      Impact of infusion line compliance on syringe pump performance.

      1 , , ,
      Paediatric anaesthesia

      Read this article at

      ScienceOpenPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          To determine the impact of infusion line compliance on the time to occlusion alarm (occlusion time) and the size of the subsequent occlusion release bolus, we studied three different infusion lines in combination with a 50-ml and 10-ml syringe. The mean occlusion time of the 50-ml syringe amounted to 20.3 +/- 0.28 min and increased to 26.2 +/- 0.19 min with the infusion with the infusion line yeielding the largest compliance (P<0.0001). The mean occlusion time of the 10-ml syringe was 2.5 +/- 0.75 min and showed a 2.7-fold variation with the three infusion lines (2.6+/-0.08, 5.6 +/- 0.28 and 6.9 +/- 0.35 min, respectively; P<0.0001). Occlusion times correlated strongly to be calculated estimated compliance of the syringe-infusion line assembly (Pearson's r=0.998, P<0.0001). Infusion line compliance has an important impact on time to occlusion, particularly when using small, low compliant infusion pump syringes.

          Related collections

          Author and article information

          Journal
          Paediatr Anaesth
          Paediatric anaesthesia
          1155-5645
          1155-5645
          2000
          : 10
          : 6
          Affiliations
          [1 ] Department of Anaestesia, University Children's Hospital of Zurich, Switzerland.
          Article
          11186873
          41536cfb-6f23-442e-9925-b451eac499ca
          History

          Comments

          Comment on this article