      Capsular Ligament Function After Total Hip Arthroplasty

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background:

          The hip joint capsule passively restrains extreme range of motion, protecting the native hip against impingement, dislocation, and edge-loading. We hypothesized that following total hip arthroplasty (THA), the reduced femoral head size impairs this protective biomechanical function.

          Methods:

          In cadavers, THA was performed through the acetabular medial wall, preserving the entire capsule, and avoiding the targeting of a particular surgical approach. Eight hips were examined. Capsular function was measured by rotating the hip in 5 positions. Three head sizes (28, 32, and 36 mm) with 3 neck lengths (anatomical 0, +5, and +10 mm) were compared.

          Results:

          Internal and external rotation range of motion increased following THA, indicating late engagement of the capsule and reduced biomechanical function (p < 0.05). Internal rotation was affected more than external. Increasing neck length reduced this hypermobility, while too much lengthening caused nonphysiological restriction of external rotation. Larger head sizes only slightly reduced hypermobility.

          Conclusions:

          Following THA, the capsular ligaments were unable to wrap around the reduced-diameter femoral head to restrain extreme range of motion. The posterior capsule was the most affected, indicating that native posterior capsule preservation is not advantageous, at least in the short term. Insufficient neck length could cause capsular dysfunction even if native ligament anatomy is preserved, while increased neck length could overtighten the anterior capsule.

          Clinical Relevance:

          Increased understanding of soft-tissue balancing following THA could help to prevent instability and improve early function. This study illustrates how head size and neck length influence the biomechanical function of the hip capsule in the early postoperative period.

          Related collections

          Most cited references45

          • Record: found
          • Abstract: found
          • Article: not found

          Hip contact forces and gait patterns from routine activities.

          In vivo loads acting at the hip joint have so far only been measured in few patients and without detailed documentation of gait data. Such information is required to test and improve wear, strength and fixation stability of hip implants. Measurements of hip contact forces with instrumented implants and synchronous analyses of gait patterns and ground reaction forces were performed in four patients during the most frequent activities of daily living. From the individual data sets an average was calculated. The paper focuses on the loading of the femoral implant component but complete data are additionally stored on an associated compact disc. It contains complete gait and hip contact force data as well as calculated muscle activities during walking and stair climbing and the frequencies of daily activities observed in hip patients. The mechanical loading and function of the hip joint and proximal femur is thereby completely documented. The average patient loaded his hip joint with 238% BW (percent of body weight) when walking at about 4 km/h and with slightly less when standing on one leg. This is below the levels previously reported for two other patients (Bergmann et al., Clinical Biomechanics 26 (1993) 969-990). When climbing upstairs the joint contact force is 251% BW which is less than 260% BW when going downstairs. Inwards torsion of the implant is probably critical for the stem fixation. On average it is 23% larger when going upstairs than during normal level walking. The inter- and intra-individual variations during stair climbing are large and the highest torque values are 83% larger than during normal walking. Because the hip joint loading during all other common activities of most hip patients are comparably small (except during stumbling), implants should mainly be tested with loading conditions that mimic walking and stair climbing.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            The definition and measurement of acetabular orientation.

            The orientation of an acetabulum or an acetabular prosthesis may be described by its inclination and anteversion. Orientation can be assessed anatomically, radiographically, and by direct observation at operation. The angles of inclination and anteversion determined by these three methods differ because they have different spatial arrangements. There are therefore three distinct definitions of inclination and anteversion. This paper analyses the differences between the definitions and provides nomograms to convert from one to another. It is recommended that the operative definitions be used to describe the orientation of prostheses and that the anatomical definitions be used for dysplastic acetabula.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Single-incision anterior approach for total hip arthroplasty on an orthopaedic table.

              Dislocation remains the leading early complication of total hip arthroplasty; surgical approach and implant positioning have been recognized as factors influencing total hip arthroplasty stability. We describe a total hip arthroplasty technique done through a single, tissue sparing anterior approach that allows implantation of the femoral and acetabular components without detaching or sectioning any of the muscles and tendons around the hip joint. A series of 437 consecutive, unselected patients who had 494 primary total hip arthroplasty surgeries done through an anterior approach on an orthopaedic table from September 1996 to September 2004 was reviewed. There were 54 hybrid and 442 uncemented hips in the 437 patients (57 bilateral). The average patient age was 64 years. Radiographic analysis showed an average abduction angle of 42 degrees , with 96% in the range of 35 degrees to 50 degrees abduction. The average cup anteversion was 19 degrees with 93% within the target range of 10 degrees to 25 degrees . Postoperative leg length discrepancy averaged 3 +/- 2 mm (range, 0-26 mm). Three patients sustained dislocations for an overall dislocation rate of 0.61%, and no patients required revision surgery for recurrent dislocation. There were 17 operative complications, including one deep infection, three wound infections, one transient femoral nerve palsy, three greater trochanter fracture, two femoral shaft fractures four calcar fractures, and three ankle fractures. Operative time averaged 75 minutes (range 40-150 minutes), and the average blood loss was 350 mL (range, 100-1300 mL). The mean hospital stay was 3 days (range, 1-17 days). The anterior approach on the orthopaedic table is a minimally invasive technique applicable to all primary hip patients. This technique allows accurate and reproducible component positioning and leg-length restoration and does not increase the rate of hip dislocation. Therapeutic study, Level IV-1 (case series). See the Guidelines for Authors for a complete description of levels of evidence.
                Bookmark

                Author and article information

                Journal
                J Bone Joint Surg Am
                J Bone Joint Surg Am
                jbjsam
                jbjsam
                The Journal of Bone and Joint Surgery. American Volume
                The Journal of Bone and Joint Surgery, Inc.
                0021-9355
                1535-1386
                18 July 2018
                18 July 2018
                18 July 2018
                : 100
                : 14
                : e94 (pp. 1-10)
                Affiliations
                [1 ]Department of Mechanical Engineering, Imperial College London, London, United Kingdom
                [2 ]The London Hip Unit, London, United Kingdom
                Author notes
                [a ]E-mail address for J.R.T. Jeffers: j.jeffers@ 123456imperial.ac.uk
                Article
                JBJS-D-17-00251 %art-id%
                10.2106/JBJS.17.00251
                6636890
                30020129
                4a2b402d-e62a-4db4-8449-ae6f271a01c7
                Copyright © 2018 by The Authors. Published by The Journal of Bone and Joint Surgery, Incorporated.

                This is an open access article distributed under the Creative Commons Attribution License 4.0 (CCBY), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                Page count
                Pages: 10
                Categories
                0010
                0070
                Scientific Articles
                Custom metadata
                TRUE
                ONLINE-ONLY

                Comments

                Comment on this article