      Experimental verification of Advanced Collapsed‐cone Engine for use with a multichannel vaginal cylinder applicator

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Model‐based dose calculation algorithms have recently been incorporated into brachytherapy treatment planning systems, and their introduction requires critical evaluation before clinical implementation. Here, we present an experimental evaluation of Oncentra ® Brachy Advanced Collapsed‐cone Engine (ACE) for a multichannel vaginal cylinder (MCVC) applicator using radiochromic film. A uniform dose of 500 cGy was specified to the surface of the MCVC using the TG‐43 dose formalism under two conditions: (a) with only the central channel loaded or (b) only the peripheral channels loaded. Film measurements were made at the applicator surface and compared to the doses calculated using TG‐43, standard accuracy ACE (sACE), and high accuracy ACE (hACE). When the central channel of the applicator was used, the film measurements showed a dose increase of (11 ± 8)% (k = 2) above the two outer grooves on the applicator surface. This increase in dose was confirmed with the hACE calculations, but was not confirmed with the sACE calculations at the applicator surface. When the peripheral channels were used, a periodic azimuthal variation in measured dose was observed around the applicator. The sACE and hACE calculations confirmed this variation and agreed within 1% of each other at the applicator surface. Additionally for the film measurements with the central channel used, a baseline dose variation of (10 ± 4)% (k = 2) of the mean dose was observed azimuthally around the applicator surface, which can be explained by offset source positioning in the central channel.

          Related collections

          Most cited references25

          • Record: found
          • Abstract: found
          • Article: not found

          Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: current status and recommendations for clinical implementation.

          The charge of Task Group 186 (TG-186) is to provide guidance for early adopters of model-based dose calculation algorithms (MBDCAs) for brachytherapy (BT) dose calculations to ensure practice uniformity. Contrary to external beam radiotherapy, heterogeneity correction algorithms have only recently been made available to the BT community. Yet, BT dose calculation accuracy is highly dependent on scatter conditions and photoelectric effect cross-sections relative to water. In specific situations, differences between the current water-based BT dose calculation formalism (TG-43) and MBDCAs can lead to differences in calculated doses exceeding a factor of 10. MBDCAs raise three major issues that are not addressed by current guidance documents: (1) MBDCA calculated doses are sensitive to the dose specification medium, resulting in energy-dependent differences between dose calculated to water in a homogeneous water geometry (TG-43), dose calculated to the local medium in the heterogeneous medium, and the intermediate scenario of dose calculated to a small volume of water in the heterogeneous medium. (2) MBDCA doses are sensitive to voxel-by-voxel interaction cross sections. Neither conventional single-energy CT nor ICRU∕ICRP tissue composition compilations provide useful guidance for the task of assigning interaction cross sections to each voxel. (3) Since each patient-source-applicator combination is unique, having reference data for each possible combination to benchmark MBDCAs is an impractical strategy. Hence, a new commissioning process is required. TG-186 addresses in detail the above issues through the literature review and provides explicit recommendations based on the current state of knowledge. TG-43-based dose prescription and dose calculation remain in effect, with MBDCA dose reporting performed in parallel when available. In using MBDCAs, it is recommended that the radiation transport should be performed in the heterogeneous medium and, at minimum, the dose to the local medium be reported along with the TG-43 calculated doses. Assignments of voxel-by-voxel cross sections represent a particular challenge. Electron density information is readily extracted from CT imaging, but cannot be used to distinguish between different materials having the same density. Therefore, a recommendation is made to use a number of standardized materials to maintain uniformity across institutions. Sensitivity analysis shows that this recommendation offers increased accuracy over TG-43. MBDCA commissioning will share commonalities with current TG-43-based systems, but in addition there will be algorithm-specific tasks. Two levels of commissioning are recommended: reproducing TG-43 dose parameters and testing the advanced capabilities of MBDCAs. For validation of heterogeneity and scatter conditions, MBDCAs should mimic the 3D dose distributions from reference virtual geometries. Potential changes in BT dose prescriptions and MBDCA limitations are discussed. When data required for full MBDCA implementation are insufficient, interim recommendations are made and potential areas of research are identified. Application of TG-186 guidance should retain practice uniformity in transitioning from the TG-43 to the MBDCA approach.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous media.

