      Quality control analysis of test items based on Philips Digital Diagnost DR

      research-article

      Read this article at

      ScienceOpenPublisher
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Objective To explore the quality control results of Digital Diagnost digital radiography (DR), and also to analyze the differences in signal transfer characteristics (STP) of DR from different manufacturers, in order to provide a reference for quality control of testers and daily supervision of administrative staff.

          Methods According to WS 76—2020 “Specification for testing of quality control in medical X-ray diagnostic equipment”, the test items such as automatic exposure control (AEC) repeatability, AEC response, detector dose indication (DDI), STP, high-contrast resolution, and low-contrast resolution of Digital Diagnost DR were qualitatively controlled, and several common DRs on the market were selected to analyze their STP functional relationships.

          Results The baseline values of DDI, high-contrast resolution, and low-contrast resolution of Digital Diagnost DR were 19 733, 2.8 Lp/mm, and H5 of image central pixel value, respectively; the quality control results of other test items met the requirements of WS 76—2020. The STP functional relationships of common DRs were mainly linear and logarithmic, but different models of DR from the same manufacturer might also have different STP functional relationships.

          Conclusion When testing some special test items of DR, attention should be paid to the influence of a variety of factors, while refining the test steps and condition settings for the test items. The simulation should be performed in accordance with the functional relationship provided by the manufacturer when analyzing the STP.

          Abstract

          摘要: 目的 探讨 Digital Diagnost 型 DR 质量控制检测结果, 同时分析不同厂家 DR 的信号传递特性 (STP) 的差异, 以期为检测人员的质控和行政人员日常监管提供参考。 方法 依据 WS 76—2020 《医用 X 射线诊断设备质量控制检 测规范》, 对 Digital Diagnost 型 DR 的 AEC 重复性、AEC 响应、探测器剂量指示 (DDI)、STP、高对比度分辨力、低对 比度分辨力等检测项目进行质量控制检测, 同时选取市场上常见 DR 分析其 STP 函数关系。 结果 Digital Diagnost 型 DR 的 DDI、高对比度分辨力和低对比度分辨力的基线值分别为影像中央像素值 19 733、2.8 Lp/mm 和 H5; 其 他检测项目质控结果均满足 WS 76—2020 的要求。常见 DR 的 STP 函数关系主要为线性和对数关系, 但相同厂家不 同型号 DR 也可能具有不同的 STP 函数关系。 结论 DR 部分专用检测项目检测时, 应注意多种因素的影响, 同时细 化检测项目的检测步骤与条件设置。在对 STP 进行分析时应按照厂家提供的函数关系进行模拟。

          Related collections

          Author and article information

          Journal
          CJRH
          Chinese Journal of Radiological Health
          Chinese Preventive Medical Association (Ji’an, China )
          1004-714X
          01 October 2022
          01 December 2022
          : 31
          : 5
          : 553-557
          Affiliations
          [1] 1Hangzhou Hospital for the Prevention and Treatment of Occupational Disease, Hangzhou 310014 China
          [2] 2Hubei Centers for Disease Control and Prevention, Wuhan 430070 China
          [3] 3Hunan Prevention and Treatment Institute for Occupational Diseases, Changsha 410021 China
          [4] 4Institute of Radiation Medicine, Chinese Academy of Medical Sciences, Tianjin 300192 China
          Author notes
          Corresponding author: ZHAI Hezheng, E-mail: zhaihezheng@ 123456irm-cams.ac.cn
          Article
          j.issn.1004-714X.2022.05.005
          10.13491/j.issn.1004-714X.2022.05.005
          59c19f77-8067-4bfa-a7e0-8c863145f3b0
          © 2022 Chinese Journal of Radiological Health

          This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 Unported License (CC BY-NC 4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc/4.0/.

          History
          Categories
          Journal Article

          Medicine,Image processing,Radiology & Imaging,Bioinformatics & Computational biology,Health & Social care,Public health
          Signal transfer characteristics,Quality control,DR

          Comments

          Comment on this article