      Performance characteristics of a novel megavoltage cone-beam-computed tomography device.

      1 , , ,
      Physics in medicine and biology
      IOP Publishing

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          In this work, the image quality of a novel megavoltage cone-beam-computed tomography (CBCT) scanner is compared to three other image-guided radiation therapy devices by analysing images of different-sized quality assurance phantoms. The following devices are compared in terms of image uniformity, signal-to-noise ratio, contrast-to-noise ratio (CNR), electron density to HU conversion, presampling modulation transfer function (MTF(pre)) and combined spatial resolution and noise (Q-factor): (i) the Siemens Artiste kilovoltage (kV) (121 kV) CBCT device, (ii) the Artiste treatment beam line (TBL), 6 MV, (iii) the Tomotherapy (3.5 MV) fan-beam CT and (iv) Siemens' novel approach using a carbon target for a dedicated imaging beam line (IBL), 4.2 MV. Machine settings were selected to produce the same imaging dose for all devices. For a head phantom, IBL scans display CNR values 2.6 ± 0.3 times higher than for the TBL at the same dose level (for a CT-number range of -200 to -60 HU). kV CBCT, on the other hand, displays CNR values 7.9 ± 0.3 times higher than the IBL. There was no significant deviation in spatial resolution between IBL, TBL and Tomotherapy in terms of 50% and 10% MTF(pre). For kV CBCT, the MTF(pre) was significantly higher than those for other devices. In our Q-factor analysis, the IBL (14.6) scores higher than the TBL (7.9) and Tomotherapy (9.7) due to its lower noise level. The linearity of electron density to HU conversion is demonstrated for different-sized phantoms. Employing the IBL instead of the TBL significantly reduces the imaging dose by up to a factor of 5 at a constant image quality level, providing an immediate benefit for the patient.

          Related collections

          Author and article information

          Journal
          Phys Med Biol
          Physics in medicine and biology
          IOP Publishing
          1361-6560
          0031-9155
          Feb 07 2012
          : 57
          : 3
          Affiliations
          [1 ] DKFZ, Im Neuenheimer Feld 280, 69120 Heidelberg, Germany. m.fast@dkfz.de
          Article
          10.1088/0031-9155/57/3/N15
          22251668
          4d207163-36ec-420c-8af7-b3c74a209e6d
          History

          Comments

          Comment on this article