic ICD shock is to be expected under these circumstances,
a manually initiated ICD emergency shock at 40 joules was delivered in order to terminate
VF. The manually recorded real-time electrogram (EGM) is shown in Figure 2.
Two major questions may arise:
1.
Is the missed VF detection caused by too-small signal amplitudes?
2.
Why is the tachycardia rate faster in the far-field EGM (FF-EGM) than in the right
ventricle?
Discussion
The RV signal amplitude during sinus rhythm was indeed very low (2.9–3.6 mV) and was
thus below the minimum requirement of at least 5 mV. Signal amplitudes became even
less after VF induction. In Figure 2A there is definitely ventricular undersensing.
Comparing the VF signals with the calibration bar in front of the EGM tracing indicates
that signal amplitudes are probably below 0.8 mV, the default minimum sensing threshold
of the device. There are no ventricular sense markers in Figure 2A. Thus, the lower
rate interval can time out, and a ventricular pace event is released after 1500 ms
(VVI 40). VF detection becomes impossible owing to subthreshold signal amplitudes.
In Figure 2B, the sensitivity of the ICD becomes sufficient to detect all intraventricular
signals. Interestingly, the rate of tachycardia in the FF-EGM appears to be faster
than in the RV channel. The FF channel suggests ventricular fibrillation. The intraventricular
intervals, however, are in a range between 294 and 368 ms and are therefore longer
than the temporarily programmed VF intervention interval of 270 ms for defibrillation
threshold testing. In the present case we can only speculate because a single-chamber
ICD does not provide left ventricular (LV) electrograms, as would be available in
an ICD with resynchronization therapy (cardiac resynchronization therapy defibrillator;
CRT-D). The observations of dissimilar ventricular tachycardia (VT) rhythms (dissociated
RV and LV rhythms) and the implications for ICD detection were described by Barold
et al.
3
Our observation may be due to the presence of dissimilar ventricular rhythms, recorded
by a single-chamber ICD. The fast VF rhythm shown in the FF channel can be explained
if we assume that there was VF in the left ventricle and a VT in the right ventricle.
The exact reason for dissimilar ventricular rhythms is still unclear. Interventricular
conduction block may be invoked to explain the long RR intervals in the RV-EGM in
the presence of a faster rate in the FF-EGM. The RV lead tip may have been partially
protected by local scar tissue (entrance block), whereas the fast VT involved the
LV and most of the RV apart from the RV electrode tip. The relatively regular shorter-cycle-length
atrial EGMs as recorded by the floating electrodes are probably due to retrograde
atrial activation from the slower VT in the right ventricle.
After repositioning of the lead in another low septal area, a signal amplitude ranging
from 4.9 to 7.1 mV was obtained. A renewed VF induction was performed 2 days after
the implantation procedure with a programmed minimum sensing threshold of 0.8 mV.
The second VF sensing test was successful and VF detection could be fulfilled after
8/12 without undetected ventricular events.
In summary, the situation that led to an undetected VF may be 2-fold. Firstly, low-amplitude
and subthreshold VF signals resulted in undersensing. Secondly, the RV rate was too
slow and far below the programmed VF zone owing to dissimilar ventricular rhythms,
as presented in this case.
Similar observations were already made by Josephson
4
and described as atypical VF. The surface electrocardiogram presented classic VF;
however, the bulk of the heart demonstrated regular discrete ventricular activity.
Fragmented, asynchronous activity was only seen in the area of the large infarction
in the left ventricle.
The present case also bears similarities to a well-documented case report by Yunoki
et al.
5
In that report, both the FF-EGM and the external defibrillator recordings present
VF-like electrical activity. This VF, however, remained undetected because RR intervals
on the RV-EGM stayed prolonged at 400 ms (150 beats/minute) for the duration of the
episode, well below the programmed tachycardia intervention rate.
Other published case reports described failure of VF detection because of too-long
RR intervals but in the presence of VF-like signal morphologies in the FF-EGM, despite
large R signal amplitudes during sinus rhythm.6, 7, 8 The possible presence of “dissimilar
VT rhythms” was, however, not taken into account and discussed in these articles.
The availability of an LV-EGM with CRT-D devices renders the diagnosis of dissimilar
VT rhythms less problematic, as is the case with single- or dual-chamber ICDs.
Clinical implications and conclusion
Dissimilar ventricular rhythms during VT have been observed in both directions: LV
rate faster than RV, and RV rate faster than LV. As long as the RV delivers the shortest
intervals, ICD detection and therapy delivery is guaranteed. ICD detection may fail
under worst-case conditions if the LV rate is faster than the RV. Both repositioning
of the ICD lead and reinduction of VF might be appropriate measures. If the unit is
not equipped with a DF-4 lead, interchanging the RV and LV lead plugs could be an
option in a CRT-D device to permit VF detection from the left side. The presence of
dissimilar ventricular rhythms in a single- or dual-chamber ICD remains a critical
issue that could possibly be solved by introduction of a suitable hemodynamic sensor
in future ICDs.
A VF induction has luckily been performed, which is no longer a standard procedure.
Without induced VF, dissimilar rhythms would never have been discovered as the root
cause. Understanding the origin of underdetection is of the utmost importance in order
to avoid dramatic consequences at the occurrence of spontaneous tachyarrhythmias.