      Diagnostic performance of the loop-mediated isothermal amplification (LAMP) based illumigene ® malaria assay in a non-endemic region

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Light microscopy and antigen-based rapid diagnostic tests are the primary diagnostic tools for detecting malaria, although being labour-intensive and frequently challenged by lack of personnel’s experience and low levels of parasite density. The latter being especially important in non-endemic settings. Novel molecular techniques aim to overcome this drawback. The objective of this study was to assess the diagnostic performance of the illumigene malaria assay ® (Meridian Bioscience) compared to microscopy, RDT and real-time PCR. This loop-mediated isothermal amplification (LAMP) assay is a qualitative in vitro diagnostic test for the direct detection of Plasmodium spp. DNA in human venous whole blood samples.

          Methods

          The illumigene assay was assessed on a retrospective panel of stored blood samples (n = 103) from returned travellers and external quality control samples (n = 12). Additionally the assay was prospectively assessed on 30 fresh routine samples with a request for malaria diagnosis. The illumigene assay was compared to microscopy, RDT and Plasmodium species specific real-time PCR.

          Results

          In the retrospective evaluation, the illumigene assay showed 100% agreement with the real-time PCR, RDT and microscopy yielding a sensitivity and specificity of 100% (95% CI 95.1–100% and 89.7–100%, respectively). Seven samples from patients recently treated for Plasmodium falciparum infection that were RDT positive and microscopy negative yielded positive test results. The performance of the illumigene assay equals that of microscopy combined with RDT in the prospective panel with three false negative RDT results and one false negative microscopy result. Excellent concordance with PCR was observed. The limit of detection of the assay approached 0.5 parasites/µL for both P. falciparum and Plasmodium vivax.

          Conclusion

          In non-endemic regions where the diagnostic process for malaria infections is questioned by lack of experience and low levels of parasite densities, the illumigene assay can be of value. Due to its high sensitivity, the LAMP assay may be considered as primary diagnostic test. The results of this study indicate that negative screen results do not need further confirmation. However, before implementation, this approach needs to be confirmed in larger, prospective studies. A shortcoming of this assay is that no species identification nor determination of parasite density are possible.

          Related collections

          Most cited references19

          • Record: found
          • Abstract: found
          • Article: not found

          A review of malaria diagnostic tools: microscopy and rapid diagnostic test (RDT).

          The absolute necessity for rational therapy in the face of rampant drug resistance places increasing importance on the accuracy of malaria diagnosis. Giemsa microscopy and rapid diagnostic tests (RDTs) represent the two diagnostics most likely to have the largest impact on malaria control today. These two methods, each with characteristic strengths and limitations, together represent the best hope for accurate diagnosis as a key component of successful malaria control. This review addresses the quality issues with current malaria diagnostics and presents data from recent rapid diagnostic test trials. Reduction of malaria morbidity and drug resistance intensity plus the associated economic loss of these two factors require urgent scaling up of the quality of parasite-based diagnostic methods. An investment in anti-malarial drug development or malaria vaccine development should be accompanied by a parallel commitment to improve diagnostic tools and their availability to people living in malarious areas.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Update on rapid diagnostic testing for malaria.

            To help mitigate the expanding global impact of malaria, with its associated increasing drug resistance, implementation of prompt and accurate diagnosis is needed. Malaria is diagnosed predominantly by using clinical criteria, with microscopy as the current gold standard for detecting parasitemia, even though it is clearly inadequate in many health care settings. Rapid diagnostic tests (RDTs) have been recognized as an ideal method for diagnosing infectious diseases, including malaria, in recent years. There have been a number of RDTs developed and evaluated widely for malaria diagnosis, but a number of issues related to these products have arisen. This review highlights RDTs, including challenges in assessing their performance, internationally available RDTs, their effectiveness in various health care settings, and the selection of RDTs for different health care systems.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Detection of four Plasmodium species by genus- and species-specific loop-mediated isothermal amplification for clinical diagnosis.

              Loop-mediated isothermal amplification (LAMP), a novel nucleic acid amplification method, was developed for the clinical detection of four species of human malaria parasites: Plasmodium falciparum, P. vivax, P. malariae, and P. ovale. We evaluated the sensitivity and specificity of LAMP in comparison with the results of microscopic examination and nested PCR. LAMP showed a detection limit (analytical sensitivity) of 10 copies of the target 18S rRNA genes for P. malariae and P. ovale and 100 copies for the genus Plasmodium, P. falciparum, and P. vivax. LAMP detected malaria parasites in 67 of 68 microscopically positive blood samples (sensitivity, 98.5%) and 3 of 53 microscopically negative samples (specificity, 94.3%), in good agreement with the results of nested PCR. The LAMP reactions yielded results within about 26 min, on average, for detection of the genus Plasmodium, 32 min for P. falciparum, 31 min for P. vivax, 35 min for P. malariae, and 36 min for P. ovale. Accordingly, in comparison to the results obtained by microscopy, LAMP had a similar sensitivity and a greater specificity and LAMP yielded results similar to those of nested PCR in a shorter turnaround time. Because it can be performed with a simple technology, i.e., with heat-treated blood as the template, reaction in a water bath, and inspection of the results by the naked eye because of the use of a fluorescent dye, LAMP may provide a simple and reliable test for routine screening for malaria parasites in both clinical laboratories and malaria clinics in areas where malaria is endemic.
                Bookmark

                Author and article information

                Contributors
                anne-sophie.dekoninck@medina.be
                lcnops@itg.be
                mattias.hofmans@ugent.be
                jjacobs@itg.be
                dvandenbossche@itg.be
                +32 93323408 , jan.philippe@uzgent.be
                Journal
                Malar J
                Malar. J
                Malaria Journal
                BioMed Central (London )
                1475-2875
                17 October 2017
                17 October 2017
                2017
                : 16
                : 418
                Affiliations
                [1 ]ISNI 0000 0004 0626 3303, GRID grid.410566.0, Department of Laboratory Medicine, , Ghent University Hospital (GUH), ; De Pintelaan 185, 9000 Ghent, Belgium
                [2 ]ISNI 0000 0001 2153 5088, GRID grid.11505.30, Institute of Tropical Medicine (ITM) Antwerp, ; Nationalestraat 155, 2000 Antwerp, Belgium
                Article
                2065
                10.1186/s12936-017-2065-8
                5645927
                29041927
                449c2f78-0cc5-44ec-baa6-b5ece8198fcd
                © The Author(s) 2017

                Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License ( http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

                History
                : 27 March 2017
                : 13 October 2017
                Categories
                Research
                Custom metadata
                © The Author(s) 2017

                Infectious disease & Microbiology
                illumigene malaria assay,lamp assay,plasmodium,loop-mediated isothermal amplification,malaria

                Comments

                Comment on this article