      Correlation between detector array measurements and a computer algorithm for enhanced dynamic wedge profiles.

      Medical dosimetry : official journal of the American Association of Medical Dosimetrists
      Algorithms, Humans, Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted, instrumentation

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Verification of dosimetric data computed by a treatment planning system is necessary in the commissioning process for any clinical software, just as it is necessary for any annual quality assurance testing. Direct verification of the dosimetric data is achievable when calculating the enhanced dynamic wedge (EDW) off-axis ratio at each point of interest. As mathematical models for hand-calculating such factors are still evolving, measurement of these external beam arrangements has proven significantly more accurate. This research presents a correlation for measured and planned data, specific for 6- and 18-MV photon beams on a Varian 21EX linear accelerator, using the EDW mode. Field generation was created using the Varian Eclipse treatment planning system. On treatment field delivery, the Sun Nuclear MapCHECK diode array was used to plot each beam profile in 2 dimensions. Wedge angles of 10 degrees , 15 degrees , 20 degrees , 25 degrees , 30 degrees , 45 degrees and 60 degrees were studied here, under isocentric geometry, at a fixed water equivalent depth of 15 cm. Field size dependence was considered with each wedge and energy combination, where symmetric apertures of 5 x 5 cm(2), 10 x 10 cm(2), 15 x 15 cm(2) and 20 x 20 cm(2) were used. Accurate dosimetric results were found to be achievable when using this treatment planning system to within 2.8% maximum deviation, and to within 1% deviation averaged over all. The diode array also proved to be simple and ideally suited for EDW measurements.

          Related collections

          Author and article information

          Journal
          17707202
          10.1016/j.meddos.2007.02.001

          Chemistry
          Algorithms,Humans,Radiotherapy Dosage,Radiotherapy Planning, Computer-Assisted,instrumentation

          Comments

          Comment on this article