            A Ahnesjö (2015)
            A method for photon beam dose calculations is described. The primary photon beam is raytraced through the patient, and the distribution of total radiant energy released into the patient is calculated. Polyenergetic energy deposition kernels are calculated from the spectrum of the beam, using a database of monoenergetic kernels. It is shown that the polyenergetic kernels can be analytically described with high precision by (A exp( -ar) + B exp( -br)/r2, where A, a, B, and b depend on the angle with respect to the impinging photons and the accelerating potential, and r is the radial distance. Numerical values of A, a, B, and b are derived and used to convolve energy deposition kernels with the total energy released per unit mass (TERMA) to yield dose distributions. The convolution is facilitated by the introduction of the collapsed cone approximation. In this approximation, all energy released into coaxial cones of equal solid angle, from volume elements on the cone axis, is rectilinearly transported, attenuated, and deposited in elements on the axis. Scaling of the kernels is implicitly done during the convolution procedure to fully account for inhomogeneities present in the irradiated volume. The number of computational operations needed to compute the dose with the method is proportional to the number of calculation points. The method is tested for five accelerating potentials; 4, 6, 10, 15, and 24 MV, and applied to two geometries; one is a stack of slabs of tissue media, and the other is a mediastinum-like phantom of cork and water. In these geometries, the EGS4 Monte Carlo system has been used to generate reference dose distributions with which the dose computed with the collapsed cone convolution method is compared. Generally, the agreement between the methods is excellent. Deviations are observed in situations of lateral charged particle disequilibrium in low density media, however, but the result is superior compared to that of the generalized Batho method.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO.

              This report addresses uncertainties pertaining to brachytherapy single-source dosimetry preceding clinical use. The International Organization for Standardization (ISO) Guide to the Expression of Uncertainty in Measurement (GUM) and the National Institute of Standards and Technology (NIST) Technical Note 1297 are taken as reference standards for uncertainty formalism. Uncertainties in using detectors to measure or utilizing Monte Carlo methods to estimate brachytherapy dose distributions are provided with discussion of the components intrinsic to the overall dosimetric assessment. Uncertainties provided are based on published observations and cited when available. The uncertainty propagation from the primary calibration standard through transfer to the clinic for air-kerma strength is covered first. Uncertainties in each of the brachytherapy dosimetry parameters of the TG-43 formalism are then explored, ending with transfer to the clinic and recommended approaches. Dosimetric uncertainties during treatment delivery are considered briefly but are not included in the detailed analysis. For low- and high-energy brachytherapy sources of low dose rate and high dose rate, a combined dosimetric uncertainty <5% (k=1) is estimated, which is consistent with prior literature estimates. Recommendations are provided for clinical medical physicists, dosimetry investigators, and source and treatment planning system manufacturers. These recommendations include the use of the GUM and NIST reports, a requirement of constancy of manufacturer source design, dosimetry investigator guidelines, provision of the lowest uncertainty for patient treatment dosimetry, and the establishment of an action level based on dosimetric uncertainty. These recommendations reflect the guidance of the American Association of Physicists in Medicine (AAPM) and the Groupe Européen de Curiethérapie-European Society for Therapeutic Radiology and Oncology (GEC-ESTRO) for their members and may also be used as guidance to manufacturers and regulatory agencies in developing good manufacturing practices for sources used in routine clinical treatments.
                Bookmark

                Author and article information

                Contributors
                cawstong@ualberta.ca
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                20 March 2017
                May 2017
                : 18
                : 3 ( doiID: 10.1002/acm2.2017.18.issue-3 )
                : 16-27
                Affiliations
                [ 1 ] Department of Medical Physics Cross Cancer Institute Edmonton AB T6G 1Z2 Canada
                [ 2 ] Department of Oncology University of Alberta Edmonton AB T6G 2R3 Canada
                Author notes
                [*] [* ]Author to whom correspondence should be addressed Brie Cawston‐Grant E‐mail: cawstong@ 123456ualberta.ca .
                Article
                ACM212061
                10.1002/acm2.12061
                5689852
                28317325
                5a8928ab-2f83-4fe7-a908-bcd4f3eecfc7
                © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

                This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 10 August 2016
                : 09 January 2017
                : 26 January 2017
                Page count
                Figures: 7, Tables: 3, Pages: 12, Words: 8177
                Categories
                87.53.Bn
                87.53.Bw
                87.55.d-
                87.55.kd
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                acm212061
                May 2017
                Converter:WILEY_ML3GV2_TO_NLMPMC version:5.2.5 mode:remove_FC converted:16.11.2017

                ace,brachytherapy,gynecology,phantoms,radiochromic film
                ace, brachytherapy, gynecology, phantoms, radiochromic film

                Comments

                Comment on this article