      Performance assessment of standard algorithms for dynamic R-T interval measurement: comparison between R-Tapex and R-T(end) approach.

      Medical & Biological Engineering & Computing
      Algorithms, Computer Simulation, Electrocardiography, Ambulatory, Heart, physiopathology, Humans, Models, Cardiovascular, Sensitivity and Specificity

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Three automatic approaches to ventricular repolarisation duration measurement (R-Tapex, R-T(end threshold) and R-T(end fitting) methods) are compared on computer-generated and real ECG signals, in relation to their reliability in the presence of the most common electrocardiographic artefacts (i.e. additive broadband noise and additive and multiplicative periodical disturbances). Simulations permit the evaluation of the amount of R-T beat-to-beat variability induced by the artefacts. The R-T(end threshold) method performs better than the R-T(end fitting) one, and, hence, the latter should be used with caution when R-T(end) variability is addressed. Whereas the R-Tapex method is more robust with regard to broadband noise than the R-T(end threshold) one, the reverse situation is observed in the presence of periodical amplitude modulations. A high level of broadband noise dose not prevent the detection of the central frequency of underlying R-T periodical changes. Comparison between the power spectra of the beat-to-beat R-T variability series obtained from three orthogonal ECG leads (X,Y,Z) is used to assess the amount of real and artefactual variability in 13 normal subjects at rest. The R-Tapex series displays rhythms at high frequency (HF) with a percentage power on the Z lead (57.1 +/- 4.9) greater than that on the X and Y leads (41.9 +/- 4.6 and 46.1 +/- 4.9, respectively), probably because of respiratory-related artefacts affecting the Z lead more remarkably. More uniform HF power distributions over X,Y,Z leads are observed in the R-T(end threshold) series (31.8 +/- 3.8, 39.2 +/- 4.1 and 35.1 +/- 4.2, respectively), thus suggesting minor sensitivity of the R-T(end threshold) measure to respiratory-related artefacts.

          Related collections

          Author and article information

          Journal
          9614746
          10.1007/BF02522855

          Chemistry
          Algorithms,Computer Simulation,Electrocardiography, Ambulatory,Heart,physiopathology,Humans,Models, Cardiovascular,Sensitivity and Specificity

          Comments

          Comment on this article