      A Vertically Integrated CMOS Microsystem for Time-Resolved Fluorescence Analysis.

      IEEE transactions on biomedical circuits and systems

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          We describe a two-chip micro-scale time-resolved fluorescence analyzer integrating excitation, detection, and filtering. A new 8×8 array of drivers realized in standard low-voltage 0.35-μm complementary metal-oxide semiconductor is bump-bonded to AlInGaN blue micro-pixellated light-emitting diodes (micro-LEDs). The array is capable of producing sample excitation pulses with a width of 777 ps (FWHM), enabling short lifetime fluorophores to be investigated. The fluorescence emission is detected by a second, vertically-opposed 16 × 4 array of single-photon avalanche diodes (SPADs) fabricated in 0.35-μm high-voltage CMOS technology with in-pixel time-gated photon counting circuitry. Captured chip data are transferred to a PC for further processing, including histogramming, lifetime extraction, calibration and background/noise compensation. This constitutes the smallest reported solid-state microsystem for fluorescence decay analysis, replacing lasers, photomultiplier tubes, bulk optics, and discrete electronics. The system is demonstrated with measurements of fluorescent colloidal quantum dot and Rhodamine samples.

          Related collections

          Author and article information

          Journal
          23853381
          10.1109/TBCAS.2010.2077290

          Comments

          Comment on this article