      Low energy peptide fragmentations in an ESI-Q-Tof type mass spectrometer.

      Journal of Proteome Research
      Amino Acid Sequence, Dissociative Disorders, Molecular Sequence Data, Molecular Weight, Peptide Fragments, chemistry, Peptide Mapping, methods, Sequence Analysis, Protein, Serine, analysis, Spectrometry, Mass, Electrospray Ionization, Threonine

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Efficient peptide sequencing relies on both high quality MS/MS data acquisition and exhaustive knowledge of gas-phase dissociation mechanisms. We report our contribution to the elaboration of more comprehensive fragmentation models required for efficient automated MS/MS spectra interpretation. Following a statistical approach, various peptides (296 sequences of variable compositions and lengths) were prepared and subjected to low-energy collision-induced dissociations (CID) in an electrospray hybrid instrument (ESI-Q-q-Tof type mass spectrometer) that has retained relatively limited attention so far. Besides, our studies were focused on low molecular weight singly charged peptides that often failed to be identified by sequencing algorithms. Only half of the studied compounds showed charge directed dissociations in accordance with the mobile proton model producing fragment ions directly related to the primary sequence. For the peptides that did not exhibit the expected fragment ion series, alternative dissociation behaviors issued from complex rearrangements were evidenced.

          Related collections

          Author and article information

          Comments

          Comment on this article