          Abstract

          A liquid chromatography/electrospray ionization tandem mass spectrometry (LC/ESI-MS/MS) method has been developed for the analysis of buprenorphine (BUP) and nor-buprenorphine (NBUP) in biological fluids. Analytes are isolated from urine and blood, after addition of d4-buprenorphine (d4-BUP) as internal standard, by solid-phase extraction. Preparation of hair involves external decontamination, mechanical pulverization, overnight incubation in acidic medium, and neutralization prior to extraction. Enzymatic hydrolysis with beta-glucuronidase may be performed to distinguish between free and total BUP. Chromatographic separation is accomplished by gradient elution on a cyanopropyl 2.1 x 150 mm column. Positive ion ESI and MS analyses are carried out in an ion trap mass spectrometer. The use of this mass analyzer allows effective collisional experiments to be performed on ESI-generated MH+ species. Abundant product ions are produced, which can be monitored together with precursor ions without losing sensitivity. Thus, assay selectivity is definitely increased with respect to LC/ESI-MS/MS methods in which only precursor ions are monitored. The method has good linearity (calibration curves were linear in the range 0.1-10 ng/mL in urine and blood, in the range 10-160 pg/mg in hair) and limits of detection of 0.05 ng/mL for both BUP and NBUP in blood and urine samples, of 4 pg/mg for both analytes in hair. Both intra- and inter-assay precision and accuracy were satisfactory at three concentrations studied: relative standard deviations were <13.7% in urine, <17.3% in blood, <17.8% in hair; percent deviation of the mean from the true value was always <10.5% in urine and blood, <16.1% in hair. The method can be used to determine both analytes in the urine and hair of drug addicts on replacement therapy, and in post-mortem blood specimens when there is suspicion of drug-related death.

          Related collections

          Author and article information

          Journal
          Rapid Commun. Mass Spectrom.
          Rapid communications in mass spectrometry : RCM
          Wiley
          0951-4198
          0951-4198
          2006
          : 20
          : 8
          Affiliations
          [1 ] Forensic Toxicology and Antidoping, University Hospital of Padova, Via Falloppio 50, I-35121 Padova, Italy.
          Article
          10.1002/rcm.2444
          16550495
          48f7dd2e-d6b8-446e-9bcb-fc29fc9bdb67
          History

          Comments

          Comment on this article