42
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Aims

          Complications after cardiac implantable electronic device (CIED) treatment, including permanent pacemakers (PMs), cardiac resynchronization therapy devices with defibrillators (CRT-Ds) or without (CRT-Ps), and implantable cardioverter defibrillators (ICDs), are associated with increased patient morbidity, healthcare costs, and possibly increased mortality.

          Methods and results

          Population-based cohort study in all Danish patients who underwent a CIED procedure from May 2010 to April 2011. Data on complications were gathered on review of all patient charts while baseline data were obtained from the Danish Pacemaker and ICD Register. Adjusted risk ratios (aRRs) with 95% confidence intervals were estimated using binary regression. The study population consisted of 5918 consecutive patients. A total of 562 patients (9.5%) experienced at least one complication. The risk of any complication was higher if the patient was a female (aRR 1.3; 1.1–1.6), underweight (aRR 1.5; 1.1–2.3), implanted in a centre with an annual volume <750 procedures (0–249 procedures: aRR 1.6; 1.1–2.2, 250–499: aRR 2.0; 1.6–2.7, 500–749: aRR 1.5; 1.2–1.8), received a dual-chamber ICD (aRR 2.0; 1.4–2.7) or CRT-D (aRR 2.6; 1.9–3.4), underwent system upgrade or lead revision (aRR 1.3; 1.0–1.7), had an operator with an annual volume <50 procedures (aRR 1.9; 1.4–2.6), or underwent an emergency, out-of-hours procedure (aRR 1.5; 1.0–2.3).

          Conclusion

          CIED complications are more frequent than generally acknowledged. Both patient- and procedure-related predictors may identify patients with a particularly high risk of complications. This information should be taken into account both in individual patient treatment and in the planning of future organization of CIED treatment.

          Related collections

          Most cited references33

          • Record: found
          • Abstract: not found
          • Article: not found

          2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA).

            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Obesity: preventing and managing the global epidemic

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study.

              Today quantitative information about the type of complications and their incidence during long-term pacemaker (PM) follow-up is scarce. To assess the incidence and determinants of short- and long-term complications after first pacemaker implantation for bradycardia. A prospective multicenter cohort study (the FOLLOWPACE study) was conducted among 1517 patients receiving a PM between January 2003 and November 2007. The independent association of patient and implantation-procedure characteristics with the incidence of PM complications was analyzed using multivariable Cox regression analysis. A total of 1517 patients in 23 Dutch PM centers were followed for a mean of 5.8 years (SD 1.1), resulting in 8797 patient-years. Within 2 months, 188 (12.4%) patients developed PM complications. Male gender, age at implantation, body mass index, a history of cerebrovascular accident, congestive heart failure, use of anticoagulant drugs, and passive atrial lead fixation were independent predictors for complications within 2 months, yielding a C-index of 0.62 (95% confidence interval 0.57-0.66). Annual hospital implanting volume did not additionally contribute to the prediction of short-term complications. Thereafter, 140 (9.2%) patients experienced complications, mostly lead-related complications (n = 84). Independent predictors for long-term complications were age, body mass index, hypertension, and a dual-chamber device, yielding a C-index of 0.62 (95% confidence interval 0.57-0.67). The occurrence of a short-term PM complication was not predictive of future PM complications. Complication incidence in modern pacing therapy is still substantial. Most complications occur early after PM implantation. Although various patient- and procedure-related characteristics are independent predictors for early and late complications, their ability to identify the patient at high risk is rather poor. This relatively high incidence of PM complications and their poor prediction underscores the usefulness of current guidelines for regular follow-up of patients with PM. Copyright © 2012 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.
                Bookmark

                Author and article information

                Journal
                Eur Heart J
                Eur. Heart J
                eurheartj
                ehj
                European Heart Journal
                Oxford University Press
                0195-668X
                1522-9645
                7 May 2014
                17 December 2013
                17 December 2013
                : 35
                : 18
                : 1186-1194
                Affiliations
                [1 ]Department of Cardiology, Aarhus University Hospital , Skejby, Denmark
                [2 ]The Danish Pacemaker and ICD Register, Odense University Hospital , Odense, Denmark
                [3 ]Department of Cardiology, Odense University Hospital , Odense, Denmark
                [4 ]Department of Public Health, Section for Epidemiology, Aarhus University , Aarhus, Denmark
                [5 ]Department of Heart, Lung, and Vascular Surgery, Odense University Hospital , Odense, Denmark
                Author notes
                [* ]Corresponding author. Tel: +45 7845 2259, Fax: +45 7845 2260, Email: reki@ 123456svf.au.dk
                Article
                eht511
                10.1093/eurheartj/eht511
                4012708
                24347317
                3fdf2e5d-22cc-44d1-9eff-70d7ec1ec836
                © The Author 2013. Published by Oxford University Press on behalf of the European Society of Cardiology.

                This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License ( http://creativecommons.org/licenses/by-nc/3.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@ 123456oup.com

                History
                : 23 May 2013
                : 11 November 2013
                : 21 November 2013
                Categories
                Clinical Research
                Arrhythmia/Electrophysiology
                Editor's choice

                Cardiovascular Medicine
                complication,predictors,cardiac resynchronization therapy,implantable cardioverter defibrillator,pacemaker

                Comments