      A prolonged run-in period of standard subcutaneous microdialysis ameliorates quality of interstitial glucose signal in patients after major cardiac surgery

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          We evaluated a standard subcutaneous microdialysis technique for glucose monitoring in two critically ill patient populations and tested whether a prolonged run-in period improves the quality of the interstitial glucose signal. 20 surgical patients after major cardiac surgery (APACHE II score: 10.1 ± 3.2) and 10 medical patients with severe sepsis (APACHE II score: 31.1 ± 4.3) were included in this investigation. A microdialysis catheter was inserted in the subcutaneous adipose tissue of the abdominal region. Interstitial fluid and arterial blood were sampled in hourly intervals to analyse glucose concentrations. Subcutaneous adipose tissue glucose was prospectively calibrated to reference arterial blood either at hour 1 or at hour 6. Median absolute relative difference of glucose (MARD), calibrated at hour 6 (6.2 (2.6; 12.4) %) versus hour 1 (9.9 (4.2; 17.9) %) after catheter insertion indicated a significant improvement in signal quality in patients after major cardiac surgery (p < 0.001). Prolonged run-in period revealed no significant improvement in patients with severe sepsis, but the number of extreme deviations from the blood plasma values could be reduced. Improved concurrence of glucose readings via a 6-hour run-in period could only be achieved in patients after major cardiac surgery.

          Related collections

          Most cited references25

          • Record: found
          • Abstract: found
          • Article: not found

          Comparing methods of measurement: why plotting difference against standard method is misleading.

          When comparing a new method of measurement with a standard method, one of the things we want to know is whether the difference between the measurements by the two methods is related to the magnitude of the measurement. A plot of the difference against the standard measurement is sometimes suggested, but this will always appear to show a relation between difference and magnitude when there is none. A plot of the difference against the average of the standard and new measurements is unlikely to mislead in this way. We show this theoretically and by a practical example.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus

            Objective To evaluate the performance of the FreeStyle Libre Flash continuous glucose monitoring (FSL-CGM) system against established central laboratory methods. Research design and methods 20 subjects (8 type 1 diabetes mellitus, 12 type 2 diabetes mellitus) were analyzed. FSL-CGM sensor measurements (inserted in arm and abdomen) were compared with capillary blood glucose results analyzed with StatStrip as semigold standard. The glucose response after a standardized oral glucose load was measured by FSL-CGM and capillary samples analyzed by perchloric acid hexokinase (PCA-HK) method, StatStrip and FSL test strip (FSLC), and a commonly used CGM system (iPro2). Results FSL-CGM arm sensor readings showed 85.5% of paired readings falling within Clarke Error Grid (ISO 15197:2013) zone A when compared with StatStrip. For FSL-CGM abdomen and FSLC, these percentages were 64% and 98%, respectively. The overall correlation of FSL-CGM in the arm and the StatStrip indicates a performance with lower results with the FSL-CGM in the arm than expected based on the StatStrip in the lower glucose ranges, and higher results than expected in the higher ranges. Following a standardized glucose load, a slower rise in glucose level was observed for FSL-CGM arm as compared with PCA-HK, StatStrip, FSLC, and iPro2 during the first 45–60 min after glucose load ingestion. Conclusions Certain matters need attention while using the FSL-CGM in daily life including the observed lower values in the lower ranges, and the underestimation of the effect of a meal on glucose response. These effects of such deviations can partly be overcome by optimizing the available user instructions. Trial registration number TC5348; results.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Continuous glucose monitoring systems for type 1 diabetes mellitus.

