9
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: not found

      Rapid Slow Off-Rate Modified Aptamer (SOMAmer)-Based Detection of C-Reactive Protein Using Isotachophoresis and an Ionic Spacer.

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          We present an on-chip electrophoretic assay for rapid protein detection with a SOMAmer (Slow Off-Rate Modified Aptamer) reagent. We used isotachophoresis (ITP) coupled with an ionic spacer to both react and separate SOMAmer-protein complex from free SOMAmer reagent. ITP accelerates the reaction kinetics as the ionic spacer concurrently separates the reaction products. We developed a numerical and analytical model to describe ITP spacer assays, which involve low-mobility, nonfocusing targets that are recruited into the ITP zone by higher-mobility, ITP-focused probes. We demonstrated a proof-of-concept of this assay using C-reactive protein (CRP) in buffer, and achieved a 2 nM limit of detection (LOD) with a combined 20 min assay time (10 min off-chip preparation of reagents and 10 min on-chip run). Our findings suggest that this approach has potential as a simple and rapid alternative to other homogeneous immunoassays. We also explore the extension of this assay to a diluted serum sample spiked with CRP, where we observe decreased sensitivity (an LOD of 25 nM in 20-fold diluted serum). We describe the challenges in extending this assay to complex samples and achieving higher sensitivity and specificity for clinical applications.

          Related collections

          Author and article information

          Journal
          Anal. Chem.
          Analytical chemistry
          American Chemical Society (ACS)
          1520-6882
          0003-2700
          Jul 07 2015
          : 87
          : 13
          Affiliations
          [1 ] †Department of Mechanical Engineering, Stanford University, Stanford, California 94305, United States.
          [2 ] §SomaLogic, Inc., Boulder, Colorado 80301, United States.
          Article
          10.1021/acs.analchem.5b00886
          26024067
          3f559adf-b84c-425b-94fb-c8ba376cda1b
          History

          Comments