              Self-monitoring of blood glucose is essential to optimise glycaemic control in type 1 diabetes mellitus. Continuous glucose monitoring (CGM) systems measure interstitial fluid glucose levels to provide semi-continuous information about glucose levels, which identifies fluctuations that would not have been identified with conventional self-monitoring. Two types of CGM systems can be defined: retrospective systems and real-time systems. Real-time systems continuously provide the actual glucose concentration on a display. Currently, the use of CGM is not common practice and its reimbursement status is a point of debate in many countries. To assess the effects of CGM systems compared to conventional self-monitoring of blood glucose (SMBG) in patients with diabetes mellitus type 1. We searched The Cochrane Library, MEDLINE, EMBASE and CINAHL for the identification of studies. Last search date was June 8, 2011. Randomised controlled trials (RCTs) comparing retrospective or real-time CGM with conventional self-monitoring of blood glucose levels or with another type of CGM system in patients with type 1 diabetes mellitus. Primary outcomes were glycaemic control, e.g. level of glycosylated haemoglobin A1c (HbA1c) and health-related quality of life. Secondary outcomes were adverse events and complications, CGM derived glycaemic control, death and costs. Two authors independently selected the studies, assessed the risk of bias and performed data-extraction. Although there was clinical and methodological heterogeneity between studies an exploratory meta-analysis was performed on those outcomes the authors felt could be pooled without losing clinical merit. The search identified 1366 references. Twenty-two RCTs meeting the inclusion criteria of this review were identified. The results of the meta-analyses (across all age groups) indicate benefit of CGM for patients starting on CGM sensor augmented insulin pump therapy compared to patients using multiple daily injections of insulin (MDI) and standard monitoring blood glucose (SMBG). After six months there was a significant larger decline in HbA1c level for real-time CGM users starting insulin pump therapy compared to patients using MDI and SMBG (mean difference (MD) in change in HbA1c level -0.7%, 95% confidence interval (CI) -0.8% to -0.5%, 2 RCTs, 562 patients, I(2)=84%). The risk of hypoglycaemia was increased for CGM users, but CIs were wide and included unity (4/43 versus 1/35; RR 3.26, 95% CI 0.38 to 27.82 and 21/247 versus 17/248; RR 1.24, 95% CI 0.67 to 2.29). One study reported the occurrence of ketoacidosis from baseline to six months; there was however only one event. Both RCTs were in patients with poorly controlled diabetes.For patients starting with CGM only, the average decline in HbA1c level six months after baseline was also statistically significantly larger for CGM users compared to SMBG users, but much smaller than for patients starting using an insulin pump and CGM at the same time (MD change in HbA1c level -0.2%, 95% CI -0.4% to -0.1%, 6 RCTs, 963 patients, I(2)=55%). On average, there was no significant difference in risk of severe hypoglycaemia or ketoacidosis between CGM and SMBG users. The confidence interval however, was wide and included a decreased as well as an increased risk for CGM users compared to the control group (severe hypoglycaemia: 36/411 versus 33/407; RR 1.02, 95% CI 0.65 to 1.62, 4 RCTs, I(2)=0% and ketoacidosis: 8/411 versus 8/407; RR 0.94, 95% CI 0.36 to 2.40, 4 RCTs, I(2)=0%).Health-related quality of life was reported in five of the 22 studies. In none of these studies a significant difference between CGM and SMBG was found. Diabetes complications, death and costs were not measured.There were no studies in pregnant women with diabetes type 1 and in patients with hypoglycaemia unawareness. There is limited evidence for the effectiveness of real-time continuous glucose monitoring (CGM) use in children, adults and patients with poorly controlled diabetes. The largest improvements in glycaemic control were seen for sensor-augmented insulin pump therapy in patients with poorly controlled diabetes who had not used an insulin pump before. The risk of severe hypoglycaemia or ketoacidosis was not significantly increased for CGM users, but as these events occurred infrequent these results have to be interpreted cautiously.There are indications that higher compliance of wearing the CGM device improves glycosylated haemoglobin A1c level (HbA1c) to a larger extent. 
                Bookmark

                Author and article information

                Contributors
                julia.mader@medunigraz.at
                Journal
                Sci Rep
                Sci Rep
                Scientific Reports
                Nature Publishing Group UK (London )
                2045-2322
                19 January 2018
                19 January 2018
                2018
                : 8
                : 1262
                Affiliations
                [1 ]ISNI 0000 0000 8988 2476, GRID grid.11598.34, Division of Endocrinology and Diabetology, Department of Internal Medicine, , Medical University of Graz, ; Graz, Austria
                [2 ]ISNI 0000 0001 0658 8800, GRID grid.4827.9, Diabetes Research Group, School of Medicine, , Swansea University, ; Swansea, United Kingdom
                [3 ]ISNI 0000 0001 0658 8800, GRID grid.4827.9, Applied Sport, Technology, Exercise and Medicine Research Centre (A-STEM), College of Engineering, , Swansea University, ; Swansea, United Kingdom
                [4 ]Department of Cardiology and Intensive Care Medicine, LKH Graz Süd-West, Graz, Austria
                [5 ]ISNI 0000 0000 8988 2476, GRID grid.11598.34, Intensive Care Unit, Department of Internal Medicine, , Medical University of Graz, ; Graz, Austria
                [6 ]ISNI 0000 0000 8988 2476, GRID grid.11598.34, Department of Anaesthesiology, , Medical University of Graz, ; Graz, Austria
                [7 ]ISNI 0000 0004 0644 9589, GRID grid.8684.2, Institute for Biomedicine and Health Sciences, , Joanneum Research GmbH, HEALTH, ; Graz, Austria
                [8 ]ISNI 0000 0000 8988 2476, GRID grid.11598.34, Division of Gastroenterology and Hepatology, Department of Internal Medicine, , Medical University of Graz, ; Graz, Austria
                Author information
                http://orcid.org/0000-0001-7854-4233
                Article
                19768
                10.1038/s41598-018-19768-2
                5775252
                29352162
                49f30bec-a21f-49a3-9b06-8b8901e42f6a
                © The Author(s) 2018

                Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.

                History
                : 4 July 2017
                : 8 January 2018
                Categories
                Article
                Custom metadata
                © The Author(s) 2018

                Uncategorized
                Uncategorized

                Comments

                Comment on this article